Declarations of Conformity for this where in the EULA. Keysight shall be product and for other Keysight prod- under no obligation to update, revise or ucts may be downloaded from the WARNING otherwise modify the Software.
Caution, risk of danger (refer to this manual Caution, risk of electric shock for specific Warning or Caution information) Earth (ground) terminal Frame or chassis (ground) terminal Equipment protected throughout by double Protective earth (ground) terminal insulation or reinforced insulation Keysight U8903B User’s Guide...
Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards for design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer’s failure to comply with these requirements.
Environmental Conditions The U8903B is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for this instrument. Environmental cond ition Requirement Operating condition – 0 °C to 55 °C Temperature Storage condition –...
To return this unwanted instrument, contact your nearest Keysight Service Center, or visit www.keysight.com/environment/product for more information. Sales and Technical Support To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites: – www.keysight.com/find/U8903B...
U8903B Bluetooth generator and analyzer block diagram ...56 Figure 2-5 Direct U8903B Bluetooth audio connection to DUT example ..57 Figure 2-6 Wireless U8903B Bluetooth audio connection to DUT example .
Page 18
Figure 3-26 Digital Generator > Waveform Config menu page (Rectangular noise) Figure 3-27 Bluetooth Generator > Waveform Config menu page (Rectangular noise) Figure 3-28 Digital Generator > Waveform Config menu page (Triangular noise) . . 118 Keysight U8903B User’s Guide...
Page 19
Bluetooth Analyzer > Link Config > Common Settings menu page 1 . .150 Figure 3-64 Bluetooth Analyzer > Link Config > Common Settings menu page 2 . .150 Figure 3-65 Bluetooth Analyzer > Link Config > Common Settings menu page 3 . .151 Keysight U8903B User’s Guide...
Page 20
Analog Analyzer > Functions > SMPTE IMD menu page 1 ..210 Figure 4-34 Digital Analyzer > Functions > SMPTE IMD menu page 1 ... 211 Keysight U8903B User’s Guide...
..........37 Table 1-3 U8903B front panel description ....... .38...
Page 26
Bluetooth Analyzer > Link Config > Common Settings menu description Table 3-36 Bluetooth Analyzer > Link Config > A2DP Settings menu description . 153 Table 3-37 Bluetooth Analyzer > Link Config > HFP Settings menu description . . 154 Keysight U8903B User’s Guide...
Page 27
Table 4-21 Analog/Digital/Bluetooth Analyzer > Functions > X-Talk menu description Table 4-22 Analog Analyzer > Filter Config menu description ....231 Keysight U8903B User’s Guide...
Page 32
THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight U8903B User’s Guide...
Page 33
Access the help mode Update the U8903B Perform self-test Add or remove U8903B options This chapter teaches you how to set up the U8903B for the first time. A summary of all the features of the U8903B is also given.
FFT graph analysis, as well as sweep capability for frequency and amplitude. The U8903B also allows you to create test sequences. Refer to Chapter 7, "Test Sequence Application" for more information. To search for firmware updates for the U8903B, go to the Keysight U8903B firmware update Web site at www.keysight.com/find/audioanalyzer_firmware.
Also, include a brief description of the problem. Ventilation The U8903B can operate within the temperature range of 0 °C to 55 °C. The U8903B is cooled by drawing air through the sides and bottom at the front of the U8903B, and exhausting it through the ventilation holes on the sides and top at the rear of the U8903B.
Getting Started Standard Shipped Items Verify that you have received the following items. If anything is missing or damaged, please contact the nearest Keysight Sales Office. – U8903B Performance Audio Analyzer – Product Reference CD-ROM – Power cord – USB cable –...
Bluetooth card with 2 RF connectors. Bluetooth card with 2 RF connectors at a secondary option slot. U8903B-BL2 This is only applicable when U8903B-AN8 is selected. Wide bandwidth option - 1.5 MHz (fixed perpetual license). N3431A This is only available for channel 1 and channel 2 analog analyzer.
Press Stop All to stop all generator and test sequence operations. Refer to “RUN CONTROL Panel” on page for more information. Enables access to the U8903B commonly used graph functions. Refer to “GRAPH Panel” on GRAPH panel page for more information.
Page 39
Table 1-3 U8903B front panel description (continued) Item Description Enables access to the U8903B system functions. Refer to “SYSTEM Panel” on page SYSTEM panel more information. The headphone jack can be operated in stereo or mono mode. Refer to “Input Headphone jack Configuration (Analog Analyzer)”...
Receives and transmits audio signal through the RF1 and RF2 connectors. The RF1 connector Bluetooth RF ports is the default port and the RF2 port can be selected in the U8903B’s GUI. Only one RF port can be used at a time.
Page 41
AC power port Connects to an AC line voltage. VGA port Allows an external monitor to be connected to the U8903B. Allows the U8903B to be controlled remotely over the GPIB (General Purpose Interface Bus) GPIB port interface. USB port Allows an external USB flash storage to be connected to the U8903B.
Displays a maximum of 7 softkey menu options for the selected panel. Status bar Displays the status of an action such as information, warning, or error messages. Refer to “Display mode” on page for more information on the U8903B display layout. NOTE Keysight U8903B User’s Guide...
Getting Started Power on the U8903B Connect one end of the power cord to the U8903B rear panel AC power inlet and the other end to an AC voltage source. The U8903B will automatically adjust to the correct line voltage in the range of 100 VAC to 240 VAC.
CHANNEL/DATA panel and on the SYSTEM panel. When the U8903B help mode is activated, the front panel keys will not execute their NOTE normal functions when pressed. An example of a help mode is shown in Figure 1-5. The help mode icon will appear at the top-right of the title bar.
Recall menu page. EULA Press the EULA softkey to display the U8903B EULA (End User License Agreement) page. 5 Press the Load File softkey and the recall menu page is displayed to enable you to select the source location.
The System menu page is only available in the Standard mode. Refer to “Menu Key” on NOTE page for more information on the U8903B available modes. The firmware update process may take 20 minutes to 40 minutes to complete. CAUTION DO NOT power off the U8903B during the firmware update process. Keysight U8903B User’s Guide...
Getting Started Perform self-test Perform the following procedure to run the U8903B self-test. 1 Press on the SYSTEM panel. 2 From the System menu page, press the Service softkey. 3 From the Service menu page, press the Sel f-Test softkey. The Self-Test menu page is displayed as shown in Figure 1-8.
Table 1-8 System > Service > Options menu page Menu Description Press the Add Option softkey to load the U8903B option file. The Recall menu page will be displayed to select the Add Option source location. Remove Option Press the Remove Option softkey to remove the selected U8903B option.
1 Select the U8903B option to be removed from the list. 2 Press the Remove Option softkey to remove the U8903B option as shown in Figure 1-9. The System menu page is only available in the Standard mode. Refer to “Menu Key” on...
Keysight U8903B Performance Audio Analyzer User’s Guide Operation and Features Test Capabilities U8903B Block Diagram Analog audio interface Digital audio interface Bluetooth audio interface Bluetooth audio connection examples Navigation and CHANNEL/DATA Panels Menu Key Standard View Test Sequence Application (Test Seq App)
Operation and Features Test Capabilities The U8903B is capable of testing a broad range of audio-related devices and components for research and development, manufacturing, and quality assurance applications. Examples of the products that can be tested are listed follows. – Multichannel home theater systems –...
Operation and Features U8903B Block Diagram Analog audio interface Signal generation A simplified U8903B analog generator block diagram is shown as follows. ANALOG GENERATOR LOOPBACK FILTER RANGING ISOLATED GROUND Figure 2-1 U8903B analog generator block diagram The digital signal processing (DSP) generates all the required waveforms, except for square wave, digitally.
Page 54
Operation and Features Measurement A simplified U8903B analog analyzer block diagram is shown as follows. ANALOG ANALYZER AC/DC Coupling LOOPBACK RANGING FILTER 600 Ω 300 Ω 200 kΩ 100 kΩ Termination Figure 2-2 U8903B analog analyzer block diagram An audio signal can enter the analog analyzer through either the balanced (XLR) or unbalanced (BNC) input signal connector.
Operation and Features Digital audio interface A simplified U8903B digital audio interface block diagram is shown as follows. DIGITAL GENERATOR TOSLINK Serial audio Serial audio 25-pin female D-SUB connector DIGITAL ANALYZER Serial audio TOSLINK Figure 2-3 U8903B digital generator and analyzer block diagram Signal generation The digital generator generates all the required waveforms digitally.
RF interface. Finally, the waveform is transmitted to the Bluetooth audio devices through the RF1 connector. The RF1 connector is the default port and the RF2 port can be selected in the U8903B’s GUI. Measurement The audio signal enters the Bluetooth analyzer through the RF1 connector.
Operation and Features Bluetooth audio connection examples An example of a direct U8903B Bluetooth audio connection to DUT using an RF cable is shown as follows. RF cable Figure 2-5 Direct U8903B Bluetooth audio connection to DUT example An example of a wireless U8903B Bluetooth audio connection using an antenna is shown as follows.
Use the numeric keys to enter alphanumeric data by using the number keys and decimal point, or select the channel number. Numeric keys In DTMF single mode, press to generate the ‘*’ DTMF tone. Keysight U8903B User’s Guide...
Page 59
Refer to “GRAPH Panel” on page and “SYSTEM Panel” on page for more information. Escape key Press to cancel a selected action. Back space key Press to delete the character to the left of the cursor. Keysight U8903B User’s Guide...
U8903B modes description Mode Description Standard View The Standard View mode is the default mode for the U8903B. Refer to “Standard View” on page for more information. Test Seq App Refer to “Test Sequence Application (Test Seq App)” on page for more information.
SYSTEM panel to access the U8903B system configurations. You can view the error messages, configure I/O settings, update the U8903B, perform self-tests, configure the U8903B settings, configure the HP8903B mode settings, and others. Refer to “System” on page for more information on the system configurations.
Refer to Chapter 7, “Test Sequence Application” for more information. HP8903B Press and select HP8903B to access the HP8903B mode. The HP8903B mode Chapter 8 emulates the HP8903B audio analyzer behavior. Refer to , “HP8903B” for more information. Keysight U8903B User’s Guide...
Operation and Features FUNCTION Panel The FUNCTION panel provides quick access to the U8903B main functions. Table 2-3 FUNCTION panel description FUNCTION panel Description Press to switch between the analog, digital, or Bluetooth Interface interface. Press to switch between the generator or analyzer mode.
The full screen function is only available in the graph analysis mode. NOTE The full screen graph display is displayed as shown in Figure 2-9. To exit from the full screen graph display, press Figure 2-9 Full screen graph display Keysight U8903B User’s Guide...
Operation and Features Display mode The U8903B allows you to display up to ten panels on the main display in the analyzer and generator mode. Figure 2-10 4-panel view Figure 2-11 10-panel view Keysight U8903B User’s Guide...
Operation and Features The U8903B allows you to display up to two panels on the main display in the graph analysis mode. Figure 2-12 Graph analysis mode 2-panel view Keysight U8903B User’s Guide...
Refer to “Zoom” on page for more information. Edit Zoom Press to configure the section of the graph (Shift + Zoom) to be magnified. Refer to “Ed it zoom” on page for more information. Keysight U8903B User’s Guide...
If the active maker is off, the marker will be turned on before performing the previous minimum operation. Threshold Press the Threshold softkey to configure the peak search setting. Refer to “Threshold” on page for more information. Keysight U8903B User’s Guide...
Press the Min Thres softkey to set the minimum threshold value. A minimum has to be lower than the minimum threshold Min Thres value in order to be taken into consideration when a minimum search is performed. Keysight U8903B User’s Guide...
Press the Ref Mkr softkey to select the reference marker number. The reference marker has to be specified to perform the delta marker calculations in the marker table. Ref Mkr – OFF – M1 to M8 Keysight U8903B User’s Guide...
Page 71
You can choose to have the active marker readings to be at the top left corner of the graph, or at the marker location itself. When marker table is enabled, the marker readings at the marker location are automatically hidden. Keysight U8903B User’s Guide...
Press the Move to 3 dB softkey to place the marker on the nearest –3 dB value data point. Move to 6 dB Press the Move to 6 dB softkey to place the marker on the nearest –6 dB value data point. Keysight U8903B User’s Guide...
Marker > Harmonics menu page Table 2-9 Marker > Harmonics menu description Menu Description Trace Press the Trace softkey to set the trace number to place the marker. Enabled Press the Enabled softkey to show or hide the markers. Keysight U8903B User’s Guide...
Press the AutoScale Y softkey to perform autoscaling on the Y-axis by searching for the most optimum scale (top and AutoScale Y bottom values) for the Y-axis based on the data being displayed in the graph. Keysight U8903B User’s Guide...
Operation and Features Zoom The U8903B screen is split into two windows in the zoom view. The top window is a normal graph window, and the bottom window displays a magnified representation of the traces in the top window. The data in the bottom window will be more detailed due to the screen resolution difference between both windows.
The edit zoom mode is displayed as shown in Figure 2-20. Figure 2-20 Edit zoom mode display To exit the edit zoom mode, press on the CHANNEL/DATA panel and on the GRAPH panel. To exit the zoom view, press on the GRAPH panel. Keysight U8903B User’s Guide...
Table 1-5 for more information on the remote control icon. Press to lock or unlock the front panel keys in the local Lock/Unlock mode. A dialog box will pop-up for confirmation. Refer to Table 1-5 for more information on the front panel keys locked icon. Keysight U8903B User’s Guide...
Mode operations such as auto query, sweep, and test sequence will be aborted. Press the To Factory Settings softkey to reset the U8903B to the factory default settings, delete all the files saved in the To Factory Settings internal flash memory (for example, state file, arbitrary waveform, test sequence project files, and so on), and reset the GUI state to the Standard View mode.
Operation and Features Utility The Utility menu page allows you to save the U8903B state to a file, or recall the U8903B state from a file. The Utility menu page is only available in the Standard View mode. Refer to “Menu Key” on...
Press the Copy Graph From softkey to copy other channel graph state to the current channel. File Explorer Press the File Explorer softkey to access the file explorer menu. Refer to “File Explorer” on page 81. Keysight U8903B User’s Guide...
Operation and Features File Explorer File explorer allows you to rename, delete, copy, or move files from the U8903B’s internal storage or an external flash drive. Figure 2-24 File Explorer menu page Table 2-15 File Explorer menu description Menu Description Select Press the Select softkey to select the folder to browse in the file explorer mode.
System The System menu page allows you to view error messages, configure I/O settings, update the U8903B, perform self-tests, configure the U8903B settings, configure the HP8903B mode settings, and others. – The System menu page is only available in the Standard View mode. Refer to “Menu NOTE Key”...
Load File EventLog_CE.xml. The older log files are named as EventLog_CE.<yyyyMMdd>_n.xml, where yyyy = year, MM = month, dd = day, and n = version. The U8903B will keep the log files up to seven days. Keysight U8903B User’s Guide...
Press the Lan Settings softkey to configure the LAN settings. Refer to Table 2-18 for more information. GPIB Press the GPIB softkey to set the desired GPIB address. Press the FTP softkey to enable or disable the FTP control. Keysight U8903B User’s Guide...
Press the DNS 2 softkey to set the LAN DNS 2 address. DNS 2 This setting is only available when the LAN configuration mode is set to Manual. Update Refer to “Update the U8903B” on page 46. Keysight U8903B User’s Guide...
