a. Loosen the two screws securing the wash/waste switch to the rear
of the interface wall. Do not remove the screws.
b. From the front of the system push the wash/waste switch assembly
towards the back, and from the side push it towards the center of
the system and remove from the rear of the system.
c. Disconnect the wash/waste connector from the outboard cable
harness.
7. Install the new wash/waste switch assembly:
a. Place the switch assembly on the interface wall shelf, orientation
will be at an angle.
b. Pull the switch up so that the shelf is positioned between the
bracket and the switch.
c. Tighten the screws. Align the switch with the lever protruding
through the wash/waste access cavity.
d. Connect the wash/waste connector to the outboard cable harness.
8. Reassemble in the reverse order. Refer to Installing the Rear Cover,
Cartridge Interface Assembly, and Power Module Assembly on
page 378.
9. Replace the AQC frame assembly, the ampule breaker (if present),
and cover assembly.
Rev. C
Replacing Components 355