Indoor and Outdoor Mapping – Option 352 (Requires Spectrum Analyzer Mode
(Option 233 on N991xA), and GPS Receiver (Option 307))
Troubleshooting

General Troubleshooting

If your Mapping setup is not behaving as expected:
— Verify setup is correct by referring to
IMPORTANT! If you forget you N9910X-028 Ball Antenna's static IP
address, refer to this link and the Support tab for assistance:
www.keysight.com/find/fieldfox
Support Document.

Port 5000 in Use Error

— If the cmd.exe window displays a "This port was picked because
502
page
469.
— Attenuator set to 0 dB
— Preamplifier set to ON
— Correct Band, Channel, and Center Frequency entered
— GNSS set to Internal
— Frequency Ref set to GPS
— ADC errors: Use Scale/Amptd >
level (dBm).
For some data views, you need to press More to view the
softkey.
— No data displayed: Press Meas Setup 4 then
——Verify the correct Antenna Settings have been entered.
— The FieldFox's displayed annotation indicates "Check LAN
connection. Check IP address." during pinging.
Verify:
——The LAN address for first 3 fields matches the Static IP address
of FieldFox. For example, if the Ball Antenna IP address is
192.168.1.10, set FieldFox Static IP address to 192.168.1.xx.
"xx" must be different number than shown in the last field of
the for Ball Antenna. Refer to
5000 is in use." error, then:
a. Type Ctrl + c in the command window
b. Press Enter
c. The command window should display text "Terminate batch job?
(Y/N)":
Type Y and then press Enter
"How to Set Up Mapping Settings" on
Range
to adjust the absolute power
Figure 14-15 on page
and then to the Phase Array Antenna
Keysight N9927-90020 User's Guide
Range
425.