It does not affect the U8903B settings and license information. Press the Options softkey to display the installed U8903B options. You may also add or remove the U8903B options from Options the Options menu page.
Press the Power Up State softkey to select the power up state. – Last Power Up State The U8903B will start up with the last saved settings. – Default The U8903B will start up with the factory default settings. Keysight U8903B User’s Guide...
Page 88
Enable of disable the aux audio monitor. – Audio Volume Adjust the aux audio volume by using the up and down arrow keys or the knob. This is only available if the Audio Monitor is enabled. Keysight U8903B User’s Guide...
The HP8903B filter may be different according to the model and the purchased HP8903B option. The default left filter is None. – None – A-Weighing Left Fil ter – CCIR 1k wtd – CCIR 2k wtd – C-Message – CCITT – 400 Hz Keysight U8903B User’s Guide...
Press the Fan & Temperature softkey to display the temperatures of the available cards and the speed of the three 80 mm fans in the U8903B. This menu page is for service usage only. Refer to the U8903B Service Guide for more information.
Press the Aud io Mode softkey to select the aux audio mode type. Aud io Mode – Stereo – Mono Volume Press the Volume softkey to select the aux audio volume. DC Output Press the DC Output softkey to set the aux DC output value. Keysight U8903B User’s Guide...
Operation and Features Board Info Press the Board Info softkey to display the U8903B available cards information. Figure 2-34 System > Board Info menu page Legacy Sweep Press the Legacy Status softkey to enable or disable the legacy sweep. When the legacy sweep is enabled, sweep is limited to analog and the sweep channel for the analyzer and generator is paired.
Press to start or stop the test sequence in the test sequence application mode. Stop All Press to stop all generator, sweep, and test sequence operations. Keysight U8903B User’s Guide...
Operation and Features Save and Recall Save The Save menu page allows you to save the U8903B state, graph analysis trace settings, sweep function points, or screen capture of the current display into a file. Figure 2-36 Save menu page...
Operation and Features Recall The Recall menu page allows you to recall the U8903B state, graph analysis trace settings, or sweep function points from a file. Figure 2-37 Recall menu page Table 2-25 Recall menu description Menu Description Recall Press the Recall softkey to recall the selected file.
Page 96
Operation and Features THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight U8903B User’s Guide...
“Output Configuration (Analog Generator)” on page for more information. Press the References softkey to configure the output references settings that are used in the unit conversion. Refer to References “References (Analog Generator)” on page for more information. Keysight U8903B User’s Guide...
“AES3/SPDIF Output Configuration (Digital Generator)” on page for more information. Press the References softkey to configure the output references settings that are used in the unit conversion. Refer to References “References (Digital and Bluetooth Generator)” on page for more information. Keysight U8903B User’s Guide...
Configuration (Bluetooth Generator)” on page for more information. Press the References softkey to configure the output references settings that are used in the unit conversion. Refer to References “References (Digital and Bluetooth Generator)” on page for more information. Keysight U8903B User’s Guide...
Figure 3-4, Figure 3-5, and Figure 3-6 for the analog, digital, and Bluetooth generator respectively. Figure 3-4 Analog Generator > Waveform Config menu page (sine waveform) Figure 3-5 Digital Generator > Waveform Config menu page (sine waveform) Keysight U8903B User’s Guide...
For a perfect sine waveform without any DC offset, Vpp is twice Vpeak and Vrms is equivalent to Vpeak/√2. Frequency Frequency refers to the reciprocal of the period of the signal. DC Offset DC offset refers to the DC component of the waveform. Keysight U8903B User’s Guide...
Figure 3-7, Figure 3-8, and Figure 3-8 for the analog, digital, and Bluetooth generator respectively. Figure 3-7 Analog Generator > Waveform Config menu page (variable phase waveform) Figure 3-8 Digital Generator > Waveform Config menu page (variable phase waveform) Keysight U8903B User’s Guide...
Phase -> 1 refers to the phase of the channel 2 sine waveform with reference to channel 1. A positive value will cause the Phase -> 1 channel 1 output to lead the channel 2 output. A negative value will cause the channel 2 output to lag the channel 1 output. Keysight U8903B User’s Guide...
Figure 3-10, Figure 3-11, and Figure 3-12 for the analog, digital, and Bluetooth generator respectively. Figure 3-10 Analog Generator > Waveform Config menu page (dual waveform) Figure 3-11 Digital Generator > Waveform Config menu page (dual waveform) Keysight U8903B User’s Guide...
– Linear The amplitude displayed is the linear summation result of the two levels. The level of each tone can be easily calculated back from the tone ratio. Keysight U8903B User’s Guide...
Figure 3-14, and Figure 3-15 for both the analog, digital, and Bluetooth generator respectively. Figure 3-13 Analog Generator > Waveform Config menu page (SMPTE 1:1 waveform) Figure 3-14 Digital Generator > Waveform Config menu page (SMPTE 1:1 waveform) Keysight U8903B User’s Guide...
– Linear The amplitude showed is the linear summation result of the two levels. The level of each tone can be easily calculated back from the tone ratio. Keysight U8903B User’s Guide...
Figure 3-17, and Figure 3-18 for the analog, digital, and Bluetooth generator respectively. Figure 3-16 Analog Generator > Waveform Config menu page (IEC 60118 waveform) Figure 3-17 Digital Generator > Waveform Config menu page (IEC 60118 waveform) Keysight U8903B User’s Guide...
– Linear The amplitude showed is the linear summation result of the two levels. The level of each tone can be easily calculated back from the tone ratio. Keysight U8903B User’s Guide...
Figure 3-20, and Figure 3-21 for the analog, digital, and Bluetooth generator respectively. Figure 3-19 Analog Generator > Waveform Config menu page (IEC 60268 waveform) Figure 3-20 Digital Generator > Waveform Config menu page (IEC 60268 waveform) Keysight U8903B User’s Guide...
– Linear The amplitude showed is the linear summation result of the two levels. The level of each tone can be easily calculated back from the tone ratio. Keysight U8903B User’s Guide...
Gaussian noise is displayed as shown in Figure 3-22, Figure 3-23, and Figure 3-24 for the analog, digital, and Bluetooth generator respectively. Figure 3-22 Analog Generator > Waveform Config menu page (Gaussian noise) Figure 3-23 Digital Generator > Waveform Config menu page (Gaussian noise) Keysight U8903B User’s Guide...
Amplitude can be expressed as FFS, dBFS, or %FS for the digital and Bluetooth generator. For a perfect sine waveform without any DC offset, Vpp is twice Vpeak and Vrms is equivalent to Vpeak/√2. DC Offset DC Offset refers to the DC component of the waveform. Keysight U8903B User’s Guide...
Rectangular noise is displayed as shown in Figure 3-25, Figure 3-26, and Figure 3-27 for the analog, digital, and Bluetooth generator respectively. Figure 3-25 Analog Generator > Waveform Config menu page (Rectangular noise) Figure 3-26 Digital Generator > Waveform Config menu page (Rectangular noise) Keysight U8903B User’s Guide...
Amplitude can be expressed as FFS, dBFS, or %FS for the digital and Bluetooth generator. For a perfect sine waveform without any DC offset, Vpp is twice Vpeak and Vrms is equivalent to Vpeak/√2. DC Offset DC Offset refers to the DC component of the waveform. Keysight U8903B User’s Guide...
Amplitude can be expressed as FFS, dBFS, or %FS for the digital generator. Amplitude For a perfect sine waveform without any DC offset, Vpp is twice Vpeak and Vrms is equivalent to Vpeak/√2. DC Offset DC Offset refers to the DC component of the waveform. Keysight U8903B User’s Guide...
Pink noise is displayed as shown in Figure 3-29, Figure 3-30, and Figure 3-31 for the analog, digital, and Bluetooth generator respectively. Figure 3-29 Analog Generator > Waveform Config menu page (Pink noise) Figure 3-30 Digital Generator > Waveform Config menu page (Pink noise) Keysight U8903B User’s Guide...
Amplitude can be expressed as FFS, dBFS, or %FS for the digital and Bluetooth generator. For a perfect sine waveform without any DC offset, Vpp is twice Vpeak, while Vrms is equivalent to Vpeak/√2. DC Offset DC Offset refers to the DC component of the waveform. Keysight U8903B User’s Guide...
Figure 3-32, Figure 3-33, and Figure 3-34 for the analog, digital, and Bluetooth generator respectively. Figure 3-32 Analog Generator > Waveform Config menu page (square waveform) Figure 3-33 Digital Generator > Waveform Config menu page (square waveform) Keysight U8903B User’s Guide...
For a perfect sine waveform without any DC offset, Vpp is twice Vpeak and Vrms is equivalent to Vpeak/√2. Frequency Frequency refers to the frequency of the square waveform. This is only available for the digital generator. DC Offset DC Offset refers to the DC component of the waveform. Keysight U8903B User’s Guide...
DC signal is displayed as shown in Figure 3-35. Figure 3-35 Analog Generator > Waveform Config menu page (DC signal) Table 3-15 Analog Generator > Waveform Config menu description (DC signal) Menu Description Voltage Voltage refers to the amplitude of the composite signal. Keysight U8903B User’s Guide...
Figure 3-37, Figure 3-38, and Figure 3-38 for the analog, digital, and Bluetooth generator respectively. Figure 3-36 Analog Generator > Waveform Config menu page (sine burst) Figure 3-37 Digital Generator > Waveform Config menu page (sine burst) Keysight U8903B User’s Guide...
Period refers to the number of cycles from the beginning of one burst to the beginning of the next burst. The period value Period must be at least one cycle greater than the burst on. Low Level Low Level refers to the amplitude ratio of Burst On over Burst Off. Keysight U8903B User’s Guide...
For a perfect sine waveform without any DC offset, Vpp is twice Vpeak and Vrms is equivalent to Vpeak/√2. Frequency Frequency refers to the frequency for the first sine component. DC Offset DC Offset refers to the DC component of the waveform. Keysight U8903B User’s Guide...
Figure 3-40 Digital Generator > Waveform Config menu page (monotonicity) Table 3-18 Digital Generator > Waveform Config menu description (monotonicity) Menu Description Samples/Step Samples/Step refers to the duration for each half cycle of the square wave. Keysight U8903B User’s Guide...
Figure 3-41 and Figure 3-42 for the digital and Bluetooth generator respectively. Figure 3-41 Digital Generator > Waveform Config menu page (constant) Figure 3-42 Bluetooth Generator > Waveform Config menu page (constant) Keysight U8903B User’s Guide...
Audio Generator Functions Table 3-19 Digital/Bluetooth Generator > Waveform Config menu description (constant) Menu Description Voltage Press the Voltage softkey to select the output voltage or set your desired output voltage. Keysight U8903B User’s Guide...
Figure 3-43 Digital Generator > Waveform Config menu page (walking zero) Table 3-20 Digital Generator > Waveform Config menu description (walking zero) Menu Description Samples/Step Samples/Step refers to the speed the single bit 0 is incremented. Keysight U8903B User’s Guide...
Figure 3-44 Digital Generator > Waveform Config menu page (walking one) Table 3-21 Digital Generator > Waveform Config menu description (walking one) Menu Description Samples/Step Samples/Step refers to the speed the single bit 1 is incremented. Keysight U8903B User’s Guide...
Audio Generator Functions Arbitrary waveform You can load a properly formatted waveform file into the U8903B as a sequence of waveform samples. The samples are outputted at the sample rate of the generator. The waveform files must be in the .arb or .wav file type. For .wav file type, mono or stereo audio with 8, 16, or 24 bits per sample are supported.
Press the Load File softkey to load a waveform from a file. Refer to “Recall” on page for more information on the Recall Load File menu page. Info Press the Info softkey to display the loaded waveform information. Keysight U8903B User’s Guide...
Figure 3-48, Figure 3-49 and Figure 3-50 for the analog, digital, and Bluetooth generator respectively. Figure 3-48 Analog Generator > Waveform Config menu page 1 (multitone waveform) Figure 3-49 Digital Generator > Waveform Config menu page 1 (multitone waveform) Keysight U8903B User’s Guide...
Press the Custom softkey to access the custom multitone menu page. You can set the frequency, amplitude, phase, and Custom optimize the crest factor for each tone. Refer to “Custom multitone” on page for more information on the custom multitone menu page. Keysight U8903B User’s Guide...
Figure 3-52, and Figure 3-53 for both the analog, digital, and Bluetooth generator respectively. Figure 3-51 Analog Generator > Waveform Config > Custom menu page 1 (multitone waveform) Figure 3-52 Digital Generator > Waveform Config > Custom menu page 1 (multitone waveform) Keysight U8903B User’s Guide...
Press the Add Below softkey to add the tone below the selected tone for the custom multitone waveform. Remove Press the Remove softkey to remove the selected tone for the custom multitone waveform. Clear Press the Clear softkey to clear all the tones in the custom multitone waveform list. Keysight U8903B User’s Guide...
This prevents any harmonics from being misinterpreted as part of the signal. The waveform configuration for DTMF is displayed as shown in Figure 3-54. Figure 3-54 Analog Generator > Waveform Config menu page 1 (DTMF) Keysight U8903B User’s Guide...
Press the Mode softkey to switch between Single or Sequence as the dial mode. Press the numeric keys on the CHANNEL/DATA panel to generate the respective DTMF tone as listed in Table 3-27. The Dial menu page is displayed as shown in Figure 3-55. Figure 3-55 Dial menu page Keysight U8903B User’s Guide...
CHANNEL/DATA panel to generate the ‘*’ DTMF tone. Press on the CHANNEL/DATA panel to generate the ‘#’ DTMF tone. A to D Press the A to D softkeys to generate the DTMF A to D tones respectively. Keysight U8903B User’s Guide...
Press the Impedance softkey to select the output impedance value. For Bal, Com, and IEC60268 – 600 Ω – 100 Ω Impedance – 40 Ω For UnBal – 600 Ω – 50 Ω – 20 Ω Keysight U8903B User’s Guide...
Page 142
Press the Max Vol tage softkey to set the maximum voltage. The maximum voltage value set will limit the maximum Max Vol tage amplitude output from the generator and prevent any excessive high voltage values from damaging the device under test (DUT). Keysight U8903B User’s Guide...
Press the Connector softkey to select the output connector type. – Bal Connector Balanced mode outputs a digital signal on the XLR output connector at the U8903B rear panel. – UnBal Unbalanced mode outputs a digital signal on the BNC output connector at the U8903B rear panel.
– 8 to 32 Press the Mul tiplier softkey to set the external reference clock multiplier value. The selection available in the external Mul tiplier reference clock multiplier depends on the external reference clock word length. Keysight U8903B User’s Guide...
128 × sampling rate (bi-phase clock). When the synchronous clock divider is set to 128, the Divider synchronous clock is divided by 128 which is equal to the sampling rate set at the U8903B. – 1 –...
When connecting the U8903B to a PC's Bluetooth using A2DP profile where the U8903B NOTE acts as source, initiate the connection from the PC (as sink) instead of U8903B. The U8903B does not support the initialization of A2DP source profile connection from the U8903B when connecting to a PC.
Only available if Filter Class is set to Custom. Custom Press the Custom softkey to manually sets the Bluetooth device class. Refresh Device Press the Refresh Device softkey to refresh the friendly name of the selected Bluetooth device. Keysight U8903B User’s Guide...
