Troubleshooting Problems; Button And Parameter Status - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Troubleshooting Problems

Button and Parameter Status

Refer to this section when buttons and parameters at the Analysis screen
are not available.
The buttons that you touch to select the patient sample are not available.
The QC syringe button is also unavailable.
Probable Cause
You replaced the measurement
cartridge but have not yet analyzed
the controls that are scheduled for the
newly installed cartridge.
The QC! button on the Analysis screen is not available.
Probable Cause
The lot is expired for the control that is
currently scheduled, or no lot is
entered for the control.
The parameter is yellow with a line through it, and QC appears in the lower left
corner.
Probable Cause
The sensor has drifted since the last
automatic calibration.
Rev. C
Corrective Action
Analyze the Required QC samples as described in
Analyzing Required QC Samples in Section 2 in the
Rapidpoint 400 Series Operator's Manual.
Corrective Action
1. Define a new lot of the control as described in
Defining New Lots of Controls for Required QC in
Section 6 in the Rapidpoint 400 Series Operator's
Manual.
2. Analyze the scheduled controls as described in
Analyzing Required QC Samples in Section 2 in
the Rapidpoint 400 Series Operator's Manual.
Corrective Action
For pH, pO
, pCO
, Na
2
2
perform a 2-point calibration.
For Hct or tHb, perform a full calibration.
+
+
++
-
, K
, Ca
, Cl
, and Glu,
(Continued)
Troubleshooting 155