Declarations of Conformity for this where in the EULA. Keysight shall be product and for other Keysight prod- under no obligation to update, revise or ucts may be downloaded from the WARNING otherwise modify the Software.
Caution, risk of danger (refer to this manual Caution, risk of electric shock for specific Warning or Caution information) Earth (ground) terminal Frame or chassis (ground) terminal Equipment protected throughout by double Protective earth (ground) terminal insulation or reinforced insulation Keysight U8903B Service Guide...
Failure to comply with these precautions or with specific warnings elsewhere in this manual violates safety standards for design, manufacture, and intended use of the instrument. Keysight Technologies assumes no liability for the customer’s failure to comply with these requirements.
Environmental Conditions The U8903B is designed for indoor use and in an area with low condensation. The table below shows the general environmental requirements for this instrument. Environmental cond ition Requirement Operating condition – 0 °C to 55 °C Temperature Storage condition –...
Page 6
Regulatory Information The U8903B complies with the following safety and Electromagnetic Compatibility (EMC) compliances: Safety compliance – IEC 61010-1:2010/EN 61010-1:2010 – Canada: CAN/CSA-C22.2 No. 61010-1-12 – USA: ANSI/UL Std. No. 61010-1 (3rd Edition) EMC compliance – IEC 61326-1:2005/EN61326-1:2006 – Canada: ICES-001:2004 –...
Page 7
Forty years is the expected MSIP-REM-ATI- the home. useful life of the product. WCAUD8903B Keysight U8903B Service Guide...
Page 8
To return this unwanted instrument, contact your nearest Keysight Service Center, or visit www.keysight.com/environment/product for more information. Sales and Technical Support To contact Keysight for sales and technical support, refer to the support links on the following Keysight websites: – www.keysight.com/find/U8903B...
Table 5-5 A2DP APTX functional test points and limits for analyzer and generator Table 5-6 U8903B Bluetooth test to be repeated and its prerequisite ...94 Table A-1 Test record for DC accuracy verification .
Page 17
Performance Audio Analyzer Service Guide Performance Verification Overview Introduction Self-Test Keysight Calibration Services Calibration Interval Recommended Test Equipment Test Consideration This chapter provides an overview on performance verification of the U8903B. It also includes a list of recommended test equipment.
Performance Verification Overview Introduction This section provides a brief overview for verifying the performance of the U8903B. Performance verifications allow you to verify that the U8903B is operating within its published specifications. Figure 1-1 illustrates the process flow for carrying out the self-test, performance tests, and repair.
If all tests pass, you have a high confidence that the U8903B is operational. Keysight Calibration Services When your U8903B is due for calibration, contact your local Keysight Service Center for a low-cost recalibration. The U8903B is supported on automated calibration and adjustment systems, which allow Keysight to provide this service at competitive prices.
– The verification and adjustment procedures are based on the assumption that the recommended test equipment is being used. Substituting with an alternative test equipment may require modification of some procedures. Keysight U8903B Service Guide...
Page 22
Performance Verification Overview THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight U8903B Service Guide...
Sine wave amplitude accuracy and flatness verification Sine wave frequency accuracy verification Crosstalk verification Square wave rise time verification Square wave amplitude accuracy verification The performance verification procedures described in this chapter verify that the U8903B analog generator is operating within its published specifications.
Performance Verification for Analog Generator Introduction This section lists the procedures to test the electrical performance of the U8903B analog generator using the published specifications given in the U8903B Performance Audio Analyzer User’s Guide (U8903-90045) as the performance standards. If the analog generator fails any of the tests or if any abnormal test results are obtained, adjustment will need to be carried out accordingly.
Performance Verification for Analog Generator Performance Verification DC accuracy verification This test verifies the accuracy of the DC signal generated by the U8903B over a range of amplitudes using two digital multimeters to ensure that it is within the published specification.
– Output grounding configuration: Floating ground – Waveform: DC – Output state: ON 4 Set both the U8903B output channel 1 and channel 2 to be tested to produce DC voltage with amplitude of V as given in the “Test record for DC accuracy verification”...
– Waveform: Sine – Output state: ON 4 Set both the U8903B output channel 1 and channel 2 to be tested to output a frequency of 1000 Hz. 5 Set both the U8903B output channel 1 and channel 2 to be tested to output an amplitude of V as given in the “Test record for sine wave amplitude accuracy...