Disconnects the audio media connection to the connected Bluetooth device via the A2DP profile. – Start Streaming Starts or resumes the audio media connection to the connected Bluetooth device. – Pause Streaming Stops or pauses the audio media connection to the connected Bluetooth device. Keysight U8903B User’s Guide...
Page 149
Emulates a button press on the connected remote headset device. Only applicable for HFP profile. – Send AT Command Sends the AT command from the U8903B’s Bluetooth card to the connected remote device. Press the HFP Operation softkey to select the HFP (Audio Gateway) profile operation settings. This is only available if the HFP profile is connected and the Common Settings >...
Audio Generator Functions Common settings Figure 3-63 Bluetooth Analyzer > Link Config > Common Settings menu page 1 Figure 3-64 Bluetooth Analyzer > Link Config > Common Settings menu page 2 Keysight U8903B User’s Guide...
Menu Description Press the Profile softkey to select the U8903B’s Bluetooth card profile. There are two sets of profiles available and the selected profile set will be accessible by the connected Bluetooth device. – A2DP SRC, HFP AG, AVRCP TRGT...
Page 152
Loopback available if the HFP profile is connected. Press the Link Status softkey to enable or disable the link status of the U8903B’s Bluetooth card. This is only available if Link Status the HFP profile or A2DP profile is connected.
– No Transmission No retransmission effort. – Power Saving Retransmit One or more retransmission effort. Optimized for power consumption. – Link Quality One or more retransmission effort. Optimized for link quality. – No Preference No retransmission effort preference. Keysight U8903B User’s Guide...
Bluetooth Analyzer > Link Config > AVRCP log view menu description Menu Description Clear Log Press the Clear Log softkey to clear the AVRCP operation log history. Refresh Log Press the Refresh Log softkey to refresh the AVRCP operation log history. Keysight U8903B User’s Guide...
Press the Word Length softkey to set the master clock word length value. The word length cannot be less than the audio resolution. Word Length – 8 to 32 Press the Resolution softkey to set the audio resolution. Resolution – 8 to 24 Keysight U8903B User’s Guide...
Page 158
Press the Vol tage softkey to select the output voltage or set your desired output voltage. – 1.2 Vpp – 1.5 Vpp – 1.8 Vpp Vol tage – 2.5 Vpp – 3 Vpp – 3.3 Vpp – Custom Keysight U8903B User’s Guide...
Save Save File menu page. Press the Load File softkey to load the bits from a file. Refer to “Recall” on page for more information on the Recall Load File menu page. Keysight U8903B User’s Guide...
Press the Emphasis softkey to select the emphasis type. – No pre-emphasis Emphasis – 50/15 μs – Reserved 1 – Reserved 2 Press the Channel Mode softkey to set the channel mode value. Channel Mode – 0 to 3 Keysight U8903B User’s Guide...
Press the Clk Accuracy softkey to select the clock accuracy type. – Level 1 Clk Accuracy – Level 2 – Level 3 – Reserved Press the Max Word Len softkey to select the maximum word length. Max Word Len – 24 bits – 20 bits Keysight U8903B User’s Guide...
Page 162
– 176.4 kHz – 192 kHz – Reserved 1 – Reserved 2 Press the CGMA-A softkey to select the CGMA-A type. – Copying Permitted CGMA-A – Condition Not Used – One Generation Copy – Copying Denied Keysight U8903B User’s Guide...
Press the Emphasis softkey to select the emphasis type. – Not indicated Emphasis – No pre-emphasis – 50/15 μs – CCITT J.17 Press the Freq Mode softkey to select the frequency mode. Freq Mode – Locked – Unlocked Keysight U8903B User’s Guide...
Page 166
– User defined – Reserved for Metadata – As in IEC60958-3 Press the Aux Bits softkey to select the auxiliary bits. – 20-bit not defined Aux Bits – 24-bit main audio – 20-bit single – Reserved Keysight U8903B User’s Guide...
Page 167
Press the Channel Origin softkey to set the channel origin value (4 alphanumeric digit). Channel Dest Press the Channel Dest softkey to set the channel destination value (4 alphanumeric digit). Press the Local Addr softkey to set the local address value. Local Addr – 0 to 2 –1 Keysight U8903B User’s Guide...
8 – 11 0110: Reserved for user application 0111: SCDSR (see byte 3 for ID) 1000: SCDSR (stereo left) 1001: SCDSR (stereo right) 1111: Multichannel (see byte 3 for ID) Single Channel Double Sample Rate (SCDSR) Keysight U8903B User’s Guide...
Page 169
32-bit binary number representing the sample count of the first sample of the channel 112 – 143 Local sample address code status block 144 – 175 Time of day code 32-bit binary number representing tiem source encoding in samples since midnight Keysight U8903B User’s Guide...
Page 170
Description 0: Data in byte range is reliable 176 – 183 Reliability flags 1: Data in byte range is unreliable 00000000: Not implemented 184 – 191 CRCC nnnnnnnn: Error check code for bits 0 - 183 Keysight U8903B User’s Guide...
Figure 3-73 Analog Generator > References menu page Table 3-45 Analog Generator > References menu description Menu Description Ref. Imp Press the Ref. Imp softkey to set the reference impedance for the unit conversion of dBm measurements. Keysight U8903B User’s Guide...
Press the Vol ts/FS softkey to set the volts/FS reference for the analog output voltage (in Volts) of the DAC used when Vol ts/FS converted to digital full scale (1 FS). This is useful in DAC tests for the conversion of digital units to analog units. Keysight U8903B User’s Guide...
Page 173
Keysight U8903B Performance Audio Analyzer User’s Guide Audio Analyzer Measurement Functions Audio Analyzer Measurement mode (Digital Analyzer) Functions Measurement Functions Frequency measurement AC voltage level measurement DC voltage level measurement THD+N ratio and THD+N level measurements SINAD measurement THD ratio and THD level measurements...
Page 174
Audio Analyzer Measurement Functions Wave File (Analog, Digital, and Bluetooth Analyzer) Statistics This chapter describes the U8903B performance audio analyzer measurement functions and configurations. Keysight U8903B User’s Guide...
Press on the CHANNEL/DATA panel to switch to channel 1 or channel 2 respectively, and start configuring the active channel. The U8903B can also be expanded with additional analog analyzer channels. Refer to “U8903B Options” on page for more information.
Press the Wave File softkey to configure the wave file settings. Refer to “Wave File (Analog, Digital, and Bluetooth Wave File Analyzer)” on page for more information. Statistics Press the Statistics softkey to configure the statistics settings. Refer to “Statistics” on page for more information. Figure 4-2 Digital Analyzer menu page (standard measurement mode) Keysight U8903B User’s Guide...
Press the Bits Analysis softkey to set the bits analysis configuration. Refer to “Bits Analysis (Digital Analyzer)” on Bits Analysis page for more information. Statistics Press the Statistics softkey to configure the statistics settings. Refer to “Statistics” on page for more information. Keysight U8903B User’s Guide...
Press the Wave File softkey to configure the wave file settings. Refer to “Wave File (Analog, Digital, and Bluetooth Wave File Analyzer)” on page for more information. Statistics Press the Statistics softkey to configure the statistics settings. Refer to “Statistics” on page for more information. Keysight U8903B User’s Guide...
Process delay mode allows you to view the process delay in seconds. When the digital analyzer analysis mode is set to Process Delay, you are not allowed to NOTE perform any other types of measurements. Figure 4-4 Digital Analyzer menu page (process delay measurement mode) Keysight U8903B User’s Guide...
This test will only be meaningful if the selected output and input connectors are the same type. The U8903B is able to analyze all waveform types generated by the digital generator In BERT mode except for the following: – Monotonicity –...
Digital Analyzer > Meas Mode > BERT menu description Menu Description Press the Unit softkey to select the BERT result in either hex or decimal. Unit – Hex – Dec Reset BERT Press the Reset BERT softkey to reset the BERT result. Keysight U8903B User’s Guide...
Audio Analyzer Measurement Functions Functions The U8903B supports up to four functions to be measured in one single query. The function configuration for analog, digital, and Bluetooth analyzer are displayed as shown in Figure 4-6, Figure 4-7, and Figure 4-8.
Press the Meas. Func softkey to display the list of available analyzer measurement function types. Select the desired measurement function from the drop-down list. Refer to “Measurement Functions” on page for the respective Meas. Func measurement functions settings. This setting is only available when the multichannel mode is set to Off. Keysight U8903B User’s Guide...
Frequency measurement Frequency is a common and basic measurement function which is expressed in hertz (Hz). The U8903B uses software algorithm to detect the period of a repetitive waveform, and the frequency is computed from the reciprocal of the period.
Press the Format softkey to select the format of the returned measurement reading. – Off Format Selecting Off will return the measurement reading without any formatting (raw value). – Delta Selecting Delta will return the measurement result by deducting the reference frequency from the measured frequency. Keysight U8903B User’s Guide...
Page 186
Press the Set resul t as ref. from softkey to store the measurement result from the selected channel as the reference Set resul t as ref. from frequency. Ref. Freq Press the Ref. Freq softkey to set the reference frequency. Keysight U8903B User’s Guide...
AC voltage level measurement is the most common measurement function of an AC signal. The U8903B provides two types of AC level detection consisting of rms and peak-to-peak. When the rms detection method is selected, the AC voltage measurement is expressed as an rms value.
Audio Analyzer Measurement Functions Figure 4-13 Digital Analyzer > Functions > AC Voltage menu page 1 Figure 4-14 Bluetooth Analyzer > Functions > AC Voltage menu page 1 Keysight U8903B User’s Guide...
Press the Set resul t as ref. from softkey to store the measurement result from the selected channel as the reference level. Ref. Level Press the Ref. Level softkey to set the reference level value. There is only one reference level for each channel. Keysight U8903B User’s Guide...
Page 190
Press the Cal SPL softkey to set the calibrator level value. This setting is only available when the unit is set to dBSPL. Press the Detector softkey to select the AC level detection type. Detector – RMS – Pk-Pk Keysight U8903B User’s Guide...
Bluetooth analyzer. The settings for DC voltage level measurement are displayed as shown in Figure 4-15, Figure 4-16, and Figure 4-17 for the analog, digital, and Bluetooth analyzer respectively. Figure 4-15 Analog Analyzer > Functions > DC Voltage menu page 1 Keysight U8903B User’s Guide...
Audio Analyzer Measurement Functions Figure 4-16 Digital Analyzer > Functions > DC Voltage menu page 1 Figure 4-17 Bluetooth Analyzer > Functions > DC Voltage menu page 1 Keysight U8903B User’s Guide...
Press the Set resul t as ref. from softkey to store the measurement result from the selected channel as the reference level. Ref. Level Press the Ref. Level softkey to set the reference level value. There is only one reference level for each channel. Keysight U8903B User’s Guide...
Bluetooth analyzer. The settings for THD+N ratio measurement are displayed as shown in Figure 4-18, Figure 4-19, and Figure 4-20 for the analog, digital, and Bluetooth analyzer respectively. Figure 4-18 Analog Analyzer > Functions > THD+N Ratio menu page 1 Keysight U8903B User’s Guide...
Selecting Auto will allow the U8903B to search for the fundamental frequency automatically. – Gen Lock Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency Freq Lock value of the respective generator channel. You can select the generator channel by setting the Src Channel in the Meas.
Figure 4-21, Figure 4-22, and Figure 4-23 for the analog, digital, and Bluetooth analyzer respectively. Figure 4-21 Analog Analyzer > Functions > THD+N Level menu page 1 Figure 4-22 Digital Analyzer > Functions > THD+N Level menu page 1 Keysight U8903B User’s Guide...
– V – ΔV – dBSPL – x Unit Digital and Bluetooth Analyzer – FFS – dBFS – %FS – V – dBu – dBV – dBr – x – LSB – Hex – Dec – dBSPL Keysight U8903B User’s Guide...
Page 199
Selecting Auto will allow the U8903B to search for the fundamental frequency automatically. – Gen Lock Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency Freq Lock value of the respective generator channel. You can select the generator channel by setting the Src Channel in the Meas.
Figure 4-24, Figure 4-25, and Figure 4-26 for the analog, digital, and Bluetooth analyzer respectively. Figure 4-24 Analog Analyzer > Functions > SINAD menu page 1 Figure 4-25 Digital Analyzer > Functions > SINAD menu page 1 Keysight U8903B User’s Guide...
Press the Set resul t as ref. from softkey to store the measurement result from the selected channel as the reference ratio. Ref. Ratio Press the Ref. Ratio softkey to set the reference ratio value. This is only available for the digital analyzer. Precision Press the Precision softkey to enable or disable the precision mode. Keysight U8903B User’s Guide...
Page 202
Selecting Auto will allow the U8903B to search for the fundamental frequency automatically. – Gen Lock Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency Freq Lock value of the respective generator channel. You can select the generator channel by setting the Src Channel in the Meas.
Bluetooth analyzer. The settings for THD ratio measurement are displayed as shown in Figure 4-27, Figure 4-28, and Figure 4-29 for the analog, digital, and Bluetooth analyzer respectively. Figure 4-27 Analog Analyzer > Functions > THD Ratio menu page 1 Keysight U8903B User’s Guide...
Bluetooth Analyzer > Functions > THD Ratio menu page 1 Table 4-12 Analog/Digital/Bluetooth Analyzer > Functions > THD Ratio menu description Menu Description Press the Unit softkey to select the measurement unit. – dB Unit – ΔdB – % – x Keysight U8903B User’s Guide...
Page 205
Freq Lock Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency value of the respective generator channel. You can select the generator channel by setting the Src Channel in the Meas.
Figure 4-30, Figure 4-31, and Figure 4-32 for the analog, digital, and Bluetooth analyzer respectively. Figure 4-30 Analog Analyzer > Functions > THD level menu page 1 Figure 4-31 Digital Analyzer > Functions > THD level menu page 1 Keysight U8903B User’s Guide...
– V – ΔV – dBSPL – x Unit Digital and Bluetooth Analyzer – FFS – dBFS – %FS – V – dBu – dBV – dBr – x – LSB – Hex – Dec – dBSPL Keysight U8903B User’s Guide...
Page 208
Press the Even Harmonic softkey to select the even harmonics. – ALL – 2 Even Harmonic – 4 – 6 – 8 Press the Odd Harmonic softkey to select the odd harmonics. – ALL – 3 Odd Harmonic – 5 – 7 – 9 Keysight U8903B User’s Guide...
Page 209
Freq Lock Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency value of the respective generator channel. You can select the generator channel by setting the Src Channel in the Meas.
Bluetooth analyzer. The settings for SMPTE IMD measurement are displayed as shown in Figure 4-33, Figure 4-34, and Figure 4-35 for the analog, digital, and Bluetooth analyzer respectively. Figure 4-33 Analog Analyzer > Functions > SMPTE IMD menu page 1 Keysight U8903B User’s Guide...
– Gen Lock Selecting Gen Lock will allow the U8903B to search for the upper and lower frequencies of the input signal based on the upper and lower frequencies of the respective generator channel. You can select the generator channel by setting the Freq Lock Src Channel in the Meas.
