Replacing the Display/Printer Assembly

BIOHAZARD:
CAUTION:
devices. Use an antistatic mat and wrist strap when handling PC boards.
Remove the diskette drive from the interface wall before removing
NOTE:
the display module. Refer to Replacing the Diskette Drive Assembly on
page 323.
Tools and Supplies
Phillips screwdriver, #1
small flat-blade screwdriver
torque driver
1. Turn the system off and disconnect all the external cables from the
back of the system.
2. Remove the ampule breaker or Automatic QC cartridge and open the
front door.
3. Remove the rear cover and power module assembly. Refer to
Removing the Rear Cover and Power Module Assembly on page 375.
4. Remove the cartridge interface assembly. Refer to Replacing the
Cartridge Interface Assembly on page 309.
326
Refer to Biohazards and Warnings.
Electrical shock hazard. Refer to Biohazards and Warnings.
Observe precautions for handling electrostatic-sensitive
Rev. C