VVM (Vector Voltmeter) Mode - Option 308
VVM Mode Settings

How to set Output Power

— Press Meas Setup 4.
— Then
— Then

Averaging

Averaging helps to reduce the effects of random noise on a measurement. The
FieldFox computes the displayed value over several consecutive readings.
Average count determines the number of readings to average. The higher the
average count, the greater the amount of noise reduction.
An average counter is shown in the left edge of the screen as Avg N. This
shows the number of previous readings that have been averaged together.
When the counter reaches the specified count, then a 'running average' of the
last N readings is displayed.
Average count = 1 means there is NO averaging.
This setting can be changed at any time without affecting calibration accuracy.

How to set Averaging

— Press Meas Setup 4
— Then
— Then
— Enter a value using the numeric keypad. Enter 1 for NO averaging.
— Press Enter.
To restart the averaging process, press Meas Setup 4, then More, then
590
More
Output Power
— High (Default setting) Sets output power to the maximum achievable
power at all displayed frequency.
— Low Sets output power to approximately –45 dBm.
— Man Set output power to an arbitrary value.
— Then press
Power Level
— Then enter a value using the numeric keypad, the
rotary knob.
— Press Enter.
More
Avg N
where N is the current count setting.
arrows, or the
Keysight N9927-90020 User's Guide