Figure 4-37, and Figure 4-38 for the analog, digital, and Bluetooth analyzer respectively. Figure 4-36 Analog Analyzer > Functions > DFD 60268 2nd menu page 1 Figure 4-37 Digital Analyzer > Functions > DFD 60268 2nd menu page 1 Keysight U8903B User’s Guide...
Set resul t as ref. from Press the Set resul t as ref. from softkey to store the measurement result from the selected channel as the reference ratio. Ref. Ratio Press the Ref. Ratio softkey to set the reference ratio. Keysight U8903B User’s Guide...
The SNR measurement is accomplished by an internal routine that alternatively switches the U8903B generator output on and off. When the U8903B output is in the off state, the U8903B output will be terminated at the DUT input.
Press the Set resul t as ref. from softkey to store the measurement result from the selected channel as the reference ratio. Ref. Ratio Press the Ref. Ratio softkey to set the reference ratio. SNR Delay Press the SNR Delay softkey to set the SNR delay. Keysight U8903B User’s Guide...
The settings for SNR (fast mode) measurement are displayed as shown in Figure 4-40 and Figure 4-41 for the analog and Bluetooth respectively. Figure 4-40 Analog Analyzer > Functions > SNR (Fast) menu page 1 Keysight U8903B User’s Guide...
Set resul t as ref. from Press the Set resul t as ref. from softkey to store the measurement result from the selected channel as the reference ratio. Ref. Ratio Press the Ref. Ratio softkey to set the reference ratio. Keysight U8903B User’s Guide...
Page 219
Selecting Auto will allow the U8903B to search for the fundamental frequency automatically. – Gen Lock Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency Freq Lock value of the respective generator channel. You can select the generator channel by setting the Src Channel in the Meas.
DAC in the FFT spectrum. J-test measurement function mode is applicable for the analog analyzer only. The J-test measurement is displayed as shown in Figure 4-42. Figure 4-42 Analog Analyzer > Functions > J-Test menu page Keysight U8903B User’s Guide...
Menu Description Press the Unit softkey to select the measurement unit. – FFS – dBFS – %FS – V – dBu Unit – dBV – dBr – x – LSB – Hex – Dec – dBSPL Keysight U8903B User’s Guide...
Page 222
Press the Vol t/FS softkey to set the volts/FS reference value. Cal SPL Press the Cal SPL softkey to set the calibrator level value. This setting is only available when the unit is set to dBSPL. Keysight U8903B User’s Guide...
Menu Description Press the Unit softkey to select the measurement unit. – FFS – dBFS – %FS – V – dBu Unit – dBV – dBr – x – LSB – Hex – Dec – dBSPL Keysight U8903B User’s Guide...
Page 224
Press the Vol t/FS softkey to set the volts/FS reference value. Cal SPL Press the Cal SPL softkey to set the calibrator level value. This setting is only available when the unit is set to dBSPL. Keysight U8903B User’s Guide...
1 to measure the input, and use the analog analyzer channel 2 to measure the output. The U8903B always uses channel 1 of the analog generator as the reference channel, while the selected analog analyzer channel becomes the reference channel for interchannel phase measurements.
Page 226
Audio Analyzer Measurement Functions Figure 4-45 Analog Analyzer > Functions > Phase menu page Figure 4-46 Digital Analyzer > Functions > Phase menu page Keysight U8903B User’s Guide...
Freq Lock Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency value of the respective generator channel. You can select the generator channel by setting the Src Channel in the Meas.
The settings for crosstalk measurement are displayed as shown in Figure 4-48, Figure 4-49, and Figure 4-50 for the analog, digital, and Bluetooth analyzer respectively. Figure 4-48 Analog Analyzer > Functions > X-Talk menu page 1 Keysight U8903B User’s Guide...
Analog/Digital/Bluetooth Analyzer > Functions > X-Talk menu description Menu Description Ref. Channel Press the Ref. Channel softkey to set the driven channel. Press the Unit softkey to select the measurement unit. – dB Unit – ΔdB – % – x Keysight U8903B User’s Guide...
Page 230
Freq Lock Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency value of the respective generator channel. You can select the generator channel by setting the Src Channel in the Meas.
– 15 kHz – 20 kHz – 22 kHz – 30 kHz – 40 kHz – 50 kHz – 80 kHz – Custom (Refer to “Append ix 18: User-defined Filter File Format” on page for more information.) Keysight U8903B User’s Guide...
Page 232
Press the Deemphasis softkey to select the de-emphasis value. – None Deemphasis – 50 μs – 75 μs – Custom Press the Notch Fil ter softkey to configure the notch filter settings. Refer to “Notch Filter” on page for more Notch Fil ter information. Keysight U8903B User’s Guide...
Press the Center Freq softkey to set the frequency of the component to be removed from the input signal. Band wid th Press the Band wid th softkey to set the band width of the signal component to be removed. Keysight U8903B User’s Guide...
Press the HPF softkey to select the high-pass filter. – None – 20 Hz – 100 Hz – 400 Hz – Custom (Refer to “Append ix 18: User-defined Filter File Format” on page for more information.) Keysight U8903B User’s Guide...
Page 235
This is only available for digital analyzer. Press the Sample Rate softkey to select the sample rate. – 32 kHz – 44.1 kHz Sample Rate – 48 kHz – 88.2 kHz – 96 kHz – 176.4 kHz – 192 kHz Keysight U8903B User’s Guide...
Press the Weighting softkey to select the weighting filter. – None – A – CCIR 1k Weighting – CCIR 2k – CCITT – C-Message – Custom (Refer to “Append ix 18: User-defined Filter File Format” on page for more information.) Keysight U8903B User’s Guide...
Press the Center Freq softkey to set the frequency of the component to be removed from the input signal. Band wid th Press the Band wid th softkey to set the band width of the signal component to be removed. Keysight U8903B User’s Guide...
– For the frequencies searching algorithm when the frequency lock is set to Gen Lock. Press the Trigger softkey to configure the trigger settings. Refer to “Trigger (Analog, Digital, and Bluetooth Trigger Analyzer)” on page for more information. Keysight U8903B User’s Guide...
Page 239
Audio Analyzer Measurement Functions For details on audio measurement ranging, please refer to “Append ix 29: Aud io NOTE Measurement Ranging” on page to set the input voltage range. Keysight U8903B User’s Guide...
The higher the sample size, the higher the accuracy of the measurement result. – 2048 – 4096 – 8192 – 16384 – 32768 Sample Size – 65536 – 131072 – 262144 – 524288 – 1M – 2M Keysight U8903B User’s Guide...
Page 241
Press the Mov Average softkey to set the average points. Higher number of average points should be used when the analyzed data is noisy. Mov Average – 1 to 20 points Press the Trigger softkey to configure the trigger settings. Refer to “Trigger (Analog, Digital, and Bluetooth Trigger Analyzer)” on page for more information. Keysight U8903B User’s Guide...
– 131072 – 262144 – 524288 Press the Mov Average softkey to set the average points. Higher number of average points should be used when the analyzed data is noisy. Mov Average – 1 to 20 points Keysight U8903B User’s Guide...
Page 243
– For the frequencies searching algorithm when the frequency lock is set to Gen Lock. Press the Trigger softkey to configure the trigger settings. Refer to “Trigger (Analog, Digital, and Bluetooth Trigger Analyzer)” on page for more information. Keysight U8903B User’s Guide...
The trigger menu page for analog, digital, and Bluetooth analyzer is displayed as shown in Figure 4-59, Figure 4-60, and Figure 4-61. Figure 4-59 Analog Analyzer > Meas Config > Trigger menu page Figure 4-60 Digital Analyzer > Meas Config > Trigger menu page Keysight U8903B User’s Guide...
When the Trigger Interface is Analog, the Trigger Level is in Vpeak. Trigger Level When the Trigger Interface is Digital or Bluetooth, the Trigger Level is in Hex. Press the Trigger Level softkey to set the trigger level. Keysight U8903B User’s Guide...
Press the Impedance softkey to select the input impedance value. This setting is only available when Connector is set to UnBal or Bal. – 100 kΩ (for UnBal) Impedance – 200 kΩ (for Bal) – 600 Ω – 300 Ω Auto Range Press the Auto Range softkey to enable or disable auto range. Keysight U8903B User’s Guide...
Page 247
– 1 V – 320 mV Press the Band wid th softkey to select the input band width value. This setting is only available with Option N3431A. Refer to “U8903B Options” on page for more information. Band wid th – 90 kHz (This is the default setting with sampling rate of 192 kHz) –...
DSI connector type routes the digital signal from the 25-pin female D-SUB connector at the rear panel to the digital analyzer. Press the Impedance softkey to select the input impedance. Unbalanced connector type – 75 Ω Impedance – HiZ Balanced connector type – 110 Ω – HiZ Keysight U8903B User’s Guide...
Page 249
The audio signal frequency will be translated according to the reference sample rate value set in the Ref SR. This is only available when frequency scaling is Custom. Ref SR Press the Ref SR softkey to set the reference sample rate value. Keysight U8903B User’s Guide...
Audio Analyzer Measurement Functions Link Configuration (Bluetooth Analyzer) The link configuration for the Bluetooth analyzer is identical to the link configuration for the Bluetooth generator. Please refer to “Link Configuration (Bluetooth Generator)” on page for more information. Keysight U8903B User’s Guide...
Press the Data Shift Cnt softkey to set the number for the data bits to be shifted in relative to the frame clock. This is only available when the format is Left or Right. Press the Data Shift Dir softkey to select the data shift direction. Data Shift Dir – Left – Right Keysight U8903B User’s Guide...
Page 252
– Rising – Falling Press the Vol tage softkey to select the input voltage value. – 1.2 Vpp – 1.5 Vpp – 1.8 Vpp Vol tage – 2.5 Vpp – 3 Vpp – 3.3 Vpp – Custom Keysight U8903B User’s Guide...
Save in Hex Save menu page. Press the Save in Xml softkey to save the status bits to an XML file. Refer to “Save” on page for more information on Save in Xml the Save menu page. Keysight U8903B User’s Guide...
"1" indicates that the data is 1 at the particular bit, and "0" indicates that the data is 0 at the particular bit. – Active Active bits indicate bits that have changed state during the measurement period. "1" indicates a bit has changed state, and "0" indicates no change. Keysight U8903B User’s Guide...
Wave File (Analog, Digital, and Bluetooth Analyzer) The U8903B allows you to record and save the input measurement signal into a wave file. The recorded wave file can be used as a test signal. The wave file settings for analog, digital, and Bluetooth analyzer is displayed as shown in Figure 4-67, Figure 4-68, and Figure 4-69 respectively.
Press the Duration softkey to set the recording duration of the wave file. Press the Record softkey to start recording and save the measurement into a wave file. Refer to “Save” on page Record more information on the Save menu page. Keysight U8903B User’s Guide...
The statistics settings for analog, digital, and Bluetooth analyzer is displayed as shown in Figure 4-70, Figure 4-71, and Figure 4-72 respectively. Figure 4-70 Analog Analyzer > Statistics menu page Figure 4-71 Digital Analyzer > Statistics menu page Keysight U8903B User’s Guide...
No. of Read ing Press the No. of Read ing softkey to set the number of readings used for the statistics calculation. Press the respective softkeys to select the statistics calculation type. The U8903B allows you to display three types of statistics data.
Graph Analysis Graph Analysis – The U8903B graph analysis mode is only available in the Standard View mode. Refer to NOTE “Menu Key” on page for more information. – When the graph analysis mode is activated, all filters will be turned off.
Press the Running Mode softkey to select the graph analysis running mode. Running Mode – Continuous – Single Press the Measurement softkey to configure the graph analysis measurement settings. Refer to “Measurement Measurement Settings” on page for more information. Keysight U8903B User’s Guide...
If the sample size of an analyzer card channel is set to 52428, the rest of the analyzer card channels sample size will be set to 262144 (if the initial sample size is more than 262144). Keysight U8903B User’s Guide...
Page 263
The Min hold will compare the current data with the previous data and retains whichever value that is lower. – Max The Max hold will compare the current data with the previous data and retain whichever value that is higher. Keysight U8903B User’s Guide...
Press the Bottom softkey to set the bottom axis limit value. If the value entered is more than the top value, the top value Bottom will automatically be set to a value of (Bottom + 10). Keysight U8903B User’s Guide...
Page 265
– Linear – Log Press the Y-Scale softkey to select the Y-axis scale type. The bottom and top values must be more than 0 in order to use the log scale. Y-Scale – Linear – Log Keysight U8903B User’s Guide...
Press the Unit softkey to select the trace unit type. – V – dBV Unit – dBu – W – dBm – dBSPL Press the Persistence softkey to configure the persistence settings. Refer to “Persistence” on page for more Persistence information. Keysight U8903B User’s Guide...
Page 267
Axis – Primary – Secondary Press the Color softkey to select the color of the active trace. – Yellow – Cyan – White Color – Pink – Green – Orange – Red – Purple Keysight U8903B User’s Guide...
Clear Memory Press the Clear Memory softkey to clear the trace from the memory buffer. [a] Files saved in the internal volatile memory will be erased after power cycle or reset (refer to “Preset” on page 78). Keysight U8903B User’s Guide...
Press the Function softkey to select the math function to be applied on the active trace. – None – f(Source) + x Function – f(Source) - x – f(Source) * x – f(Source) / x Press the X softkey to set the x value in the selected math function. Keysight U8903B User’s Guide...
For example, a value of 5 indicates that up to a total of 5 previous sets of trace data will be displayed in the graph Persist Count before they are removed. This setting is only available when persistence is enabled. Keysight U8903B User’s Guide...
Graph view is the default display option. In the graph view, the frequency, phase, and time domain analysis are performed. The graph view is displayed as shown in Figure 5-8. Figure 5-8 Graph Analysis > Display Option > Graph menu page Keysight U8903B User’s Guide...
The delta marker readings are also listed in the table. The marker table view is displayed as shown in Figure 5-10. Figure 5-10 Graph Analysis > Display Option > Marker Table menu page Keysight U8903B User’s Guide...
You can also observe the THD information for the individual channels. The harmonics display is displayed as shown in Figure 5-12. Figure 5-12 Graph Analysis > Display Option > Harmonics menu page Keysight U8903B User’s Guide...
Press the Display softkey to select the type of harmonic analysis data to be displayed. Display – Harmonics – THD Press the THD unit softkey to select the unit for THD measurements. THD unit – dB – % Keysight U8903B User’s Guide...
In the signal analysis view, you can view the analyzer and generator channel information. The signal analysis display is displayed as shown in Figure 5-14. Figure 5-14 Graph Analysis > Display Option > Signal Analysis menu page Keysight U8903B User’s Guide...
Graph Analysis Measurement Settings The U8903B allows you to observe up to two measurements running simultaneously in the graph analysis mode. The measurements will be displayed at the bottom of the graph. The measurement settings for graph analysis is displayed as shown in Figure 5-15.
Press the Function No softkey to select the function to be displayed based on the function number (1 to 4). The function number corresponds to the respective measurement functions in the analyzer. The allowable measurement functions are as follows. – Frequency Function No – DC voltage – THD ratio – THD level Keysight U8903B User’s Guide...
Page 278
Graph Analysis THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight U8903B User’s Guide...
Page 279
Keysight U8903B Performance Audio Analyzer User’s Guide Sweep Function Sweep Parameter Group Delay Points settings Sweep channels Plot View Axis settings Plot settings Edit Points This chapter describes the U8903B sweep parameter and the configurations.
