IQ Analyzer (IQA) – Option 351 (CPU2 Only)
I/Q Analyzer (IQA) Mode Settings
FFT Window
K-B (120 dB)
Gaussian (60 dB)
Gaussian (70 dB)
Gaussian (80 dB)
Sinc
Tukey25
Changing between the Continuous and Single captures and adjusting
the Trigger settings
— Press Sweep 3.
— Then choose from the following:
Keysight N9927-90020 User's Guide
IF BW
Sets the Digital IF bandwidth of the instrument. The
relationship between the IF bandwidth and sample rate can be
described by the following equation:
sample rate = 1.25 * IF BW.
(Default: 10 MHz, minimum: 10 Hz, maximum: 10 MHz)
Single
When pressed, Single automatically sets Continuous to OFF
and causes the application to make one acquisition.
Continuous
Toggles Continuous between ON and OFF. When
Continuous is ON, each Run/Hold press causes the application to
toggle between continuous acquisitions and Hold. When Continuous
is OFF, each Run/Hold or Single press causes the application to
make one acquisition.
Restart
When pressed, Restart causes the FieldFox to restart trace
averaging at count of 1. When in Single sweep mode, the IQA will
perform a complete set of averages before returning to the hold
state.
Start Time
(Waveform only) Sets the start time of the waveform
measurement (default: 0s, minimum: 0s, maximum: dynamic). In
Waveform mode Setting the Start and stop time enables you to
display a subset of the data on the display in order to speed up the
Description
This is a Kaiser-Bessel window whose maximum side-lobe
level is -120 dB.
This is a Gaussian window whose maximum side-lobe level is
-60 dB.
This is a Gaussian window whose maximum side-lobe level is
-70 dB.
This is a Gaussian window whose maximum side-lobe level is
-80 dB.
This window is the central lobe of the sin(x)/x (often called
sinc(x)) function.
This is a tukey (cosine-tapered) window with alpha equal to
0.25.