Pulse Measurements Mode – Option 330
Pulse Measurement Settings
9 - Duty Cycle
10 - Pulse Rep Freq
How to enable Auto Analysis
— Press Mkr ->/Tools.
— Then
— Then choose from the following:
Pulse Top
Pulse Top is NOT the same as the pulse Peak. Pulse Top is used to set and
report the reference power level. You enter a value in percent, and the Pulse
Top value is reported in the Auto Analysis table in dB down from 100% of the
reference power level.
How to set Pulse Top
— Press Mkr ->/Tools
— Then
— Then enter a percent value between 0 and 100 using the numeric keypad,
Grid
Enables and disables the display of a 10 x 10 grid on the FieldFox screen.
How to enable Grid display
— Press Mkr ->/Tools
— Then
Keysight N9927-90020 User's Guide
Auto Analysis
—
OFF
Analysis OFF
—
ON
The Analysis values appear below trace 1.
Pulse Top
the
arrows, or the rotary knob.
— When set to 10%, 10 dB down from the 100% acquisition is reported.
— When set to 50%, 3 dB down from the 100% acquisition is reported.
— When set to 90%, 0.5 dB down from the 100% acquisition is reported.
Grid
—
OFF
Grid OFF
—
ON
A grid of X/Y lines appears on the screen.
The ratio of the pulse width / pulse period.
The rate, in pulses / sec, at which pulses
recur.