NOTE on page for more information. The U8903B sweep mode allows you to perform sweeps and to display the results in a plot view. Press on the FUNCTION panel to access the sweep mode. The Sweep Parameter menu page is the default page and is displayed as shown in Figure 6-1.
This setting is only available when the legacy sweep is disabled. Press the Plot View softkey to display the Plot View menu page. Plot View Refer to “Plot View” on page for more information. Keysight U8903B User’s Guide...
Group Delay measurement function mode is only applicable for digital interface. Figure 6-2 Sweep Parameter > App. Type > Group Delay menu page Keysight U8903B User’s Guide...
Press the Stop softkey to set the sweep stop value. This setting is only available when the sweep spacing is Log or Linear. Press the Step softkey to set the sweep step value. Adjusting the sweep step value will change the sweep points value. Step This setting is only available when the sweep spacing is Log or Linear. Keysight U8903B User’s Guide...
Sweep Parameter > Channels menu description Menu Description Source Press the Source softkey to select the source channel(s). Press the Measure softkey to select the measure channel(s). The available selection depends on the number of installed Measure analog analyzer cards. Keysight U8903B User’s Guide...
Edit Points Refer to “Ed it Points” on page for more information. Press the Save Pts softkey to save the sweep points to a file. Save Pts Refer to “Save” on page for the Save menu page. Keysight U8903B User’s Guide...
Press the Bottom softkey to set the bottom axis value. Press the X-Scale softkey to set the X-axis scale type. X-Scale – Linear – Log Press the Y-Scale softkey to set the Y-axis scale type. Y-Scale – Linear – Log Keysight U8903B User’s Guide...
Press the X Data softkey to select the X-axis trace data source. – S: <sweep parameter> – F1: <measurement function 1> X Data – F2: <measurement function 2> – F3: <measurement function 3> – F4: <measurement function 4> Keysight U8903B User’s Guide...
Page 288
– Average Hold Type Each point is averaged with the corresponding point from the previous result. – Max Always keep the maximum value of each point. – Min Always plot the minimum value of each point. Keysight U8903B User’s Guide...
Press the Save Points softkey to save the sweep points to a file. Save Points Refer to “Save” on page for the Save menu page. If the sweep points are changed, the sweep spacing mode will be set to Custom. Refer to NOTE Table 6-2 for more information. Keysight U8903B User’s Guide...
Page 290
Sweep Function THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight U8903B User’s Guide...
Test Sequence Application Voice quality Measurement Results Report Properties This chapter describes the various configurations for the U8903B test sequence application. Keysight U8903B User’s Guide...
Test Sequence Application Test Sequence Application The U8903B test sequence application (Test Seq App) mode allows you to create a series of automated measurements on a DUT. You can customize the order in which the measurements are performed, edit the pass or fail limits, add sub steps, and create a detailed report of the measurements.
This setting is only available when the On/Off key is pressed with the test or measurement tab selected. Press the Start Selected Measurement softkey to start the test for the selected measurement. Start Selected Measurement This setting is only available when the On/Off key is pressed with the measurement tab selected. Keysight U8903B User’s Guide...
Press the up or down arrow key to select the test sequence, IO configuration, or measurements tabs. The respective menu page will be displayed. Refer to “Test Sequence” on page 297, “IO Configuration” on page 298, and “Measurements” on page for more information. Keysight U8903B User’s Guide...
Press the Fail Message softkey to edit the fail message. If one or more measurements in the test sequence fails, the fail Fail Message message is displayed. The default message is “Failed” This setting is only available when Pass/Fail Msg is enabled. Keysight U8903B User’s Guide...
Press the Ed it softkey to move or copy the selected test sequence, or paste a copied test sequence after the selected test Ed it sequence. Properties Press the Properties softkey to set the test sequence name. Keysight U8903B User’s Guide...
Press the Settings softkey to configure the IO settings. Refer to “Settings” on page for more information. Press the Properties softkey to configure the IO configuration properties. Refer to “Properties” on page for more Properties information. Keysight U8903B User’s Guide...
DUT. Press the Reference softkey to set the output references for generator. Reference – Impedance Set the reference impedance for the unit conversion of dBm measurements. Keysight U8903B User’s Guide...
Press the Link Config softkey to configure the Bluetooth link settings. Refer to “Link Configuration (Bluetooth Link Config Generator)” on page 146. Press the Reference softkey to set the output references for generator. References – Volt/FS Set the volts/FS reference value. Keysight U8903B User’s Guide...
Press the Impedance softkey to select the input impedance value. This setting is only available when Connector is set to UnBal or Bal. – 100 kΩ (for Unbalanced) Impedance – 200 kΩ (for Balanced) – 600 Ω – 300 Ω Keysight U8903B User’s Guide...
Page 303
(This is the default setting with sampling rate of 192 kHz) Band wid th – 1.5 MHz – This setting is only available with Option N3431A. Refer to “U8903B Options” on page for more information. Press the Reference softkey to set the input references for analyzer.
Page 304
Device, the profile used will be fixed to the connected profile(s). – A2DP Profiles – HFP – AVRCP Press the Link Config softkey to configure the Bluetooth link settings. Refer to “Link Configuration (Bluetooth Link Config Generator)” on page 146. Keysight U8903B User’s Guide...
Page 305
Set the reference ratio value. – Frequency References Set the reference frequency value. – Sound Lvl Set the sound level value. – Calibrator Lvl Set the calibration level value value. – Volt/FS Set the volts/FS reference value. Keysight U8903B User’s Guide...
Press the Failure softkey to edit the failure message. If one or more measurements in the test sequence fails, the failure Failure message is displayed. The default message is “Failed”. This setting is only available when Pass/Fail Msg is enabled. Keysight U8903B User’s Guide...
Settings – Prompt Refer to “Prompt sub-step settings” on page for more information. – Send SCPI Refer to “Send SCPI sub-step settings” on page for more information. Properties Press the Properties softkey to rename the sub-step. Keysight U8903B User’s Guide...
Press the Timeout softkey to set the prompt timeout value in seconds. This setting is only available if Timeout is selected in the dialog settings. Timeout Minimum: 1 s Maximum: 3600 s (1 hour) Keysight U8903B User’s Guide...
An optional delay time can be set to wait for completion. The next sub-step in the test sequence will not be executed until the delay time has passed. Keysight U8903B User’s Guide...
Test Sequence Application Measurements The U8903B allows up to 20 results in a single measurement. You can move a selected measurement within the same test sequence. An example of the measurement menu page is displayed as shown in Figure 7-15.
Page 311
– SMPTE level sweep – DFD level sweep – DC level sweep – Receiver sensitivity – Measurement recorder – Voice quality Press the Properties softkey to configure the measurement properties. Refer to “Properties” on page for more Properties information. Keysight U8903B User’s Guide...
Display a prompt window with Abort, Retry, and Ignore commands. Abort will immediately stop the test sequence. Retry will rerun the selected measurements. Ignore will flag the measurement as failed and continue with the test sequence. – Continue Seq. Flag the measurement as failed and continue with the test sequence. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The AC level signal generation settings menu page is displayed as shown in Figure 7-17. Figure 7-17 TSA > Project > Test > AC Level > Settings > Signal Generation menu page Keysight U8903B User’s Guide...
Set the phase value. This setting is only available when channel 2 is selected. Signal analysis The AC level signal analysis settings menu page is displayed as shown in Figure 7-18. Figure 7-18 TSA > Project > Test > AC Level > Settings > Signal Analysis menu page Keysight U8903B User’s Guide...
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The frequency signal generation settings menu page is displayed as shown in Figure 7-19. Figure 7-19 Frequency > Signal Generation settings menu page Keysight U8903B User’s Guide...
Waveform Config Set the frequency value. – Amplitude Set the amplitude value. – DC Offset Set the DC offset value. – Phase->1st Ch Set the phase value. This setting is only available when channel 2 is selected. Keysight U8903B User’s Guide...
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The phase signal generation settings menu page is displayed as shown in Figure 7-21. Figure 7-21 Phase > Signal Generation settings menu page Keysight U8903B User’s Guide...
Waveform Config Set the frequency value. – Amplitude Set the amplitude value. – DC Offset Set the DC offset value. – Phase->1st Ch Set the phase value. This setting is only available when channel 2 is selected. Keysight U8903B User’s Guide...
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to Input Range channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The SNR signal generation settings menu page is displayed as shown in Figure 7-23. Figure 7-23 SNR > Signal Generation settings menu page Keysight U8903B User’s Guide...
Waveform Config Set the frequency value. – Amplitude Set the amplitude value. – DC Offset Set the DC offset value. – Phase->1st Ch Set the phase value. This setting is only available when channel 2 is selected. Keysight U8903B User’s Guide...
Freq Lock Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency value of the respective generator channel. This Gen Lock method is only useful if you use the internal audio generator.
Page 326
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The THD+N signal generation settings menu page is displayed as shown in Figure 7-25. Figure 7-25 THD+N > Signal Generation settings menu page Keysight U8903B User’s Guide...
Set the phase value. This setting is only available when channel 2 is selected. Signal analysis The THD+N signal analysis settings menu page is displayed as shown in Figure 7-26. Figure 7-26 THD+N > Signal Analysis settings menu page Keysight U8903B User’s Guide...
Freq Lock Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency value of the respective generator channel. This Gen Lock method is only useful if you use the internal audio generator.
Page 330
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The DC level signal generation settings menu page is displayed as shown in Figure 7-27. Figure 7-27 DC Level > Signal Generation settings menu page Keysight U8903B User’s Guide...
Disable the Track 1st Ch to set the individual channels. Waveform Config – Frequency Set the frequency value. – Amplitude Set the amplitude value. – DC Offset Set the DC offset value. Keysight U8903B User’s Guide...
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The crosstalk signal generation settings menu page is displayed as shown in Figure 7-29. Figure 7-29 Crosstalk > Signal Generation settings menu page Keysight U8903B User’s Guide...
Set the amplitude value. – DC Offset Set the DC offset value. Signal analysis The crosstalk signal analysis settings menu page is displayed as shown in Figure 7-30. Figure 7-30 Crosstalk > Signal Analysis settings menu page Keysight U8903B User’s Guide...
Freq Lock Selecting Gen Lock will allow the U8903B to use the driven channel frequency value. You can select the generator channel by setting the Driven Ch. This Gen Lock method is only useful if you use the internal audio generator.
The signal generation is disabled if None is selected for the output configuration channel. NOTE The SMPTE IMD signal generation settings menu page is displayed as shown in Figure 7-31. Figure 7-31 SMPTE IMD > Signal Generation settings menu page Keysight U8903B User’s Guide...
Set the amplitude value. – DC Offset Set the DC offset value. Signal analysis The SMPTE IMD signal analysis settings menu page is displayed as shown in Figure 7-32. Figure 7-32 SMPTE IMD > Signal Analysis settings menu page Keysight U8903B User’s Guide...
– Gen Lock Selecting Gen Lock will allow the U8903B to search for the upper and lower frequencies of the input signal based on the upper and lower frequencies of the respective generator channel. This Gen Lock method is only useful if you use the Freq Lock internal audio generator.
The signal generation is disabled if None is selected for the output configuration channel. NOTE The DFD IMD signal generation settings menu page is displayed as shown in Figure 7-33. Figure 7-33 DFD IMD > Signal Generation settings menu page Keysight U8903B User’s Guide...
Set the amplitude value. – DC Offset Set the DC offset value. Signal analysis The DFD IMD signal analysis settings menu page is displayed as shown in Figure 7-34. Figure 7-34 DFD IMD > Signal Analysis settings menu page Keysight U8903B User’s Guide...
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The multitone analyzer signal generation settings menu page is displayed as shown in Figure 7-35. Figure 7-35 Multitone Analyzer > Signal Generation settings menu page Keysight U8903B User’s Guide...
Set thephase of all tones to random phase in degrees. – Edit Tones Edit the individual tones frequency, amplitude, and phase. – Active Channel The active channel for the absolute amplitude for each tone to be displayed in a table. Keysight U8903B User’s Guide...
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to Input Range channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The stepped frequency sweep signal generation settings menu page is displayed as shown in Figure 7-37. Figure 7-37 Stepped Frequency Sweep > Signal Generation settings menu page Keysight U8903B User’s Guide...
Disable the Waveform Config Track 1st Ch to set the individual channels. – Amplitude Set the amplitude value. – DC Offset Set the DC offset value. Keysight U8903B User’s Guide...
The result displayed is the exact value after substracted from the reference channel. Press the Harmonics softkey to select the harmonics count to be used in the THD ratio and THD level results. Harmonics – All – 2 to 9 Keysight U8903B User’s Guide...
Page 350
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The SMPTE frequency sweep signal generation settings menu page is displayed as shown in Figure 7-39. Figure 7-39 SMPTE Frequency Sweep > Signal Generation settings menu page Keysight U8903B User’s Guide...
Set the upper frequency value. This setting is only available if Swept is set to Lower Freq. – Lower Freq Set the lower frequency value. This setting is only available if Swept is set to Upper Freq. – Amplitude Set the amplitude value. – DC Offset Set the DC offset value. Keysight U8903B User’s Guide...
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The DFD frequency sweep signal generation settings menu page is displayed as shown in Figure 7-41. Figure 7-41 DFD Frequency Sweep > Signal Generation settings menu page Keysight U8903B User’s Guide...
Set the upper frequency value. This setting is only available if Swept is set to Diff Freq. – Diff Freq Set the difference frequency value. This setting is only available if Swept is set to Upper Freq. – Amplitude Set the amplitude value. – DC Offset Set the DC offset value. Keysight U8903B User’s Guide...
Press the Sample Size softkey to select the number of samples to be acquired for the measurement. – 2048 – 4096 – 8192 – 16384 – 32768 Sample Size – 65536 – 131072 – 262144 – 524288 – 1M – 2M Keysight U8903B User’s Guide...
Page 357
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
Press the GPIB Address softkey to select the GPIB address. GPIB Address – 0 to 30 Press the Init Instrument softkey to enable or disable the U8903B to send SCPI commands to initialize the connected Init Instrument signal generator. Press the Init SCPI softkey to set the SCPI commands that initialize the connected instrument. You can set the SCPI command directly or load from a file.
Page 359
Set the step size for linear spacing. This setting is only available when spacing is set to Linear. – Dwell Time Set the delay between the sweep points. – Edit Points Edit the individual points value, insert or remove points, load points, and save the points. Keysight U8903B User’s Guide...
Page 361
Enable or disable the custom notch. – Center Freq Set the center frequency value. This is only applicable when custom notch is enabled. – Band width Set the bandwidth value. This is only applicable when custom notch is enabled. Keysight U8903B User’s Guide...
Page 362
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The stepped level sweep signal generation settings menu page is displayed as shown in Figure 7-45. Figure 7-45 Stepped Level Sweep > Signal Generation settings menu page Keysight U8903B User’s Guide...
Disable the Waveform Config Track 1st Ch to set the individual channels. – Frequency Set the frequency value. – DC Offset Set the DC offset value. Keysight U8903B User’s Guide...
Freq Lock Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency value of the respective generator channel. This Gen Lock method is only useful if you use the internal audio generator.
