SA (Spectrum Analyzer) Mode (Option 233–Mixed Analyzers)
SA Mode Settings
Figure 9-10
IF Output Wide with CF-33.75 MHz
How to select IF Output
The IF Output signal is useful only in Zero Span. Learn more about Zero Span in
"Zero Span Measurements" on page
At least one sweep must be made in order to tune the FieldFox.
1. Press BW 2 >
2. Then choose from the following:

Sweep Acquisition

Available only in non-zero span measurements, when Sweep Acquisition is set
to Auto, the fastest sweep rate is achieved while maintaining full amplitude
accuracy.
For zero span measurements, refer to
page
202
Advanced
>
OFF
The translated IF signal at the IF OUT connector is no longer
guaranteed to be centered at 33.75 MHz.
Narrow
The IF output signal has approximately 5 MHz bandwidth.
Wide
The IF output signal has approximately 25 MHz bandwidth.
204.
204.
IF Output Use
"Zero Span Measurements" on
Keysight N9927-90020 User's Guide