Replacing The Sample Port - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Replacing the Sample Port

Use this procedure to replace the sample port. Replace the sample port if
you detect a problem, or if the system prompts you as a result of the
following conditions:
no sample device inserted during analysis
obstructions such as fibrin clots
bubbles in the sample
insufficient sample volume
sample not detected
problems with reagent flow
Material:
sample port
BIOHAZARD:
recommended precautions when working with biohazardous materials.
If the system prompts you, go to step 3 in the following procedure.
NOTE:
1. At the Analysis screen, touch the
2. At the Status screen, touch
3. When prompted, replace the sample port. Refer to Figure 4-1.
Touch the Video button to view a demonstration that shows how to
NOTE:
perform this procedure.
a. Squeeze the tabs on the sample port and remove it from the
system.
Dispose of the sample port according to your institution's protocol
for disposal of biohazardous materials.
b. Squeeze the tabs on the new sample port.
c. Insert the sample port on the system and release the tabs.
Rev. C
Refer to Protecting Yourself from Biohazards, for
button.
Status
.
Replace Port
Replacing Components 303