Shipping Or Storing The System - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

6. Move the system to its new location.
7. To restore power to the system, connect the power cord to the electrical
outlet and turn on the power switch.
8. After the Rapidpoint 400 series system title screen appears, the Wait
screen displays the time remaining until you can use the system. The
Analysis screen appears when the system is ready to use.
If the system has the DHCP option selected and is moved to a
CAUTION:
location that is connected to a different segment on the network, network
connection problems may occur. Reboot the system so that the DHCP
queries the network for the correct address.

Shipping or Storing the System

Use this procedure to prepare the system for long term storage, shipping
to another location for repair, or for disposal. When you are ready to
dispose of the system, observe country, state, and local codes or
requirements for recycling.
power from the system as described in this procedure.
1. Disinfect the exterior surfaces of the system. Refer to Cleaning and
Disinfecting the Exterior Surfaces in Section 3 in the Rapidpoint 400
Series Operator's Manual.
2. If prompted, enter your password, the service password, or the
password of the day.
3. Remove the AutomaticQC cartridge if present:
Refer to Replacing the AutomaticQC Cartridge in Section 3 in the
Rapidpoint 400 Series Operator's Manual, if required.
a. Touch the
b. Touch the
A message appears indicating that the cartridge does not need
replacing.
c. Touch
The system plays a video that shows how to replace the cartridge.
Rev. C
To prevent electrical shock or damage to the system, remove
button.
Status
AutomaticQC Cartridge
.
Yes
Installing and Relocating the Rapidpoint 400 Series System 397