How To Set Single Or Continuous; Step Detection Mode; How To Set Step Detection Mode; Limits - Keysight A Series User Manual

Hide thumbs Also See for A Series:
Table of Contents

Advertisement

Power Meter (USB/LAN) Mode - Option 302
USB and LAN Power Meter Settings

How to set Single or Continuous

— Press Sweep 3.
— Then choose one of the following:
You can also use Run / Hold +/- to toggle between Single and Continuous.

Step Detection Mode

The FieldFox supports the Step Detection feature that is present in Keysight
USB power sensors. When enabled, this feature reduces settling time after a
significant step in the measured power. For more information, please refer to
the USB Power Meter documentation.

How to set Step Detection mode

— Press Meas Setup 4.
— Then
— Then choose from the following:

Limits

In Power Meter Mode, Minimum and Maximum limits set a visible range of
acceptable readings between these two values. Readings that are below the
minimum or above the maximum are clearly visible on the power meter display.

How to set Limits

— Press Limit 8.
— Then choose from the following:
552
Single
Automatically sets Continuous OFF and causes FieldFox to
make ONE measurement, then hold for the next
When a data trace is displayed, the entire trace is measured, then
holds. The Hold annotation changes to an arrow --> while the
measurement occurs.
Continuous OFF ON
typical setting when battery power is not critical.
Step Detection OFF ON
— ON Enables step detection (default setting)
— OFF Disables step detection
Min Limit OFF ON
Toggles the display of the Minimum limit ON and
OFF.
Min Limit Value
Sets the Minimum limit power value. Power reading
lower than this will exceed the limit.
Makes continuous measurements. This is the
Keysight N9927-90020 User's Guide
Single
key press.

Hide quick links:

Advertisement

Table of Contents
loading