Adjusting The Measurement Setup Settings; Adjusting The Trace Settings - Keysight A Series User Manual

Hide thumbs Also See for A Series:
Table of Contents

Advertisement

IQ Analyzer (IQA) – Option 351 (CPU2 Only)
I/Q Analyzer (IQA) Mode Settings

Adjusting the Measurement Setup Settings

— Press Meas Setup 4.
— Then choose from the following:

Adjusting the Trace Settings

The Trace menu is used to modify the existing predefined view, or create a
customized view. Window and Win State softkeys are used to modify layout of
view, the Data softkey is used to specify the displayed data results for the
current window. Please note only in User Defined measurement are all of the
data results from Spectrum measurement and Waveform measurement
310
Auto Trigger
annotated in the top left corner of the FieldFox screen.
If a valid trigger signal is not received before the specified Auto
Trig Time, a acquisition will occur automatically.
Enter 0 to set Auto Trigger OFF. When Auto Trigger is OFF, the
FieldFox does NOT perform an acquisition unless a valid
trigger signal is received.
(Default: 100 ms, minimum: 1 ms, maximum: 100s)
Average Type
Opens a menu of averaging softkeys.
Then select one of the following:
——Power
(Linear) Averaging – Best for measuring true power
levels. Used in Noise Marker Average. Mathematically, trace
noise is 2.5 dB higher than when using Log Average.
annotation is displayed.
——Log
Averaging (default) – Best for displaying Trace Averaging.
Also known as video averaging.
——Voltage
(Magnitude) Averaging – Converts the amplitudes to a
magnitude unit before averaging (e.g., Volts, Amps,
microVolts, milliVolts, etc.)
Average Count
The Average Count setting is used with the Average
Trace State. In this Trace State, the Average Count setting
determines the number of acquisitions to average. The higher the
average count, the greater the amount of noise reduction (default:
10, minimum: 1, maximum: 10000).
Sample Rate
Sets the I/Q data sample rate value (default: 12.5 MHz,
maximum: 12.5 MHz, couplings: 1.25 * IF Bandwidth).
Sample Period
Sets the 1/Sample Rate value (default: 80 ns,
minimum: 80 ns, maximum: 80 ms, couplings: 1/Sample Rate).
– While waiting for a valid trigger signal,
LgAvg
VAvg
annotation is displayed.
Keysight N9927-90020 User's Guide
Wait
is
PAvg
annotation is displayed.