IQ Analyzer (IQA) – Option 351 (CPU2 Only)
I/Q Analyzer (IQA) Mode Settings
308
refresh rate when the acquisition time is long. Changing the Start
Time changes which data point is displayed on the left side of the
screen.
Couplings: When the change of Stop Time makes time domain
display length exceed the limitation, the Start Time will change to
satisfy time domain display length limitation:
——The maximum value is determined by current Sample Rate and
the maximum capture samples.
——Start Time is adjusted to the nearest sample point based on
the Sample Period.
——The maximum time domain display length is 524,288 points.
Time Domain Display Length [pts] = {[Sample
Rate[pts/s] x (Stop Time[s] - Start Time[s])] +
1}
Where:
pts = points
s = seconds
Stop Time
(Waveform only) Sets the stop time for the waveform
display (default: 100
Waveform mode Setting the Start and stop time enables you to
display a subset of the data on the display in order to speed up the
refresh rate when the acquisition time is long. Changing the Stop
Time changes which data point is displayed on the right side of the
screen.
Couplings: When the change of Stop Time makes time domain
display length exceed the limitation, the Start Time will change to
satisfy time domain display length limitation:
——The maximum value is determined by current Sample Rate and
the maximum capture samples.
——Stop Time is adjusted to the nearest sample point based on the
Sample Period.
——The maximum time domain display length is 524,288 points.
Time Domain Display Length [pts] = {[Sample
Rate[pts/s] x (Stop Time[s] - Start Time[s])] +
1}
Where:
pts = points
s = seconds
Trigger Settings
Opens the Trigger Settings softkey menu.
μ
s, minimum: 0s, maximum: dynamic). In
Keysight N9927-90020 User's Guide