Replacing The Diskette Drive Assembly - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Replacing the Diskette Drive Assembly

BIOHAZARD:
Observe precautions for handling electrostatic-sensitive
CAUTION:
devices. Use an antistatic mat and wrist strap when handling PC boards.
Tools and Supplies
Phillips screwdriver, #1
1. Turn the system off and disconnect all the external cables from the
back of the system.
2. Remove the rear cover and power module assembly. Refer to
Removing the Rear Cover and Power Module Assembly on page 375.
3. Disconnect the following cables:
Disconnect the . . .
diskette power harness
diskette drive cable
4. Remove the diskette drive from the interface wall:
If required, remove the spring clip on the bottom of the diskette drive
NOTE:
bracket to access the bottom screw.
a. Move the display screen forward.
Rev. C
Refer to Biohazards and Warnings.
Electrical shock hazard. Refer to Biohazards and Warnings.
From the . . .
Main board
UIP board
Replacing Components 323