Replacing The Connector Block Assembly - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Replacing the Connector Block Assembly

BIOHAZARD:
Observe precautions for handling electrostatic-sensitive
CAUTION:
devices. Use an antistatic mat and wrist strap when handling PC boards.
Tools and Supplies
Phillips screwdriver, #1
small flat-blade screwdriver
cutters
Allen wrench, 5/64
needle-nose pliers
pin extractor
1. Turn the system off and disconnect all the external cables from the
back of the system.
2. Remove the rear cover. Refer to Removing the Rear Cover and Power
Module Assembly on page 375.
3. Remove the cartridge interface assembly. Refer to Replacing the
Cartridge Interface Assembly on page 309.
4. Remove the four thread-forming screws securing the thermal cover.
5. Lift the handle.
Rev. C
Refer to Biohazards and Warnings.
Electrical shock hazard. Refer to Biohazards and Warnings.
Replacing Components 313