Replacing the Neon Board Assembly

BIOHAZARD:
Tools and Supplies
Phillips screwdriver, #1
1. Turn the system off and disconnect all the external cables from the
back of the system.
2. Remove the rear cover and power module assembly. Refer to
Removing the Rear Cover and Power Module Assembly on page 375.
3. Remove the neon board assembly:
a. Disconnect the neon power cable (P6).
CAUTION:
the optical fibers resulting in erroneous optics readings.
CAUTION:
the defusing material at each end of the cable as this can result in
erroneous optics readings.
b. Pull up on the spring clip securing the fiber bundle to the neon
board assembly, then grasp the ferrule and remove the fiber
bundle.
358
Refer to Biohazards and Warnings.
Electrical shock hazard. Refer to Biohazards and Warnings.
Do not bend, twist, or crimp the fiber bundle as this can damage
Handle the fiber bundle with care being careful not to damage
Rev. C