d. Wiggle the sample port from side to side to ensure it is attached
correctly.
Replacing the Sample Port
1 Touch to view the video.
2 Squeeze the tabs firmly.
3 Remove the sample port from the system.
4 Squeeze the tabs on the new sample port and install the sample port in the
system.
5 Touch the
4. Touch the
The system performs a wash to clear the system.
CAUTION:
clot or fluid is still present a system failure occurs invalidating the
measurement cartridge.
304
button when you finish installing the sample port.
Continue
button after replacing the sample port.
Continue
Do not put the same sample port onto the system because if a
Rev. C