The parameter is purple with a line through it, and QC! appears in the lower left
corner.
Probable Cause
Required QC analysis was not
performed when scheduled.
The parameter has a single line through it.
Probable Cause
The parameter failed calibration.
The parameter was turned off in
Setup and then turned on again.
Parameter has two crossed lines through it.
Probable Cause
The parameter failed calibration and is
unlikely to become available with
further calibrations.
158
Corrective Action
1. Analyze all the currently scheduled controls as
described in Analyzing Required QC Samples in
Section 2 in the Rapidpoint 400 Series Operator's
Manual.
2. If the scheduled controls do not contain the
necessary parameters, analyze QC for the
parameters as described in Analyzing
Unscheduled QC Samples in Section 2 in the
Rapidpoint 400 Series Operator's Manual. If the
results are within range, turn the parameter on as
described in Restoring Parameters in Section 2 in
the Rapidpoint 400 Series Operator's Manual.
Corrective Action
1. View the Events Log to determine the event
causing the parameter to fail, and follow the
appropriate solutions for that event.
2. Subsequent calibrations may make the
parameter available again. If the parameter
remains out of calibration, replace the cartridges
as described in Replacing the Measurement and
Wash/Waste Cartridges in Section 3 in the
Rapidpoint 400 Series Operator's Manual.
Perform a 2-point calibration to make the parameter
available again.
Corrective Action
If the failed parameter is required for sample
analysis, replace the cartridges as described in
Replacing the Measurement and Wash/Waste
Cartridges in Section 3 in the Rapidpoint 400 Series
Operator's Manual.
(Continued)