Bayer HealthCare RapidPoint 400 Series Service Manual page 399

Table of Contents

Advertisement

b. Touch
The system plays a video that shows how to replace the cartridges.
c. View the video before you begin if required.
d. Remove the wash/waste cartridge.
e. Lift up the latch that holds the measurement cartridge in place until
the cartridge is ejected.
f.
Lift the measurement cartridge up and out of the system.
g. Lower the latch.
h. Close the door.
A message appears indicating that the cartridge is not installed
correctly and the door opens again.
i.
Close the door again.
j.
Touch
The Status screen appears.
6. Shut down the system:
a. Touch the
b. Touch
A message appears asking if you really want to shut down the
system.
c. Touch
d. When the Shutdown screen appears, turn off the power switch and
disconnect the power cord from the electrical outlet and from the
system.
7. Disconnect the bar code scanner if present.
8. Remove the roll of printer paper and remove any remaining paper from
the printer by turning the paper-advance knob clockwise.
9. Empty the ampule breaker if present.
10. Push the display fully back toward the system.
11. The system is ready for storage or packing.
Rev. C
.
Yes
on the message.
Cancel
button.
Status
.
Shutdown
.
Yes
Installing and Relocating the Rapidpoint 400 Series System 399