Page 28
U8903B output channel to be tested. 11 Disable the output of U8903B. 12 Set both the U8903B output channel 1 and channel 2 to be tested to output an amplitude of 1 V. 13 Set both the U8903B output channel 1 and channel 2 to be tested to output a frequency of 1000 Hz (first frequency point in “Test record for flatness verification”...
Performance Verification for Analog Generator Sine wave frequency accuracy verification This test verifies the sine wave frequency accuracy of the signal generated by the U8903B using a frequency counter to ensure that it is within the published specification. Table 2-3...
– Output amplitude: 1 Vrms – Output state: ON 4 Set both the U8903B output channel 1 and channel 2 to be tested to produce a sine wave with amplitude of 1 Vrms and frequency of f as given in the “Test record for frequency accuracy verification”...
6 Set the U8903B analyzer Measured channel input impedance to 300 Ω and analyzer input range to 0.32 Vrms. 7 On the U8903B generator, set the Driving channel output amplitude of Vrms as given in the “Test record for crosstalk verification” on page and the Measured channel output amplitude to 0 Vrms.
Square wave rise time verification In this test, the U8903B generates a square wave swept over a set of frequencies, and the rise time is measured using a Keysight DSO8064A/8104A digital oscilloscope to ensure that it is within the published specification.
Page 33
– Waveform: Square – Output state: ON 4 Set the U8903B output channel 1 and 2 to be tested to produce a square wave with amplitude of V as given in the “Test record for square wave rise time verification”...
Square wave amplitude accuracy verification This test verifies the amplitude accuracy of the square wave signal generated by the U8903B over a range of amplitudes using two digital multimeters to ensure that it is within the published specification. Table 2-6...
Page 35
– Waveform: Square – Output state: ON 4 Set both the U8903B output channels to be tested to produce a square wave with amplitude of V and frequency of 1 kHz as given in the “Test record for square wave amplitude accuracy verification”...
Page 36
Performance Verification for Analog Generator THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight U8903B Service Guide...
CMRR verification Distortion (THD+N) verification SMTPE IMD verification DFD verification Residual noise verification Wideband AC flatness verification Wideband frequency accuracy verification The performance verification procedures described in this chapter verify that the U8903B analog analyzer is operating within its published specifications.
Performance Verification for Analog Analyzer Introduction This section lists the procedures to test the electrical performance of the U8903B analog analyzer using the published specifications given in the U8903B Performance Audio Analyzer User’s Guide (U8903-90045) as the performance standards. If the analog analyzer fails any of the tests or if any abnormal test results are obtained, adjustment will need to be carried out accordingly.
2 Configure the calibrator as follows: – Output state: Standby 3 Configure the U8903B input channels, channel 1 (3, 5, or 7) and channel 2 (4, 6, or 8) as shown in Figure 3-1 to be tested as follows: – Connection type: Unbalanced –...
U8903B input channels to be tested. 9 Repeat step 2 to step 8 to complete the verification of the remaining U8903B input channels. 10 Compare the recorded DC accuracy values to the specification. If the test fails, adjustment will need to be carried out accordingly.
2 Configure the calibrator as follows: – Output state: Standby 3 Configure the U8903B input channels, channel 1 (3, 5, or 7) and channel 2 (4, 6, or 8) as shown in Figure 3-2 to be tested as follows: – Connection type: Unbalanced –...
U8903B and frequency counter inputs. 2 Connect the function generator outputs to the frequency counter input channel 1 and 2, and to the U8903B unbalanced input channel 1 and 2 as shown in Figure 3-3. Preset the U8903B, function generator, and frequency counter.
Unbalanced input channel 2, 4, 6, 8 Frequency counter Figure 3-3 Function generator to frequency counter and U8903B connection for frequency accuracy verification 3 Configure the function generator output channel as follows: – Output load: High-Z load – Output polarity: Normal –...
“Test record for frequency accuracy verification” on page 111. 7 Set the U8903B input channels sample size as given in the “Test record for frequency accuracy verification” on page 111. 8 Set the frequency counter gate time as given in the “Test record for frequency accuracy verification”...
Performance Verification for Analog Analyzer 1 Connect the function generator to the U8903B input channel 1 (3, 5, or 7) and 2 (4, 6, or 8) via the Unbalanced connection as shown in Figure 3-4. The BNC-to-banana jack adapter and the BNC male-to-female T splitter is used for this connection. Preset both the U8903B and function generator.
U8903B input channel 2 (4, 6, or 8). 8 Repeat step 1 to step 7 to complete the verification of the other U8903B input channels. 9 Compare the recorded phase accuracy values to the specification. If the test fails, adjustment will need to be carried out accordingly.
