Setting Up Rapidqc Complete In The Rapidlink Database - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

e. Select
f.
Select
g. Select
h. Enter the system name.
i.
Enter the system serial number.
j.
Select
k. Enter the IP address for the Rapidpoint 400 series system you
copied in step 2e in Removing the Current Rapidpoint 400 Series
System.
l.
Check the Auto Save Settings that the customer had previously
chosen for the original Rapidpoint 400 series system.
m. Select

Setting Up RapidQC Complete in the Rapidlink Database

1. Select the
2. Select
3. Select
4. Choose the Rapidpoint 400 name for the new Rapidpoint 400 series
system that you just set up.
5. Select
6. Verify that all active Parameters are selected.
7. Select
type.
8. Type in Level names (1, 2, and 3).
9. Select
10. Select the Rapidpoint 400 name.
11. Select
12. Select a Level, enter the lot number, the expiration date and start date,
and package insert ranges for each parameter for each level of QC.
13. Repeat step 12 for all three levels of RapidQC Complete.
416
.
Save
.
Add System
Rapidpoint system name
from the Communication Mode options.
CompleNet
and wait for the Device Options screen to appear.
Save
Quality Control
.
QC Setup
.
Add Control
(RapidQC Complete for levels 1, 2, and 3).
Control Name
Target Value and Absolute Limits
.
Save
Setup Active Levels
from the Model list.
tab.
in the Evaluation Technique
.
Rev. C