SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings
These are referred to as
Individual
Align.
All three or a subset of the individual alignments are required for each specific
mode, depending on the type of measurement supported by it:
SA:
— InstAlign Amplitude Alignment
— RF Burst Amplitude Alignment
— Channel Equalization
RTSA:
— InstAlign Amplitude Alignment
— RF Burst Amplitude Alignment
— Channel Equalization
EMI Measurements:
— InstAlign Amplitude Alignment
— WidePath Burst Trigger Alignment
— Channel Equalization
Channel Scanner:
— InstAlign Amplitude Alignment
ERTA:
— InstAlign Amplitude Alignment
IQA:
— InstAlign Amplitude Alignment
— RF Burst Amplitude Alignment
— Channel Equalization
Built-in Power meter
— InstAlign Amplitude Alignment
Default settings and recommended use model for many applications:
— Coupled individual alignments
— Coupled state is Auto
— Individual alignments are not accessible via front panel soft-keys
— Individual alignments cannot be UnCoupled