Page 47
1 (3, 5, or 7) in which the Driving channel is channel 2 (4, 6, or 8) and the Measured channel is channel 1 (3, 5, or 7). 13 Repeat step 1 to step 12 to complete the verification of the other U8903B input channels.
Audio generator to U8903B connection for CMRR verification 3 Configure the calibrator as follows: – Output state: Standby 4 Configure the U8903B input channel 1 (3, 5, or 7) and channel 2 (4, 6, or 8) to be tested as follows: – Connection type: Balanced –...
Page 49
“Test record for CMRR verification” on page 113. 8 On the U8903B graph analysis mode, set the graph Marker 1 state to ON and assign it to Trace 1 (3, 5, or 7). Set the graph Marker 2 state to ON and assign it to Trace 2 (4, 6, or 8).
16 Compare the recorded CMRR values to the specification. If the test fails, adjustment will need to be carried out accordingly. Distortion (THD+N) verification This test verifies if the residual THD+N of the U8903B is within the published specification using an internal loopback connection to test the THD+N performance. Table 3-7...
SMTPE IMD verification This test verifies the SMTPE IMD performance of the U8903B analyzer by using an internal loopback connection to ensure that it is within the published specification. Table 3-8...
Page 52
– Input Auto Range: ON – IMD Frequency Detection Mode: Generator frequency lock 4 On the U8903B generator, set the waveform to the first SMPTE function as given in the “Test record for SMPTE IMD verification” on page 115. 5 On the U8903B generator, set the upper frequency to 7 kHz. Set the lower frequency and the amplitude of Vrms as given in the “Test record for SMPTE IMD verification”...
Vrms as given in “Test record for DFD verification” on page 116. 5 On the U8903B analyzer input channels, set the input range as given in “Test record for DFD verification” on page 116. Keysight U8903B Service Guide...
Residual noise verification This test verifies if the residual noise of the U8903B is within the published specification by using shorting connectors to terminate the U8903B input channels and measuring its own residual noise.
Wideband AC flatness verification This test verifies the wideband sine wave flatness of the U8903B analyzer by using the function generator as a source and the digital multimeter as a reference, to ensure that it is within the published specification.
U8903B audio Function generator analyzer Digital Multimeter Figure 3-7 Function generator to U8903B and digital multimeter for wideband sine wave flatness verification 2 Configure the digital multimeter as follows: – Measurement function: AC voltage – Level Filter: Enable – NPLC: 10 –...
6 On the digital multimeter, set the range as given in the “Test record for wideband sine wave flatness verification” on page 117. 7 On the U8903B analyzer, set the input range as given in the “Test record for wideband sine wave flatness verification” on page 117.
Performance Verification for Analog Analyzer 2 Connect the function generator outputs to the frequency counter input channel 1 and 2, and to the U8903B unbalanced input channel 1 and 2 as shown in Figure 3-8. Preset the U8903B, function generator, and frequency counter.
Page 59
U8903B input channel 1 to be tested. 13 Repeat step 1 to step 12 to complete the verification for the U8903B input channel 2. 14 Compare the recorded wideband frequency accuracy values to the specification. If the...
Page 60
Performance Verification for Analog Analyzer THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight U8903B Service Guide...
Page 61
Unbalanced output level accuracy verification Unbalanced input sampling rate accuracy verification Unbalanced output sampling rate accuracy verification The performance verification procedures described in this chapter verify that the U8903B digital analyzer and digital generator is operating within its published specifications.
AES3 and SPDIF digital audio interfaces (fixed perpetual license). N3436A DSI digital audio interface (fixed perpetual license). Refer to the U8903B Audio Analyzer User’s Guide (U8903- 90045) for more information on NOTE the AES3, SPDIF, and DSI interfaces specifications. Keysight U8903B Service Guide...
Equipment used ±5 ppm Frequency counter 1 Connect the Keysight 53132A/53220A frequency counter input channel 1 to the U8903B DSI MCLK output line using the DB25 to BNC cable as shown in Figure 4-1. DB25 to BNC MCLK output cable...
Page 64
Set the DSI audio resolution based on the Word Length as given in “Test record for DSI master clock output frequency accuracy verification” on page 128. Set the DSI audio word length to 32 bits. Keysight U8903B Service Guide...
Page 65
7 Verification steps for word length at 20 bits: Set the DSI audio resolution to 20 bits. b Set the DSI audio word length to 20 bits. Set the DSI Master Clock multiplier to 160. Keysight U8903B Service Guide...
