b. Slide the rear cover up and off the system.
4. Remove the power module assembly from the base:
When replacing the display module, slide the power module
NOTE:
assembly toward you until you can access the display cable. It is not
necessary to completely remove the power module assembly.
a. Remove the 14 screws securing the power module frame to the
base.
As you pull the frame back, be careful not to stress the fiber bundles
NOTE:
in the clamp on the left side of the frame.
b. Slightly spread the sides at the bottom of the power module frame
while slowly pulling the frame straight back to expose the cables.
CAUTION:
fiber bundle as this can damage the optical fibers resulting in erroneous
optics readings.
c. Disconnect the hard drive cable from the UIP board.
d. Remove the fiber bundles from the cable clamps on the frame.
e. Disconnect the patch cord (Ethernet cable) from the UIP board.
f.
Disconnect the rear panel interconnect cable from the Main board.
g. Disconnect the main power harness from the Main board.
h. Remove the fiber bundles (terminal B) from the neon board
assembly.
376