FujiFilm FN-PS551 Service Manual page 105

Network print server
Hide thumbs Also See for FN-PS551:
Table of Contents

Advertisement

Maintenance 7. Service Utility
(3)
From the [protocol] window, select the protocol ([DICOM], [TOSHIBA] or
[FINP]) for the relevant client.
NOTE
• For [DICOM], go to step (4).
• For [TOSHIBA], go to step (5).
• For [FINP] with software ver. A03 or later, go to step (6).
• When setting "DICOM: default", separate setting window is displayed for
[protocol] window, however, setting is not necessary for all parameters.
Settings can be made (setting is not possible with software ver. A05 or
later), but the following contents are set by internal process. Some of the
contents are different from the displayed default values.
• [Application Entity Title:]:
• [Attribute list error:]:
• [Attribute value out of range:]: [indicate](sends an error message)
• [Image size is larger than
• [N-Event-Report RQ:]:
• [Use Presentation LUT:]:
• [Data transfer timeout:]:
• [Use extension format:]:
For software ver. A05 or later, [avail] (available, default) or [inval] (invalid)
is set for the "DICOM: default" function.
(4)
For the DICOM protocol connection, set the following parameters.
• [Application Entity Title:]:
Enter an AE Title that identifies a client (up to 16 single-byte characters).
• [Attribute list error:]:
This parameter specifies the response to unexpected transmission errors
([not indicate] (default) ignores errors, [indicate] sends an error message).
• [Attribute value out of range:]:
This parameter specifies the response to out-of-range date errors ([not
indicate] (default) ignores errors, [indicate] sends an error message).
• [Image size is larger than image box:]:
BIB N-SET (image data receiving in frames) uses reduction to fit oversized
images into the image frame. This parameter specifies whether to issue
notification that reduction has been used ([not indicate] (default) sends no
notification, [indicate] sends an error message).
• [N-Event-Report RQ:]:
This parameter specifies whether to respond to N-Event-Report requests
([disabled] (default) send no response, [enabled] sends a response).
006-251-06
12.18.2001 FM3271
image box:]:
Maintenance-84
No relevance to "DICOM: default"
[indicate](sends an error message)
[indicate](sends an error message)
[enabled](sends a response)
[Yes](use LUT content sent by client)
(supporting software ver. A05 or later)
[enabled](timeout)
(supporting software ver. A05 or later)
[No](Not used)
FN-PS551 Service Manual Maintenance-84

Advertisement

Chapters

Table of Contents
loading

Table of Contents