Instrument Messages
Event Messages
Err#
Message
-221
Settings conflict;
Feature not supported for
this measurement.
-221
Settings conflict;
Feature not supported for
this model number
-221
Settings conflict; FFT IF
Gain High not available
when Swept IF Gain =
Manual Low
-221
Settings conflict; FFT is
not available when Signal
ID is on
-221
Settings conflict; FFT is
not available when Signal
ID is on
-221
Settings conflict; FFT is
not available with EMI
detector
-221
Settings conflict; FFT
method is unavailable for
level gating
-221
Settings conflict; FFT
sweep type is not
available while in Gated
LO
-221
Settings conflict; FFT
sweep type is not
available while in Gated
Video
-221
Settings conflict; FFT
Sweeps unavailable in
Tracking Source mode
-221
Settings conflict; Fixed
marker adjust not
available while Marker
Function is on
-221
Settings conflict; Fixed
Marker Y value is not
adjustable with Normalize
On
Instrument Messages
Verbose/Correction Information
Some functionality is available in one measurement, but not in
another. (See the measurements under the Meas key.) This error
occurs if you send a SCPI command or push a gray-out key that is
not available in the current selected measurement.
This functionality is not a part of the instrument you are using but
may be found in other models in the X-Series.
When Swept IF Gain is manually set to Low, you cannot set the
FFT IF Gain to High because that would make the Reference Level
couplings wrong in FFT mode.
When in external mixing, if you have signal identification on, then
FFT sweep type cannot be selected.
When in external mixing, if you have signal identification on, then
FFT sweep type cannot be selected.
FFT sweep type cannot be selected when the EMI detector is
being used for a trace.
If you are using level gating, you cannot select the FFT Gate
Method.
The gated LO function turns the LO on and off as it sweeps. So the
FFT sweep type is not available if you have selected gated LO.
The FFT sweep type is not available if you have selected the gated
video function.
Since FFT's do not sweep, you cannot use a Tracking Source
while doing FFT's
If a Marker Function is on for a Fixed marker, the marker's
reported value is derived from the function. Therefore, you cannot
directly set the X or Y value of a Fixed marker that has a marker
function turned on.
If Normalize is on the Amplitude scale is in dB units, so adjusting
the Y value of a Fixed marker is not possible.