Page 367
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The SMPTE level sweep signal generation settings menu page is displayed as shown in Figure 7-47. Figure 7-47 SMPTE Level Sweep > Signal Generation settings menu page Keysight U8903B User’s Guide...
Disable the Track 1st Ch to set the individual channels. Waveform Config – Upper Freq Set the upper frequency value. – Lower Freq Set the lower frequency value. – DC Offset Set the DC offset value. Keysight U8903B User’s Guide...
– Gen Lock Selecting Gen Lock will allow the U8903B to search for the upper and lower frequencies of the input signal based on the upper and lower frequencies of the respective generator channel. This Gen Lock method is only useful if you use the Freq Lock internal audio generator.
Page 371
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The DFD level sweep signal generation settings menu page is displayed as shown in Figure 7-49. Figure 7-49 DFD Level Sweep > Signal Generation settings menu page Keysight U8903B User’s Guide...
Disable the Track 1st Ch to set the individual channels. Waveform Config – Upper Freq Set the upper frequency value. – Diff Freq Set the difference frequency value. – DC Offset Set the DC offset value. Keysight U8903B User’s Guide...
Press the Sample Size softkey to select the number of samples to be acquired for the measurement. – 2048 – 4096 – 8192 – 16384 – 32768 Sample Size – 65536 – 131072 – 262144 – 524288 – 1M – 2M Keysight U8903B User’s Guide...
Page 375
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
The signal generation is disabled if None is selected for the output configuration channel. NOTE The DC level sweep signal generation settings menu page is displayed as shown in Figure 7-51. Figure 7-51 DC Level Sweep > Signal Generation settings menu page Keysight U8903B User’s Guide...
Edit the individual points value, insert or remove points, load points, and save the points. Signal analysis The DC level sweep signal analysis settings menu page is displayed as shown in Figure 7-52. Figure 7-52 DC Level Sweep > Signal Analysis settings menu page Keysight U8903B User’s Guide...
Page 379
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
U8903B controls the signal generator through Keysight 82357B USB/GPIB interface. A receiver such as a two-way radio is connected to the U8903B either by a direct connection or through an acoustic coupler. Figure 7-53...
Dwell Time Press the Dwell Time softkey to set the delay between each measured SINAD in seconds. Press the Init Instrument softkey to enable or disable the U8903B to send SCPI commands to initialize the connected Init Instrument signal generator.
Press the Target SINAD softkey to set the target SINAD value measured from the Meas Channel. Min 1st Value Press the Min 1st Value softkey to set the minimum first value. Press the Stop Cond ition softkey to select the stop condition. Stop Cond ition – Until End – On Target Keysight U8903B User’s Guide...
Page 383
Description Press the Freq Lock softkey to select the searching method for the fundamental frequency. – Auto Selecting Auto will allow the U8903B to search for the fundamental frequency automatically. Freq Lock – Custom If you have a known input signal, you can define the fundamental frequency value by setting the searching method to Custom and setting the frequency value in Fund Freq.
Page 384
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
Press the GPIB Address softkey to select the GPIB address. GPIB Address – 0 to 30 Press the Init Instrument softkey to enable or disable the U8903B to send SCPI commands to initialize the connected Init Instrument signal generator. Press the Init SCPI softkey to set the SCPI commands that initialize the connected instrument. You can set the SCPI command directly or load from a file.
Page 386
Set the step size for linear spacing. This setting is only available when spacing is set to Linear. – Dwell Time Set the delay between the sweep points. – Edit Points Edit the individual points value, insert or remove points, load points, and save the points. Keysight U8903B User’s Guide...
Freq Lock Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency value of the respective generator channel. This Gen Lock method is only useful if you use the internal audio generator.
Page 389
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
– Auto ranging is not supported for the phase result. NOTE – DC level result is not available if AC coupling is selected in the output configuration settings. Keysight U8903B User’s Guide...
Disable the Track 1st Ch to set the individual channels. Waveform Config – Frequency Set the frequency value. – Amplitude Set the amplitude value. – DC Offset Set the DC offset value. Keysight U8903B User’s Guide...
Freq Lock Selecting Gen Lock will allow the U8903B to search for the fundamental frequency based on the generator frequency value of the respective generator channel. This Gen Lock method is only useful if you use the internal audio generator.
Page 394
If Track 1st Ch is enabled, all the other channels will be set to follow the channel 1 range setting. Any changes made to channel 1 range setting will be reproduced in the other channels. Disable the Track 1st Ch to set the individual channels. – Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
Test Sequence Application Voice quality POLQA and PESQ measurements are only available with N3433A. Refer to “U8903B NOTE Options” on page for more information. This measurement is unavailable if a different interface is selected for the output and input configuration channel.
POLQA voice quality analysis may have a reduced wave file recording duration compared to NOTE the PESQ voice quality analysis based on the different option card configuration and settings in the U8903B. For more information, please refer to “POLQA Record ing Duration Analysis” on page 602. Signal generation and analysis The voice quality signal generation and analysis settings menu page is displayed as shown in Figure 7-60.
Page 397
– Rec File Set the file path for the exported recording destination. This setting is only available when Source is set to Generator, Reference File is set to Recorded, and Rec To File is set to Yes. Keysight U8903B User’s Guide...
Page 398
Press the IO Config softkey to set the IO configuration. IO Config – Output Level Set the output level value. Show/Hide POLQA Press the Show/Hide POLQA License softkey to show or hide the POLQA license information. License Keysight U8903B User’s Guide...
Press the Save Data softkey to save the selected result data to a CSV file format in the internal storage or external USB Save Data flash storage. Refer to “Save” on page for the Save menu page. Keysight U8903B User’s Guide...
Page 400
Set the X-axis right edge value. This setting is only available when Auto Scale is disabled. Graph An example of the graph result menu page is displayed as shown in Figure 7-62. Figure 7-62 Multitone Analyzer > Settings > Result (Spectrum) menu page Keysight U8903B User’s Guide...
Load limit points from a file. Refer to “Recall” on page for the Recall menu page. – Save Points Save the selected limit points to a file. Refer to “Save” on page for the Save menu page. Keysight U8903B User’s Guide...
Page 402
– Top Set the Y-axis top edge value. This setting is only available when Auto Scale is disabled. – Bottom Set the Y-axis bottom edge value. This setting is only available when Auto Scale is disabled. Keysight U8903B User’s Guide...
Test Sequence Application Report The U8903B allows you to generate a report of the test sequence results. The TSA > Report menu page is displayed as shown in Figure 7-63. Figure 7-63 TSA > Report menu page Table 7-61 TSA > Report menu description...
Prompts for the file name at the end of a test sequence. Prefix Press the Prefix softkey to set the prefix for the file name. Press the Location softkey to select the folder for the automatically saved report. Refer to “Location” on page Location more information. Keysight U8903B User’s Guide...
Page 405
Press the Select softkey to select the current folder as the location to be saved to. Press the New Folder softkey to create a new folder at the current directory or folder. Figure 7-65 Select Path menu page Keysight U8903B User’s Guide...
Page 406
Test Sequence Application THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight U8903B User’s Guide...
Keysight U8903B Performance Audio Analyzer User’s Guide HP8903B HP8903B Measurement Generator Sweep Special function code list SPCL This chapter describes the various configurations for the HP8903B mode.. The HP8903B mode is only available when GPIB is initialized successfully. Entering or NOTE exiting the HP8903B mode will cause the system to reset.
HP8903B HP8903B Press and select HP8903B to access the HP8903B menu page. The U8903B allows you to emulate the HP8903B audio analyzer behavior in the HP8903B mode. Figure 8-1 HP8903B menu page Keysight U8903B User’s Guide...
HP8903B Table 8-1 U8903B LCD display description Item Description SINAD meter Display the SINAD measurement mode readings. Display the measurement results. The left panel reading shows the frequency result while the Analyzer panel right panel reading shows the result of the selected measurement modes. Refer to “Measurement”...
Ratio Press the Ratio softkey to enable or disable the HP8903B measurement ratio mode. Press the Format softkey to select the HP8903B measurement format type. Format – Log – Lin Keysight U8903B User’s Guide...
Table 8-3 shows the measurement units that are applicable for the individual HP8903B measurement modes. Table 8-3 HP8903B unit charts Measurement Ratio On Ratio Off AC LEVEL dBm into 600 Ω DC LEVEL dBm into 600 Ω SINAD SIG/NOISE DSTN DSTN LEVEL dBm into 600 Ω Keysight U8903B User’s Guide...
10 Press the x 10 softkey to multiply the current parameter step value by 10. ÷ 10 Press the ÷ 10 softkey to divide the current parameter step value by 10. Keysight U8903B User’s Guide...
Press the Freq. Start softkey to set the HP8903B sweep frequency start value. Freq. Stop Press the Freq. Stop softkey to set the HP8903B sweep frequency stop value. Sweep Status Press the Sweep Status softkey to start or abort sweeping in the HP8903B mode. Keysight U8903B User’s Guide...
5.0 ~ 5.1 RMS Detector Display Source Settings 10.0 Activate the generator menu. 11.0 Restore last RATIO reference and enter RATIO mode if allowed Re-enter Ratio Mode 11.1 Display RATIO reference (measurement menu will be activated) Keysight U8903B User’s Guide...
Page 415
22.N 1 - Data Ready 2 - GPIB error 4 - Instrument Error The instrument powers up in the 22.2 state. 47.0 600 Ω Source Output Impedance (Instrument powers up at 600 Ω) 47.1 50 Ω Keysight U8903B User’s Guide...
Special function Generator impedance 47.0SP Change the generator impedance to 600 Ω. 47.1SP Change the generator impedance to 50 Ω. 47.0SP Change the generator impedance to 600 Ω. 47.1SP Change the generator impedance to 40 Ω. Keysight U8903B User’s Guide...
Keysight U8903B Performance Audio Analyzer User’s Guide Characteristics and Specifications For the characteristics and specifications of the U8903B Performance Audio Analyzer, refer to the data sheet at http://literature.cdn.keysight.com/litweb/pdf/5991-4551EN.pdf.
Keysight U8903B Performance Audio Analyzer User’s Guide Appendixes Appendix 1: FUNCTION panel Appendix 2: Graph Panel Menu Tree Appendix 3: System Panel Menu Tree Appendix 4: Analog Generator Menu Tree Appendix 5: Digital Generator Menu Tree Appendix 6: Bluetooth Generator Menu Tree...
Appendixes Appendix 22: U8903B Configuration Examples Example 1: Generating a sine waveform with the digital generator and measuring its voltage with the digital analyzer Example 2: Establishing a connection to a Bluetooth device and generating a sine waveform with the Bluetooth generator.
Customizable shortcut key. Refer to “Append ix 3: System Panel Menu Tree” on page Shortcut 1 more information. Customizable shortcut key. Refer to “Append ix 3: System Panel Menu Tree” on page Shortcut 2 more information. Keysight U8903B User’s Guide...
Set the left and right values of the graph to the current marker and the -> Delta reference marker location. Move to 3 dB Place the marker on the nearest –3 dB value data point. Move to 6 dB Place the marker on the nearest –6 dB value data point. Keysight U8903B User’s Guide...
Page 423
Perform autoscaling on the Y-axis by searching for the most optimum scale AutoScale Y (top and bottom values) for the Y-axis based on the data being displayed in the graph. Zoom Magnify a section of the graph. Keysight U8903B User’s Guide...
Save the current user state. Reset all modes to the default settings without Mode deleting the user-defined files. Reset the U8903B to the factory default settings, delete all the files saved in the internal flash Preset memory (for example, state file, arbitrary...
Page 425
Set the LAN subnet mask address. Gateway Set the LAN gateway address. DNS 1 Set the LAN DNS 1 address. DNS 2 Set the LAN DNS 2 address. GPIB 0 to 30 Set the desired GPIB address. Keysight U8903B User’s Guide...
Page 426
Securely erase all the files saved in the internal flash memory (for example, state file, arbitrary Secure Erase waveform, test sequence project files, and so on), It does not affect the U8903B settings and license information. Add a new U8903B option. Add Option Refer to “Append ix 13: Recall Menu Tree”...
Page 427
Volume 0 to 100 Select the aux audio volume. DC Output Set the aux DC output value. Board Info Display the U8903B available cards information. – On Legacy Status Enable or disable the legacy sweep. – Off Macro Reserved for future expansion.
Set the signal DC offset value. (Dual) Set the ratio of the amplitude of the second sine wave over the first sine Ratio wave. – RSS Summation Summation refers to the method to sum up two levels. – Linear Keysight U8903B User’s Guide...
Page 429
Set the number of cycles at which the amplitude is at the highest level. (Sine Burst) Set the number of cycles from the beginning of one burst to the beginning Period of the next burst. Low Level Set the amplitude ratio of Burst On over Burst Off. Keysight U8903B User’s Guide...
Page 430
Add the tone above the selected tone for the custom multitone waveform. Add Below Add the tone below the selected tone for the custom multitone waveform. Remove Remove the selected tone for the custom multitone waveform. Clear Clear all the tones in the custom multitone waveform list. Keysight U8903B User’s Guide...
Page 431
This setting is only available when the output connector type is set to IEC 60268. – Float Ground Select the grounding type. – Ground Max Voltage Set the maximum voltage value. Set the reference impedance for the unit conversion of dBm References Ref Impedance measurements. Keysight U8903B User’s Guide...
Set the signal frequency value. Waveform Config This is only available when the digital generator is (V. Phase) in channel 2. Phase -> 1 Set the phase of the channel 2 sine waveform with reference to channel 1. Keysight U8903B User’s Guide...
Page 433
Amplitude Set the signal amplitude value. Waveform Config (Triangular Noise) DC Offset Set the signal DC offset value. Amplitude Set the signal amplitude value. Waveform Config (Pink Noise) DC Offset Set the signal DC offset value. Keysight U8903B User’s Guide...
Page 434
Set the number of signal frequency components. – 1024 – 2048 – 4096 Select the waveform length that determines the Length – 8192 number of samples used to create one iteration of the multitone waveform. – 16384 – 32768 – 65536 Keysight U8903B User’s Guide...
Page 435
Sample Rate Set the sample rate value. Output Config – Linear PCM Audio Format – A-Law Select the encoding audio format. – μ-Law – Off – Rectangular Dither Select the dither type. – Triangular – Shaped Keysight U8903B User’s Guide...
Page 436
Select the DSI output format. – I2S – DSP Resolution 8 to 24 Set the audio resolution. – Rising BClk Edge Select the clock edge type. – Falling – Rising Fsync Polarity Select the sync polarity. – Falling Keysight U8903B User’s Guide...
Page 437
Select the copyright type. – Non-copyright Edit Bits (Consumer) – No pre-emphasis – 50/15 ìs Emphasis Select the emphasis type. – Reserved 1 – Reserved 2 Channel Mode 0 to 3 Set the channel mode value. Keysight U8903B User’s Guide...
Page 438
– 192 kHz – 768 kHz – Level 1 – Level 2 Clk Accuracy Select the clock accuracy type. – Level 3 – Reserved – 24 bits Max Word Len Select the maximum word length. – 20 bits Keysight U8903B User’s Guide...
Page 439
Select the byte and edit the byte. – Non-Linear PCM Audio Mode Select the audio mode. – Linear PCM Edit Bits (Professional) – Not indicated – No pre-emphasis Emphasis Select the emphasis type. – 50/15 μs – CCITT J.17 Keysight U8903B User’s Guide...
