a. If prompted, enter your password, the service password, or the
password of the day.
b. Touch the
c. Touch
d. Touch
e. Touch
f.
Touch
g. Ensure that the following options are selected for the
communication settings:
Setting
Baud
Parity
Data Bits
Modem
h. Touch the
8. At the Communications screen, touch the
The Wait screen appears while the system tests the connection
between the Rapidpoint 400 series system and the Rapidlink system.
At this time, the connection is unsuccessful.
9. When prompted that the connection was unsuccessful, touch
to display the Communications screen.
10. At the Rapidlink workstation, enter the communication settings as
described in Appendix B of the Rapidlink Supervisor's Manual.
11. At the Rapidpoint 400 series system, touch the
from the Communications screen.
The Wait screen appears while the system tests the connection
between the systems.
12. Refer to the Rapidlink Supervisor's Manual to complete any additional
setup that may be necessary to operate the Rapidpoint 400 series
system with the Rapidlink program.
406
button.
Status
.
Setup
Printer and Devices
.
Communications
and then touch
RS-232
button.
Continue
.
.
Configure
Default Selection
9600
Even
8
None (no modem connected)
Continue
button.
Continue
button to exit
Continue
Rev. C