SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings
When the Res BW/Video BW ratio exceeds 10,000, a
may appear to indicate that the Video BW filter has reached the maximum
capacity for averaging.
AF RBW (Audio-Frequency Resolution BW)
AF RBW - sets the audio-frequency resolution bandwidth for the selected
frequency (Default = Auto).
How to set AF RBW
1. Press BW 2 >
2. Then choose from the following:
To change this setting from Man to Auto, press AF RBW twice.
The current AF RBW setting is shown at the bottom of the screen.
# AF RBW x.xx XHz (#) means manual setting.
When the Res BW/AF RBW ratio exceeds 10,000, a
appear to indicate that the AF RBW filter has reached the maximum capacity
for averaging.
Sweep Type
In SA mode, the FieldFox uses two sweep types to process inputs signals. The
sweep type that is currently being used (FFT or Step) is displayed in the
lower-left corner of the FieldFox screen.
For a more comprehensive tutorial, see Spectrum Analysis Basics
(App Note 150) at
https://www.keysight.com/us/en/assets/7018-06714/application-notes/595
2-0292.pdf
How to set Sweep Type
1. Press BW 2 >
2. Then choose from the following:
200
AF RBW
—
Auto
Couples AF BW to the Resolution BW in a 1-to-1 ratio.
—
Man
Enter a AF BW value using the numeric keypad, the
arrows, or the rotary knob. Then press a multiplier if necessary or
press Enter. Setting AF BW higher than RBW may smooth the trace.
Advanced
>
Meas UNCAL
Sweep Type
Keysight N9927-90020 User's Guide
Meas UNCAL
warning
warning may