Page 66
Compute the master clock error using the following equation. MCLK is the master clock rate calculated where MCLK = MUL × Sampling Rate (Hz). Δ × ppm = [(F – MCLK) / MCLK] 1000000 error measure Keysight U8903B Service Guide...
Page 67
Set the DSI Master Clock Multiplier based on the Master Clock Multiplier as given in “Test record for DSI master clock output frequency accuracy verification” on page 128. Measure the Master Clock Rate twice and average the two measurements. Keysight U8903B Service Guide...
Balanced output level accuracy verification This test measures the balanced output level accuracy on the balanced digital audio output of the U8903B using the Keysight DSO8064A oscilloscope to ensure that it is within the published specification. Table 4-3 Balanced output level accuracy specification...
– Function 1 to subtract channel 1 and channel 2 – Function 1: Enable – Function 1 offset: 0 3 Configure the U8903B digital output channel to be tested as follows: – Digital output type: Balanced – Digital output voltage: 2.5 V –...
Page 70
10 Repeat step 4 to step 9 for the rest of the F as given in “Test record for balanced output level accuracy verification” on page to complete the verification of the different sampling rates. Keysight U8903B Service Guide...
1 Connect the Keysight 53132A/53220A frequency counter input channel 1 and 2 to the U8903B Balanced digital output channel using the XLR female to two BNC male split AES cable with separated + and – output lines as shown in Figure 4-3.
11 Connect the U8903B Balanced digital output to its own balanced digital input using the AES cable as shown in Figure 4-4. Balanced...
Page 73
13 Set the digital output sampling rate F as given in “Test record for balanced input sampling rate accuracy verification” on page 147. 14 Set the U8903B digital analyzer to measure the sampling rate and record the measurement reading as SR meas...
1 Connect the Keysight 53132A/53220A frequency counter input channel 1 and 2 to the U8903B balanced digital output channel using the XLR female to two BNC male split AES cable with separated + and – output lines as shown in Figure 4-5.
Page 75
– SINE waveform amplitude: 1 FFS – Digital output state: ON 4 Set the U8903B calibration digital output mode to BCOM mode for AES/SPDIF. 5 Set the digital output sampling rate F as given in “Test record for balanced output sampling rate accuracy verification”...
Performance Verification for Digital Generator and Digital Analyzer Unbalanced output level accuracy verification This test measures the level accuracy on the Unbalanced digital audio output of the U8903B using DSO8064A digital oscilloscope to ensure that it is within the published specification. Table 4-7...
10 Repeat step 4 to step 9 for the rest of the F as given in “Test record for unbalanced output level accuracy verification” on page to complete the verification of the different sampling rates. Keysight U8903B Service Guide...
Unbalanced input sampling rate accuracy verification This test measures the sampling rate accuracy on the unbalanced digital audio input of the U8903B using the Keysight 53132A/53220A frequency counter and its own digital audio unbalanced generator to ensure that it is within the published specification.
11 Connect the U8903B unbalanced digital output to its own balanced digital input using the SPDIF or 75 Ω impedance cable as shown in Figure 4-8. Unbalanced digital output SPDIF or 75 Ω...
Page 80
13 Set the digital output sampling rate F as given in “Test record for unbalanced input sampling rate accuracy verification” on page 155. 14 Set the U8903B digital analyzer to measure the sampling rate and record the measurement reading as SR meas...
Performance Verification for Digital Generator and Digital Analyzer Unbalanced output sampling rate accuracy verification This test measures the sampling rate accuracy on the unbalanced digital audio output of the U8903B using the Keysight 53132A/53220A frequency counter to ensure that it is within the published specification. Table 4-10...
Page 82
Performance Verification for Digital Generator and Digital Analyzer 4 Set the U8903B calibration digital output mode to BCOM mode for AES/SPDIF. 5 Set the digital output sampling rate F as given in “Test record for unbalanced output sampling rate accuracy verification” on page 158.
U8903B Bluetooth Test Description Handsfree Profile (HFP) test description Advanced Audio Distribution Profile (A2DP) test description U8903B Bluetooth Retest Procedure The performance verification procedures described in this chapter verify that the U8903B ® Bluetooth card is operating within its published specifications.
Bluetooth card with 2 RF connectors at a secondary option slot. U8903B-BL2 This is only applicable when U8903B-AN8 is selected. Refer to the U8903B Audio Analyzer User’s Guide (U8903- 90045) for more information on NOTE the Bluetooth interfaces specifications. Keysight U8903B Service Guide...
