3. A yellow dialog box appears and indicates that the measurement
cartridge needs to be replaced.
4. Touch
Continue
5. Touch
Replace
The door opens.
If you are unsure which mode the system is in, close the door with no
cartridges in place.
If the system is in . . .
the USER mode
the FOR DEMONSTRATION
ONLY mode
6. Once you verify the system is in the USER mode, install new real
measurement and wash/waste cartridges, and close the door.
7. The system begins an initialization process with a countdown clock
that begins at 20:00.
The system stays in the USER mode when rebooted with a valid
NOTE:
real measurement cartridge in place.
When replacing an AutomaticQC cartridge, if you eject the
NOTE:
old cartridge and do not have the replacement cartridge available, the
Return button appears on the screen, after a 5 minute wait period. Touch
this button to interrupt the replacement. The system continues to calibrate
or run samples while you obtain the AutomaticQC cartridge.
424
.
on the
Cartridge Status
Then . . .
a yellow dialog box appears indicating that the
measurement cartridge needs to be replaced.
the system starts a simulated initialization.
In other words, when in FOR
DEMONSTRATION ONLY mode, the system
indicates that a measurement cartridge is in
place when it is not.
screen.
Rev. C