Message
D70 Optics Error:
9, 12
Qualifiers:
9 Alignment
EEPROM Read Error
12 Wavelength Shift
Error
206
Probable Cause and Corrective Action
Service Solutions
Perform service solutions only after all other corrective actions
have been completed.
For Error 3:
1. Repeat 2-point calibrations.
2. Verify the lamp is in the correct position. Remove and
reinstall the halogen lamp. Then shut down and restart the
system.
3. Verify that the rear cover does not shift the lamp position.
Remove the rear cover, reinstall the lamp, and restart the
system (with cover removed).
4. Replace the fiber bundle assembly.
For Error 4 and 7:
1. Repeat 2-point calibrations.
2. Check that the screws are tight on the polychromator to
ensure it is properly grounded.
3. Remove the polychromator ribbon cable and reattach.
Power up the system.
4. If the error persists, replace the polychromator module.
For Error 11:
1. Repeat 2-point calibrations.
2. Perform the Wavelength Calibration test without the
cartridge.
3. Check all the connections to the fiber bundle assembly.
4. Replace the neon board assembly.
The system detects a failure in the CO-ox optical components.
The system can measure samples and report results for
NOTE:
parameters that are still available.
1. Shut down the system as described in Shutting Down the
System, page 256, wait 10 seconds, and then turn the
system on.
2. If the message appears again, record the message,
including the qualifier, and call for technical assistance.
(Continued)
Rev. C