When Setup Is Complete - Bayer HealthCare CLINITEK 500 Operating Manual

Urine chemistry analyzer
Hide thumbs Also See for CLINITEK 500:
Table of Contents

Advertisement

SELECTING YOUR OPTIONS
2.
e. Display: The display can be tested to ensure that
all pixels (the lighted elements on the display) are
being lit and turned off appropriately. When the
Display key is touched, the entire screen will be lit
for several seconds, then be blank. This series will
be repeated twice more before returning to the
previous menu. If there are numerous faulty pixels,
or if they are located in critical areas, the display
may need to be replaced.
2.
f. Printer: The internal and/or 80-column external
printer can be tested to ensure that all characters
are being printed correctly. Ensure that the exter-
nal printer is turned on (if one is being used). Touch
the Printer key and follow the directions on the
screen. Examine the printout for its readability. The
display automatically returns to the previous menu.
3. When all tests have been completed, touch
to any previous Setup menu, or touch
the Ready/Run screen.
3.16

When Setup is Complete

Touch
When you have finished selecting the setup parame-
ters, touch
screen. The changes you made will NOT be saved if you
do not touch
It is suggested that you print a copy of the setup report
to verify your selections and to retain in your files. Touch
Menu from the Ready/Run screen, then touch Print. From
the Print menu that is displayed, select the option Setup
report. (If printing from the internal printer, make a
photocopy of the report, since the thermal print may fade
over time.)
The setup parameters are stored both in the CLINITEK
500 Analyzer and in the program card. If a new program
to return
card is installed, you can select to use your current setup
to return to
configuration from the Analyzer memory. Conversely, if
you receive a new CLINITEK 500 Analyzer, but have
retained your old program card, you can select to use your
current setup configuration from the program card. If both
the Analyzer and the program card are new, you can use
the printout obtained previously to reselect the desired
parameters.
to save!
to return the display to the Ready/Run
.
Revised 5/01