Single / Continuous / Restart; How To Set Single Or Continuous; Points; How To Set Number Of Points - Keysight A Series User Manual

Hide thumbs Also See for A Series:
Table of Contents

Advertisement

SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings

Single / Continuous / Restart

This setting determines whether the FieldFox measures continuously or only
once each time the
Single or to conserve battery power or to allow you to save or analyze a
specific trace.

How to set Single or Continuous

1. Press Sweep 3.
2. Then choose one of the following:

Points

Points is the number of measurements that are displayed along the X-axis. The
higher number of data points, the better the ability to resolve closely spaced
signals and the slower the sweep speed.

How to set number of points

1. Press Sweep 3 >
2. Then choose from:

Trace Display States (SA Mode)

In SA Mode you can display up to four of the following types of trace states. All
SA settings are applied to all displayed traces.
and SA Mode)" on page
212
Single
or Run / Hold +/- button is pressed. Use Hold /
Continuous
ON
Causes the SA to make continuous sweeps. This is the
typical setting for making measurements. Each Run / Hold
+/- press causes the SA to toggle between continuous
sweep and Hold.
OFF
Causes the SA to Hold. Each Run / Hold +/- or
press causes the SA to make ONE sweep, then Hold. The
Hold annotation changes to an arrow --> while the sweep
occurs.
Single
Automatically sets Continuous OFF and causes the SA to
make ONE sweep, then Hold.
Restart
Resets the Average counter to 1. The Average counter also
is used with MaxHold and MinHold Trace States. Learn more in
"Average Count" on page
Points [current setting]
101 | 201 | 401 | 601 | 801 | 1001

# Points

to enter a trace points value manually (Maximum = 20001).
704.
214.
See also
"Marker Trace (IQA
Keysight N9927-90020 User's Guide
Single