Programming examples
Source-measure example — Table 8 shows a typical command sequence to source 10V
and measure current on the 10mA range.
Measure-only example — Table 9 shows a typical command sequence to measure current
only.
Table 8
Command sequence for source-measure example
Command*
*RST
:SOUR:FUNC VOLT
:SOUR:VOLT:MODE FIX
:SOUR:VOLT:RANG 20
:SOUR:VOLT:LEV 10
:SENS:FUNC "CURR"
:SENS:CURR:PROT 10e-3
:SENS:CURR:RANG 10e-3
:OUTP ON
:READ?
*SourceMeter must be addressed to talk after sending :READ? to trigger and acquire data.
Table 9
Command sequence for measure current only example
Command*
*RST
:SOUR:FUNC VOLT
:SOUR:VOLT:MODE FIX
:SOUR:VOLT:RANG 0.2
:SOUR:VOLT:LEV 0
:SENS:FUNC "CURR"
:SENS:CURR:PROT 100e-3
:SENS:CURR:RANG 10e-3
:OUTP ON
:READ?
*SourceMeter must be addressed to talk after sending :READ? to trigger and acquire data.
Comments
Restore GPIB defaults (one-shot source-measure mode).
Select voltage source function.
Select fixed voltage source mode.
Select 20V source range.
Set source output level to 10V.
Select current measurement function.
Set compliance limit to 10mA.
Select the 10mA measurement range.
Turn the output on.
Trigger and acquire one data string.
Comments
Restore GPIB defaults (one-shot measure mode).
Select voltage source function.
Select fixed voltage source mode.
Select 200mV source range.
Set source output level to 0.000mV.
Select current measurement function.
Set compliance limit to 100mA.
Select the 10mA measurement range.
Turn the output on.
Trigger and acquire one data string.
Quick Results Guide
17