7 Set the Device Under Test (DUT) GPIB address to 28. Test procedure: 1 Connect the Keysight U8903B Tester RF1 port to the DUT RF1 port using the N-type (male) to N-type (male) RF cable as shown in Figure 5-1.
4 Click Connect Instruments to establish the connection between the U8903B Tester unit and the DUT. A green box on the U8903B-BT Tester will appear as shown in Figure 5-3 if the U8903B Tester unit is detected and is able to communicate on the GPIB bus.
9 Select the checkbox for every test in the VEE test program to run a full test. 10 Click Run Selected Test to begin the testing procedure. 11 A pop-up window will appear as a reminder to properly connect the U8903B Tester unit to the DUT as shown in Figure 5-7.
Performance Verification for Bluetooth Card U8903B Bluetooth Test Description The U8903B Bluetooth card supports two Bluetooth audio profiles which are the Handsfree Profile (HFP) and the Advanced Audio Distribution Profile (A2DP). The VEE test program is used to test these two profiles for their functionality. The tests focus on profile connectivity as well as audio generation and measurement functionality.
Figure 5-9 U8903B Bluetooth test for HFP profile (BT Role: Responder, Audio Role: HF) – The second set of Bluetooth tests for the HFP profile is performed with the DUT’s Bluetooth profile role configured as Responder and the audio role configured as Handsfree.
Advanced Audio Distribution Profile (A2DP) test description Figure 5-10 U8903B Bluetooth test for A2DP profile (BT Role: Initator, Audio Role: SRC) – The first set of Bluetooth tests for the A2DP profile is performed with the DUT’s Bluetooth profile role configured as Initiator and the audio role configured as Source.
Performance Verification for Bluetooth Card Figure 5-11 U8903B Bluetooth test for A2DP profile (BT Role: Responder, Audio Role: SNK) – The second set of Bluetooth tests for the A2DP profile is performed with the DUT’s Bluetooth profile role configured as Responder and the audio role configured as Sink.
A2DP APTX functional test points and limits for analyzer and generator Amplitude accuracy Frequency (Hz) Amplitude (FFS) Frequency limit (ppm) THD+N ratio limit (dB) limit (dB) 1000 ±130 -0.3 to 0.1 ≤-45 ±130 -0.8 to 0 ≤-45 18000 ±130 -1 to 0 ≤-18 Keysight U8903B Service Guide...
Figure 5-12, select the appopriate tests and include any prerequisite tests shown in Table 5-6. Figure 5-12 U8903B Bluetooth retest procedure Table 5-6 U8903B Bluetooth test to be repeated and its prerequisite Test number to be repeated Test number that needs to be included 1a and/or 1b 2a and/or 2b...
Page 95
Performance Verification for Bluetooth Card Table 5-6 U8903B Bluetooth test to be repeated and its prerequisite (continued) Test number to be repeated Test number that needs to be included 5a and/or 5b 4 and 5 4 and 5 9a and/or 9b –...
Page 96
Performance Verification for Bluetooth Card THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight U8903B Service Guide...
Keysight U8903B Audio Analyzer Service Guide Appendix A Analog Generator Performance Verification Test record for DC accuracy verification Test record for sine wave amplitude accuracy verification Test record for flatness verification Test record for frequency accuracy verification Test record for crosstalk verification...
Test record for sine wave amplitude accuracy verification (Vrms) (Vrms) Digital multimeter AC accuracy (%) AC accuracy (%) range (Channel 1) (Channel 2) (Channel 1) (Channel 2) 0.008 0.01 0.01 0.01 0.015 0.02 0.03 0.04 0.05 0.08 Keysight U8903B Service Guide...
Test record for flatness verification Table A-3 Test record for flatness verification (Vrms) (Vrms) Digital mul timeter Flatness (dB) Flatness (dB) (Hz) range (Channel 1) (Channel 2) (Channel 1) (Channel 2) 1000 12250 18250 20000 30000 40000 50000 60000 Keysight U8903B Service Guide...
7 and 8. Table A-8 Test record for DC accuracy verification U8903B input range DC accuracy (%) DC accuracy (%) (Channel 1) (Channel 2) (Channel 1) (Channel 2) –140 –100 –32 –10 –3.2 –1 –0.32 0.32 0.32 0.32 Keysight U8903B Service Guide...
