TDR (Time Domain Reflectometry) - Option 215
TDR Measurement Settings
The Loss value for the measurement is interpolated from the Freq/Loss pairs at
the Center Freq:
The correction for loss at 5 meters in one direction: 0.18 dB/m * 5m = 0.9 dB.
All TDR measurements correct for loss for travel down the DUT and back, so
double the correction: 0.9 dB * 2 = 1.8 dB.

Window Settings

Window settings provide the ability to choose between optimizing TDR
measurements for resolving closely-spaced faults or for the ability to measure
low-level faults.

How to select Window settings

— Press Meas Setup 4.
— Then Settings.
— Then press
— Then press Edit.
— Then press
— Then press
— Again press Done.
94
1 GHz = 0.1 dB/m
1.8 GHz = 0.18 dB/m
2 GHz = 0.2 dB/m
arrows to move to the Window row.
Window
and choose from the following:
Maximum
(Maximum – Optimized for dynamic range) the noise floor
is lowered to provide the ability to measure low-level responses.
(Default setting)
Medium
– Compromise between Min and Max window settings.
Minimum
– Best Response Resolution, providing the ability to resolve
between two closely-spaced responses.
Done