Page 440
User Bits Select the user bits type. – User defined – Reserved for Metadata – As in IEC60958-3 – 20-bit not defined – 24-bit main audio Aux Bits Select the auxiliary bits. – 20-bit single – Reserved Keysight U8903B User’s Guide...
Page 441
Channel Origin digit). Set the channel destination value (4 alphanumeric Channel Dest digit). Local Addr Set the local address value. 0 to 2 –1 Time of Day Set the time of day. 0 to 2 –1 Keysight U8903B User’s Guide...
Page 442
Load File Recall Menu Tree” on page for the recall menu tree. Set the volts/FS reference for the analog output References Volts/FS voltage (in Volts) of the DAC used when converted to digital full scale (1 FS). Keysight U8903B User’s Guide...
Set the signal DC offset value. (Dual) Set the ratio of the amplitude of the second sine wave Ratio over the first sine wave. – RSS Summation – Summation refers to the method to sum up two levels. – Linear Keysight U8903B User’s Guide...
Page 444
(Sine Burst) the highest level. Set the number of cycles from the beginning of one Period burst to the beginning of the next burst. Low Level Set the amplitude ratio of Burst On over Burst Off. Keysight U8903B User’s Guide...
Page 445
Set the number of signal frequency components. – 1024 – 2048 – 4096 Select the waveform length that determines the number Length – 8192 of samples used to create one iteration of the multitone waveform. – 16384 – 32768 – 65536 Keysight U8903B User’s Guide...
Page 446
1 to 16 identified. – All Link Config Device Scan Filter Class Select the filter class type. – Custom Custom Set the Bluetooth device class. Refresh the friendly name of the selected Bluetooth Refresh Device device. Keysight U8903B User’s Guide...
Page 447
– Send AT Command – Volume Up – Volume Down – Mute – Play – Stop Select the audio/video remote control profile operation AVRCP Operation settings. – Pause – Rewind – Fast Forward – Forward – Backward Keysight U8903B User’s Guide...
Page 448
Profile Select the U8903B’s Bluetooth card profile. – A2DP SNK, HFP HF, AVRCP CTRL Friendly Name Set the friendly name of the U8903B’s Bluetooth card. Device PIN Set the passkey of the U8903B’s Bluetooth card. – Auto Device Class Select the device class based on the configured profile.
Page 449
– 2EV5 – 3EV5 Clear Log Clear the AVRCP operation log history. AVRCP Log View Refresh Log Refresh the AVRCP operation log history. Set the reference impedance for the unit conversion of References Ref Impedance dBm measurements. Keysight U8903B User’s Guide...
Select the reading format of the returned measurement reading. Functions – Delta (Frequency) Store the measurement result from the selected channel as the reference Set result as ref. from Channels 1 to 8 frequency. Ref. Freq Set the reference frequency value. Keysight U8903B User’s Guide...
Page 451
Store the measurement result from the selected channel as the reference Set result as ref. from Channels 1 to 8 frequency. Set the reference level value. Ref. Level There is only one reference level for each channel. Keysight U8903B User’s Guide...
Page 452
This setting is only available when the frequency lock is set to Custom. – Enable Enable or disable the analog notch filter. Analog Notch – Disable This filter is used to emulate the analog notch filter used in the HP8903B. Keysight U8903B User’s Guide...
Page 453
This setting is only available when the frequency lock is set to Custom. – Enable Enable or disable the analog notch filter. Analog Notch – Disable This filter is used to emulate the analog notch filter used in the HP8903B. Keysight U8903B User’s Guide...
Page 454
Channels 1 to 8 ratio. Set the reference ratio value. Ref. Ratio There is only one reference ratio for each channel. – ALL – 2 Even Harmonic – 4 Display the even harmonics order selection. – 6 – 8 Keysight U8903B User’s Guide...
Page 455
Set the calibration level value. Cal SPL This setting is only available when the unit is set to dBSPL. – ALL – 2 Even Harmonic – 4 Select the even harmonics values. – 6 – 8 Keysight U8903B User’s Guide...
Page 456
Set the upper frequency value. Upper Freq This setting is only available when the frequency lock is set to Custom. Set the lower frequency value. Lower Freq This setting is only available when the frequency lock is set to Custom. Keysight U8903B User’s Guide...
Page 457
Store the measurement result from the selected channel as the reference Set result as ref. from Channels 1 to 8 frequency. Set the reference ratio value. Ref. Ratio There is only one reference ratio for each channel. Keysight U8903B User’s Guide...
Page 458
– Auto Freq Lock – Gen Lock Select the searching method for the fundamental frequency. – Custom Set the fundamental frequency value. Fund Freq This setting is only available when the frequency lock is set to Custom. Keysight U8903B User’s Guide...
Page 459
Enable or disable the notch filter. State – Enable – Disable Notch Filter Center Freq Set the frequency of the component to be removed from the input signal. Bandwidth Set the band width of the signal component to be removed. Keysight U8903B User’s Guide...
Page 460
– 100 V – 32 V Select the input voltage range value. Range – 10 V This setting is only available when the auto range is disabled. – 3.2 V – 1 V – 320 mV Keysight U8903B User’s Guide...
Page 461
Level 1 Level 2 Level 3 Description Select the input bandwidth value. – 90 kHz Bandwidth This setting is only available with Option N3431A. Refer to “U8903B – 1.5 MHz Options” on page for more information. Input Config – DC Coupling Select the input coupling type.
Select the reading format of the returned Format Functions measurement reading. – Delta (Frequency) Store the measurement result from the selected Set result as ref. from Channels 1 to 8 channel as the reference frequency. Ref. Freq Set the reference frequency value. Keysight U8903B User’s Guide...
Page 463
Store the measurement result from the selected Set result as ref. from Channels 1 to 2 channel as the reference frequency. Set the reference voltage value. Ref. Voltage There is only one reference level for each channel. Keysight U8903B User’s Guide...
Page 464
Set the reference level value. Ref. Level There is only one reference level for each channel. Volt/FS Set the volts/FS reference value. – On Precision Enable or disable the precision mode. – Off Keysight U8903B User’s Guide...
Page 465
Set the reference ratio value. Ref. Ratio There is only one reference ratio for each channel. – ALL – 2 Even Harmonic – 4 Display the even harmonics order selection. – 6 – 8 Keysight U8903B User’s Guide...
Page 466
Set the volts/FS reference value. – ALL – 2 Even Harmonic – 4 Select the even harmonics values. – 6 – 8 – ALL – 3 Odd Harmonic – 5 Select the odd harmonics values. – 7 – 9 Keysight U8903B User’s Guide...
Page 467
Set result as ref. from Channels 1 to 2 channel as the reference frequency. Functions Set the reference level value. (Positive peak) Ref. Level There is only one reference level for each channel. Volt/FS Set the volts/FS reference value. Keysight U8903B User’s Guide...
Page 468
– Off Select the reading format of the returned Format – Linear measurement reading. – Delta Store the measurement result from the selected Set result as ref. from Channels 1 to 2 channel as the reference ratio. Keysight U8903B User’s Guide...
Page 470
– Digital Select the trigger interface. This is only available when the trigger source is Channel. Channel Set the channel number. This is only available when the trigger source is Channel. Trigger Level Set the trigger level. Keysight U8903B User’s Guide...
Page 471
Input Config type – 110 Ω – HiZ – MISR Freq Scaling Select the frequency scaling type. – Custom This is only available when frequency scaling is Custom. Ref SR Set the reference sample rate value. Keysight U8903B User’s Guide...
Page 472
Bit Clk Edge Select the bit clock edge. – Falling – 1.2 Vpp – 1.5 Vpp – 1.8 Vpp Voltage – 2.5 Vpp Select the input voltage value. – 3 Vpp – 3.3 Vpp – Custom Keysight U8903B User’s Guide...
Page 473
No. of Reading 2 to 20 calculation. – Min Statistics – Max Stat 1 Stat2 – Average Select the statistics calculation type. Stat3 – Std Dev – ΔMinMax Reset the statistics results of the current analog Clear analyzer. Keysight U8903B User’s Guide...
Select the reading format of the returned measurement Format Functions reading. – Delta (Frequency) Set result as ref. Store the measurement result from the selected Channels 1 to 2 from channel as the reference frequency. Ref. Freq Set the reference frequency value. Keysight U8903B User’s Guide...
Page 475
Set result as ref. Store the measurement result from the selected Channels 1 to 2 from channel as the reference frequency. Set the reference level value. Ref. Level There is only one reference level for each channel. Keysight U8903B User’s Guide...
Page 476
Set the reference level value. Ref. Level There is only one reference level for each channel. Set the calibration level value. Cal SPL This setting is only available when the unit is set to dBSPL. Keysight U8903B User’s Guide...
Page 477
Set the reference ratio value. Ref. Ratio There is only one reference ratio for each channel. – ALL – 2 Even Harmonic – 4 Display the even harmonics order selection. – 6 – 8 Keysight U8903B User’s Guide...
Page 478
Set the calibration level value. Cal SPL This setting is only available when the unit is set to dBSPL. – ALL – 2 Even Harmonic – 4 Select the even harmonics values. – 6 – 8 Keysight U8903B User’s Guide...
Page 479
Set result as ref. Store the measurement result from the selected Channels 1 to 2 from channel as the reference ratio. Set the reference ratio value. Ref. Ratio There is only one reference ratio for each channel. Keysight U8903B User’s Guide...
Page 480
Set the reference ratio value. Ref. Ratio There is only one reference ratio for each channel. Set the fundamental frequency value. Fund Freq This setting is only available when the frequency lock is set to Custom. Keysight U8903B User’s Guide...
Page 481
Filters Config State – Enable – Disable Notch Filter Set the frequency of the component to be removed from Center Freq the input signal. Set the band width of the signal component to be Bandwidth removed. Keysight U8903B User’s Guide...
Page 482
1 to 16 identified. – All Link Config Device Scan Filter Class Select the filter class type. – Custom Custom Set the Bluetooth device class. Refresh the friendly name of the selected Bluetooth Refresh Device device. Keysight U8903B User’s Guide...
Page 483
– Send AT Command – Volume Up – Volume Down – Mute – Play – Stop Select the audio/video remote control profile operation AVRCP Operation settings. – Pause – Rewind – Fast Forward – Forward – Backward Keysight U8903B User’s Guide...
Page 484
Profile Select the U8903B’s Bluetooth card profile. – A2DP SNK, HFP HF, AVRCP CTRL Friendly Name Set the friendly name of the U8903B’s Bluetooth card. Device PIN Set the passkey of the U8903B’s Bluetooth card. – Auto Device Class Select the device class based on the configured profile.
Page 485
2 to 20 calculation. – Min Statistics – Max Stat 1 Stat2 – Average – Select the statistics calculation type. Stat3 – Std Dev – ΔMinMax Reset the statistics results of the current analog Clear analyzer. Keysight U8903B User’s Guide...
Page 487
– Off Save To Memory Save the active trace to the memory buffer. Memory Load From Memory Load a trace from the memory buffer into the active trace. Clear Memory Clear the trace from the memory buffer. Keysight U8903B User’s Guide...
Page 488
– Red – Purple – Graph – Data Table – Marker Table Display Option Select the graph analysis display option. – Statistics – Harmonics – Signal Analysis Continuous Running mode Select the graph analysis running mode. Single Keysight U8903B User’s Guide...
Page 489
Select the function to be displayed based on the function number (1 to 4). Interface Select the measurement interface type. Measurement 2 Channel Select the desired channel to perform the measurement. Function No Select the function to be displayed based on the function number (1 to 4). Keysight U8903B User’s Guide...
Sweep Mode Select the sweep mode. Single Source 1 to 2 Select the source channel(s). Select the measure channel(s). Channels Measure 1 to 8 The available selection depends on the number of installed analog analyzer cards. Keysight U8903B User’s Guide...
Page 491
1> – F2: <measurement X Data Select the X-axis trace data source. function 2> – F3: <measurement function 3> – F4: <measurement function 4> – On Wrap Phase Enable of disable the wrap phase. – Off Keysight U8903B User’s Guide...
Page 492
Refer to “Append ix 12: Save Menu Tree” on page for the save menu tree. Plot View Save the sweep points to a file. Save Pts Refer to “Append ix 12: Save Menu Tree” on page for the save menu tree. Keysight U8903B User’s Guide...
Copy the marked file to the selected folder. Move Marked To Folder Move the marked file to the selected folder. Delete Delete the selected file. New Folder Set the new folder name and select OK to create a new folder. Keysight U8903B User’s Guide...
Copy the marked file to the selected folder. Move Marked To Folder Move the marked file to the selected folder. Delete Delete the selected file. New Folder Set the new folder name and select OK to create a new folder. Keysight U8903B User’s Guide...
Properties – Prompt Prefix Set the prefix for the file name. Select the folder for the automatically saved Select report. Location Create a new folder at the current directory or New Folder folder. Keysight U8903B User’s Guide...
Page 496
– Pin 2 the common IEC 60268 configuration. IEC60268 – Pin 3 This setting is only available when the output connector type is set to IEC 60268. – Float Ground Select the grounding type. – Ground Keysight U8903B User’s Guide...
Page 497
– 2 Device Select the paired Bluetooth device. Settings: Output – A2DP Configuration Profiles – HFP Select the U8903B’s Bluetooth card profile. (Bluetooth) – AVRCP Link Config Configure the Bluetooth link settings. References Volt/FS Set the output references for generator.
Page 498
Channels Select the number of input channels in use. Device Select the paired Bluetooth device. – A2DP Profiles – HFP Select the U8903B’s Bluetooth card profile. – AVRCP Link Config Configure the Bluetooth link settings. – DC Settings: Input Coupling Select the input coupling type.
Page 499
Add sub-step to the list. – Delay Add Sub-Step – Prompt Sub-Steps: – Sent SCPI Send SCPI Delete Sub-Step Delete the selected sub-step. Move or copy the selected sub-step, or paste a Edit copied sub-step after the selected sub-step. Keysight U8903B User’s Guide...
Page 500
Delete the selected test measurement. Move or copy the selected measurement, or paste Edit a copied measurement after the selected measurement. Configure the selected test measurement. Settings Refer to “Measurement settings” on page for the respective test measurement settings. Keysight U8903B User’s Guide...
Page 501
Select the prompt icon to be displayed. – None – Hand – Question Settings – Exclamation – Asterisk – Dialog Settings Add additional settings to the prompt window. – Timeout – Cancel Button – Timeout Set the prompt timeout value in seconds. Keysight U8903B User’s Guide...
Page 502
Set the delay time length after the SCPI command(s) is executed. Properties Set the prompt sub-step name. – Cancel Seq. Select the failure handling type for the selected Failure Handling – Allow Retry measurement. – Continue Seq. Keysight U8903B User’s Guide...
Page 504
Track 1st Ch Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. AC Level Refer to “Bar chart” on page for the bar chart menu tree. Gain Keysight U8903B User’s Guide...
– Enable Track 1st Ch Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. Refer to “Bar chart” on page for the bar Frequency chart menu tree. Keysight U8903B User’s Guide...
Track 1st Ch Enable or disable the tracking of the first channel. – Disable Input Range Voltage Range Set the input voltage range. Refer to “Bar chart” on page for the bar Phase chart menu tree. Keysight U8903B User’s Guide...
