Keysight A Series User Manual page 178

Hide thumbs Also See for A Series:
Table of Contents

Advertisement

SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings
3. Press
4. Then use the
5. Optionally choose from the following:
178
New
then
Yes
to clear all data from the existing Antenna or Cable
table and reset header information to default settings.
arrows to select a field.
— To edit header information, press
using the FieldFox labeler.
— Description: Enter a description for the antenna or cable
correction table. This is NOT the filename used to save the
file.
Antenna Factor
measurement Amplitude units will be available or not, and
how to convert between them. Select None (default
setting) if you just want standard Amplitude units, with a
simple dB offset supplied by the correction table. This
makes the correction basically identical to a Cable
correction. To get Field Strength units, you must choose
either dB /m (E-field or dipole antenna), or dB S/m
(H-field or loop antenna) to properly identify the Antenna
Factors in the table. The scalar measurement Amplitude
units (either standard or Field Strength) will then be
available at Ampl/more Unit. Learn how in
Units" on page
units, refer to
Freq
Interpolation: Set to LINear by the FieldFox and can
NOT be edited.
— To edit Frequency/Value pairs, enter numbers using the numeric
keypad, and a frequency suffix when necessary.
— For both
numbers indicate loss (we need ADD something positive to
compensate for it), and Negative is indicative of some
existing gain.
Antenna Factor
accurately display a far-field signal in terms that are in the
units listed below. When the antenna has low gain, or low
efficiency, the dB value to be added is larger.
Antenna Factor Unit
used in your measurement (i.e., dB /m or dB S/m).
— When one Freq/Value pair is entered, that correction value is
applied to the entire displayed frequency span.
— When two or more Freq/Value pairs are entered, the data points
between the lowest and highest frequencies are corrected. Linear
interpolation is applied between frequencies when necessary.
Edit
then modify the selected field
Unit: Determines whether Field Strength
171. For details on standard Amplitude
"How to Set Y-axis Units" on page
Cable Loss
and
Antenna Factor
describes the dB value needed to
sets the type of antenna factor unit
Keysight N9927-90020 User's Guide
"Scale and
172.
values, Positive

Hide quick links:

Advertisement

Table of Contents
loading