Co-ox Sample Chamber

The CO-ox sample chamber is part of the measurement cartridge.The
sample chamber consists of the following components:
base
gasket
seal
shunt/flag
sample chamber body
spring
cover
heater chip
The sample chamber is in its measurement position when the motor moves
the shunt into the closed position.
When the shunt returns to the open position, the measurement
cartridge reagent system cleans the sample chamber. A heater chip
ensures that the sample temperature is kept at 37

Polychromator Module

The polychromator module measures the intensities of light passed
through the sample at a number of different wavelengths.
The polychromator module has the following components:
60
0.3°C.
±