This setting is only available when SNR Mode is set to Fast and Freq Lock is set to Custom. Set the number of harmonics order to be removed. Harmonic Cnt This setting is ony available when SNR Mode is set to Fast. Keysight U8903B User’s Guide...
Page 509
Input Range – Enable Track 1st Ch Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. Refer to “Bar chart” on page for the bar chart menu tree. Keysight U8903B User’s Guide...
Page 511
Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. SINAD THD Level Refer to “Bar chart” on page for the bar THD Ratio chart menu tree. THD+N Level THD+N Ratio Keysight U8903B User’s Guide...
– Enable Track 1st Ch Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. Refer to “Bar chart” on page for the bar DC Level chart menu tree. Keysight U8903B User’s Guide...
Input Range Track 1st Ch Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. Refer to “Bar chart” on page for the bar Crosstalk chart menu tree. Keysight U8903B User’s Guide...
– Enable Track 1st Ch Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. Refer to “Bar chart” on page for the bar SMPTE Ratio chart menu tree. Keysight U8903B User’s Guide...
– Enable Track 1st Ch Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. Refer to “Bar chart” on page for the bar chart DFD Ratio menu tree. Keysight U8903B User’s Guide...
– Enable Track 1st Ch Enable or disable the tracking of the first channel. Multitone – Disable Input Range (Signal Analysis) Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
Page 517
– Level – Gain – Max Tone Level – Min Tone Level Refer to “Bar chart” on page for the bar – TD+N Level chart menu tree. – TD+N Ratio – Tone Level Keysight U8903B User’s Guide...
DC Offset Set the DC offset value. Ref Channel Set the reference channel number. – All Select the harmonics count to be used in the THD Harmonics ratio and THD level results. – 2 to 9 Keysight U8903B User’s Guide...
Page 520
Set the input voltage range. – AC Level – Gain – Phase – THD Ratio Refer to “Graph” on page for the graph menu tree. – THD Level – THD+N Ratio – THD+N Level – SINAD Keysight U8903B User’s Guide...
Enable or disable the tracking of the first channel. – Disable Upper Freq Set the upper frequency value. Waveform Config Lower Freq Set the lower frequency value. Amplitude Set the amplitude value. DC Offset Set the DC offset value. Keysight U8903B User’s Guide...
Page 522
Input Range – Enable Track 1st Ch Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. Refer to “Graph” on page for the graph SMPTE Ratio menu tree. Keysight U8903B User’s Guide...
Enable or disable the tracking of the first channel. – Disable Upper Freq Set the upper frequency value. Waveform Config Diff Freq Set the difference frequency value. Amplitude Set the amplitude value. DC Offset Set the DC offset value. Keysight U8903B User’s Guide...
Page 524
Input Range – Enable Track 1st Ch Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. Refer to “Graph” on page for the graph DFD Ratio menu tree. Keysight U8903B User’s Guide...
Select the instrument model. – Keysight 33600A – Other GPIB Address – 0 to 30 Select the GPIB address. Enable or disable the U8903B to send SCPI – On Init Instrument commands to initialize the connected – Off signal generator.
Page 527
Set the input voltage range. – AC Level – Gain – Phase – THD Ratio Refer to “Graph” on page for the graph menu tree. – THD Level – THD+N Ratio – THD+N Level – SINAD Keysight U8903B User’s Guide...
– Enable Track 1st Ch Enable or disable the tracking of the first channel. – Disable Waveform Config Frequency Set the frequency value. DC Offset Set the DC offset value. Keysight U8903B User’s Guide...
Page 530
Refer to “Graph” on page for the graph – THD Level Vs Measured Amplitude menu tree. – THD+N ratio – THD+N Level – THD+N Ratio Vs Measured Amplitude – THD+N Level Vs Measured Amplitude – SINAD Keysight U8903B User’s Guide...
Track 1st Ch Enable or disable the tracking of the first channel. – Disable Upper Freq Set the upper frequency value. Waveform Config Lower Freq Set the lower frequency value. DC Offset Set the DC offset value. Keysight U8903B User’s Guide...
Page 532
Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. – SMPTE Ratio Refer to “Graph” on page for the graph menu tree. – SMPTE Ratio Vs Measured Amplitude Keysight U8903B User’s Guide...
Track 1st Ch Enable or disable the tracking of the first channel. – Disable Upper Freq Set the upper frequency value. Waveform Config Diff Freq Set the difference frequency value. DC Offset Set the DC offset value. Keysight U8903B User’s Guide...
Page 534
Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. – DFD Ratio Refer to “Graph” on page for the graph menu tree. – DFD Ratio Vs Measured Amplitude Keysight U8903B User’s Guide...
Page 536
Input Range – Enable Track 1st Ch Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. Refer to “Graph” on page for the graph DC Level menu tree. Keysight U8903B User’s Guide...
Set the sweep stop RF power. Step Size Set the step size. Set the delay between each measured SINAD in Dwell Time seconds. Enable or disable the U8903B to send SCPI – On Init Instrument commands to initialize the connected signal – Off generator.
Page 539
– Enable Track 1st Ch Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. Refer to “Graph” on page for the graph SINAD Vs RF Power menu tree. Keysight U8903B User’s Guide...
Select the instrument model. – Keysight 33600A – Other GPIB Address – 0 to 30 Select the GPIB address. Enable or disable the U8903B to send SCPI – On Init Instrument commands to initialize the connected – Off signal generator.
Page 542
Refer to “Graph” on page for the graph – THD Level Vs Measured Amplitude menu tree. – THD+N ratio – THD+N Level – THD+N Ratio Vs Measured Amplitude – THD+N Level Vs Measured Amplitude – SINAD Keysight U8903B User’s Guide...
Page 544
Auto Range Enable or disable the auto input range. – Off Input Range – Enable Track 1st Ch Enable or disable the tracking of the first channel. – Disable Voltage Range Set the input voltage range. Keysight U8903B User’s Guide...
Page 545
Level 1 Level 2 Level 3 Description – AC Level – Gain, Phase – THD+N Ratio Refer to “Graph” on page for the graph – THD+N Level menu tree. – DC Level – Frequency – SINAD Keysight U8903B User’s Guide...
Recorded, and Rec Step is set to Delay. Enable or disable exporting the recording – Yes function. This setting is only available when Rec to File Source is set to Generator and Reference File is – No set to Recorded. Keysight U8903B User’s Guide...
Page 547
IO Config Output Level Set the output level value. Show/Hide POLQA Show or hide the POLQA license information. License – MOS- LQO Refer to “Bar chart” on page for the bar chart menu tree. – Levels Keysight U8903B User’s Guide...
Select the X-axis unit type. X-axis Set the X-axis left value. Left This setting is only available when the Autoscale is disabled. Set the X-axis right value. Right This setting is only available when the Autoscale is disabled. Keysight U8903B User’s Guide...
Refer to “Append ix 13: Recall Menu Tree” on page for the recall menu tree. Save the selected limit points to a file. Edit Limits Points Save Points Refer to “Append ix 12: Save Menu Tree” on page for the save menu tree. Keysight U8903B User’s Guide...
Page 550
– CH1 to CH8 Enable or disable the trace. State – On – Off Select the trace color. – Yellow Trace – Cyan – White Color – Pink – Green – Orange – Red – Purple Keysight U8903B User’s Guide...
Page 551
MOS-LQO and Delay results) Set the X-axis top edge value. This setting is only available when the Autoscale is disabled. Set the Y-axis bottom edge value. Bottom This setting is only available when the Autoscale is disabled. Keysight U8903B User’s Guide...
Divide the current parameter step value by 10. Freq. Start Set the HP8903B sweep frequency start value. Freq. Stop Set the HP8903B sweep frequency stop value. Sweep – On Sweep Status Start or abort sweeping in the HP8903B mode. – Off Keysight U8903B User’s Guide...
Page 553
– GPIB Address (SCPI only) – GPIB Service Request Condition (SCPI only) – Source Output Impedance (Instrument powers up at 600 Ω) Set the HP8903B special function code except those indicated as SCPI only SPCL and execute the special function. Keysight U8903B User’s Guide...
6.02 dB if a load is present. [b] Reference level is defined as a user-entered or a captured value from the current reading as a relative level for the subsequent measurement reading. It can be set to delta, linear, or log scale. Keysight U8903B User’s Guide...
Decibels relative to Full Scale Least Significant Bit FS/Vrms Ratio between cross-domain input and output levels measurements (analog input and digital output) Vrms/FS Ratio between cross-domain input and output levels measurements (digital input and analog output) Keysight U8903B User’s Guide...
Appendixes Appendix 17: Arbitrary File Format The U8903B arbitrary waveform mode allows you to load an arbitrary file or a wave file. To load waveform file for the analog arbitrary waveform, press the Waveform Config > Recall softkeys. The Recall menu page will be displayed to allow you to select the file to be loaded.
Page 560
– Summation of the Vpeak and DC offset exceeds the maximum voltage for the current output connection type. – Invalid Vpeak and DC offset values. – Total of sample points less than 32. – The arbitrary file does not exist. Keysight U8903B User’s Guide...
Page 563
The number of FIR coefficients must not be less than four, while the minimum number of sections allowed for IIR is one (seven coefficients). The delay is specified in the form of samples and within the range of 0 to 65535. Keysight U8903B User’s Guide...
DSI Ana FS-InOut In/Out Digital analyzer DSI frame sync DSI Ana Clk-InOut In/Out Digital analyzer DSI bit clock – Digital ground Sync Clock Out Sync clock output for AES3/SPDIF and DSI 14 – 25 – Digital ground Keysight U8903B User’s Guide...
NOTE on the pins assignment. Configuration 1 In this configuration, the DUT uses the U8903B internal reference clock source as the reference clock. The DUT receives the clock references and data from the U8903B. Press on the FUNCTION panel to switch between audio generator or audio analyzer mode, and press on the FUNCTION panel to switch to digital interface.
In this configuration, the DUT uses an external master clock as the reference clock. The U8903B receives the data, frame sync, and bit clock from the DUT. A typical application for this configuration is analog to digital converter (ADC) evaluation.
In this configuration, an external master clock is used to synchronize the DUT and U8903B. A phase-locked loop (PLL) is used in the U8903B to lock the incoming master clock and regenerate the frame sync and bit clock. Data will be clocked in on each bit clock.
Appendixes Configuration 4 This configuration is similar to configuration 2, except that the DUT uses the U8903B internal reference clock source as the reference clock. Press on the FUNCTION panel to switch between audio generator or audio analyzer mode, and press on the FUNCTION panel to switch to digital interface.
Appendixes Configuration 5 This configuration is similar to configuration 3, except that the DUT uses the U8903B internal reference clock source as the reference clock. Press on the FUNCTION panel to switch between audio generator or audio analyzer mode, and press on the FUNCTION panel to switch to digital interface.
Example 1: Generating a sine waveform with the digital generator and measuring its voltage with the digital analyzer In this example, you will learn how to generate a simple sine waveform from the U8903B digital generator and measure its voltage using the U8903B digital analyzer.
Page 572
You should now obtain an AC voltage reading of 1 FFS for the generated sine waveform, within the tolerance as stated in “Characteristics and Specifications” on page 417. The U8903B display should look as follows. Figure A-10 Digital analyzer voltage measurement...
10 Press Waveform Config > Amplitude, and set the amplitude to 2 FFS. 11 Press to start the signal generation on the Bluetooth generator channel 1. 12 After you have completed the steps above, the U8903B display should look as follows. Keysight U8903B User’s Guide...
Page 574
Appendixes Figure A-11 Bluetooth generator sine waveform generation Keysight U8903B User’s Guide...
Example 3: Generating a multitone waveform with the analog generator and viewing the multitone spectrum in the graph In this example, you will learn how to generate a multitone waveform from the U8903B analog generator and view the multitone spectrum in the U8903B graph.
Page 576
Appendixes Figure A-12 Analog analyer multitone waveform measurement To view the multitone spectrum in the U8903B graph, perform the following steps. 1 Press on the FUNCTION panel to access the graph analysis mode. 2 Press to enable the graph analysis mode.
Example 4: Running the Test Sequence App for the multitone analyzer measurement In this example, you will learn how to use the Test Sequence App for the multitone analyzer measurement in the U8903B. 1 Press to switch to Test Sequence App mode.
Page 578
Appendixes 7 Select the Report tab to view the test sequence results in the U8903B display. You can also save the result file as a .docx file type by pressing Save. Figure A-15 Test Sequence App report for multitone analyzer...
Appendixes Example 5: Configuring the system clock reference settings In this example, you will learn how to configure the U8903B system clock reference settings. To configure the system clock reference source to external Master clock in with word length of 24 and multiplier of 192, perform the following steps.
Appendixes Example 6: Configuring the digital generator DSI output settings In this example, you will learn how to configure the U8903B digital generator DSI output settings. To configure the digital generator DSI output settings to DSP format, sampling rate of 192 kHz, word length of 24, and multiplier of 192, perform the following steps.
Channel 1 and Channel 2 are the same Amplitude Channel 1 and Channel 2 can be different Gaussian/Rectangular/ Triangular/Pink DC Offset Channel 1 and Channel 2 are the same Constant Amplitude Channel 1 and Channel 2 are the same Keysight U8903B User’s Guide...
Page 582
Channel 1 and Channel 2 are the same Tone Phase Channel 1 and Channel 2 are the same Amplitude Channel 1 and Channel 2 can be different Arbitrary DC Offset Channel 1 and Channel 2 are the same Keysight U8903B User’s Guide...
Functions (Frequency) Format Unit Functions Format (AC Voltage) Detector Unit Functions (DC Voltage) Format Unit Format Functions (THD+N Ratio/SINAD) Freq Lock Auto Fund Freq 1000 Unit Format Functions (THD+N Level) Freq Lock Auto Fund Freq 1000 Keysight U8903B User’s Guide...
Page 597
(SMPTE IMD) Upper Freq Lower Freq 7000 Unit Functions Format (SNR) SNR Delay 0 ms Unit Format Functions Freq Lock Auto (SNR (Fast)) Fund Freq 1000 Harmonics Ref. Channel Functions Freq Lock Auto (Phase) Fund Freq 1000 Keysight U8903B User’s Guide...
100 kΩ (Unbal) Impedance 200 kΩ (Bal) Input Config Coupling Bandwidth 90 kHz Ext. Gain 0 dB Channel Left Wave File Bits/Sample Duration 10 s Show Stats No. of Reading Statistics Stat 1 Stat 2 Stat 3 Average Keysight U8903B User’s Guide...
1 In the file view, select the file to delete. 2 Press the Delete softkey. 3 A dialog box will pop up prompting confirmation to delete the file. To delete the file, select Yes and press . To abort, select No and press Keysight U8903B User’s Guide...
Appendixes Appendix 28: POLQA Recording Duration Analysis The following shows the maximum wave file recording duration allowed for POLQA analysis based on the different option card configuration and settings in the U8903B. Table A-55 U8903B option configuration Option Configuration Analog cards only...
This allows you to set the gain of the analyzer to amplify a low input signal or attenuate a high input signal as required. The U8903B provides the following input voltage ranges for both unbalanced and balanced inputs: – 140 V –...
Using manual ranging for continuous sweep and frequency response measurements For continuous sweep and frequency response measurements, manual ranging is a better option as it ensures signals across all frequency ranges are measured within the same range. Keysight U8903B User’s Guide...