Analog And Other Output Signals - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Input Signal
HCT2_HS
HCT2_LS
THM_BLK_HI
THM_BLK_LO
THM_PRE_HI
THM_PRE_LO
THM_SMP_HI
THM_SMP_LO

Analog and Other Output Signals

Output Signal
PO2_SIG
LAC_A
LAC_B
GLU_A
GLU_B
CRE_A
CRE_B
PH_SIG
HCO3_SIG
BUN_SIG
K_SIG
MG_CL_SIG
NA_SIG
CA_SIG+
72
Description
Positive signal output of the hematocrit 2 sensor. This sensor
connects to the positive input of the synchronous sampling
differential amplifier on the Contact board. The amplifier
circuitry measures the impedance (conductance) between
the HCT_HF and HCT_LF contacts of the hematocrit 2
sensor.
Signal reference of the hematocrit 2 sensor. This sensor
connects to the negative input of the synchronous sampling
differential amplifier on the Contact board.
One terminal of the block thermistor.
The other terminal of the block thermistor.
One terminal of the preheater thermistor.
The other terminal of the preheater thermistor.
One terminal of the sample thermistor.
The other terminal of the sample thermistor.
Description
Amplified and buffered pO
For future use.
For future use.
Amplified and buffered glucose (inactive) sensor signal.
Amplified and buffered glucose (active) sensor signal.
For future use.
For future use.
Amplified and buffered pH sensor signal.
Amplified and buffered HCO
For future use.
Amplified and buffered K
Amplified and buffered Cl- sensor signal.
Amplified and buffered Na
Amplified and buffered Ca
sensor signal.
2
sensor signal.
3
+
sensor signal.
+
sensor signal.
++
sensor signal.
(Continued)
Rev. C