Test record for phase accuracy verification Use the test record below for the remaining test channels 3 and 4, channels 5 and 6, and NOTE channels 7 and 8. Table A-11 Test record for phase accuracy verification (Hz) Phase accuracy (°) Keysight U8903B Service Guide...
Use the test record below for the remaining test channels 3 and 4, channels 5 and 6, and NOTE channels 7 and 8. Table A-12 Test record for crosstalk verification Crosstalk (dB) Crosstalk (dB) (Hz) U8903B input range (V) (Channel 1 to Channel 2) (Channel 2 to Channel 1) 0.32 0.32 1000 0.32 0.32...
SMPTe41 SMPTE 4 to 1 SMPTe41 SMPTE 10 to 1 SMPTe101 SMPTE 10 to 1 SMPTe101 SMPTE 10 to 1 SMPTe101 SMPTE 10 to 1 SMPTe101 SMPTE 10 to 1 SMPTe101 SMPTE 10 to 1 SMPTe101 Keysight U8903B Service Guide...
Keysight U8903B Audio Analyzer Service Guide Appendix B Digital Generator and Digital Analyzer Performance Verification Test record for DSI master clock output frequency accuracy verification Test record for balanced output level accuracy verification Test record for balanced input sampling rate accuracy verification...
Test record for balanced output level accuracy verification Sampling rate F (kHz) Level accuracy (dB) measure 28.8 28.8 28.8 28.8 28.8 28.8 28.8 28.8 28.8 28.8 28.8 28.8 28.8 28.8 28.8 28.8 28.8 28.8 28.8 28.8 28.8 Keysight U8903B Service Guide...
Page 142
Test record for balanced output level accuracy verification (continued) Sampling rate F (kHz) Level accuracy (dB) measure 44.1 44.1 44.1 44.1 44.1 44.1 44.1 44.1 44.1 44.1 44.1 44.1 44.1 44.1 44.1 44.1 44.1 44.1 44.1 44.1 44.1 Keysight U8903B Service Guide...
Page 143
Appendix B Table B-2 Test record for balanced output level accuracy verification (continued) Sampling rate F (kHz) Level accuracy (dB) measure 88.2 88.2 88.2 88.2 88.2 88.2 88.2 88.2 88.2 88.2 88.2 88.2 88.2 Keysight U8903B Service Guide...
Page 144
Appendix B Table B-2 Test record for balanced output level accuracy verification (continued) Sampling rate F (kHz) Level accuracy (dB) measure 88.2 88.2 88.2 88.2 88.2 88.2 88.2 88.2 176.4 176.4 176.4 176.4 Keysight U8903B Service Guide...
Page 145
Table B-2 Test record for balanced output level accuracy verification (continued) Sampling rate F (kHz) Level accuracy (dB) measure 176.4 176.4 176.4 176.4 176.4 176.4 176.4 176.4 176.4 176.4 176.4 176.4 176.4 176.4 176.4 176.4 176.4 Keysight U8903B Service Guide...
Page 146
Appendix B Table B-2 Test record for balanced output level accuracy verification (continued) Sampling rate F (kHz) Level accuracy (dB) measure Keysight U8903B Service Guide...
Appendix B Test record for unbalanced output sampling rate accuracy verification Table B-7 Test record for unbalanced output sampling rate accuracy verification Sampling rate F (kHz) (kHz) (ppm) measure error 28.8 44.1 88.2 Keysight U8903B Service Guide...
Page 159
Appendix B Table B-7 Test record for unbalanced output sampling rate accuracy verification (continued) Sampling rate F (kHz) (kHz) (ppm) measure error 176.4 Keysight U8903B Service Guide...
Page 160
Appendix B THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight U8903B Service Guide...
Page 161
Keysight U8903B Audio Analyzer Service Guide Appendix C Modify a Cable for the U8903B CMRR test...
Appendix C Modify a Cable for the U8903B CMRR test 1 Order the following parts as material: Table C-1 Parts to order Part Quantity U8903-61301, male XLR to female XLR, 2 m, Black 1 piece 8120-1840, BNC male to straight BNC 48-in 1 piece 2 Cut one end of the 8120-1840 BNC cable shown in Figure C-1.
XLR pin 2 and 3. Solder the BNC shielding to XLR pin 1. Figure C-5 Solder the BNC cable to the male end of the XLR cable 7 Re-install the XLR metal body. Keysight U8903B Service Guide...
Page 164
Appendix C THIS PAGE HAS BEEN INTENTIONALLY LEFT BLANK. Keysight U8903B Service Guide...