Installing And Relocating The Rapidpoint 400 Series System; Installing The System; Calibrating The - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Installing and Relocating the Rapidpoint 400 Series
System
This appendix contains procedures that describe how to install the
Rapidpoint 400 series system for the first time and how to relocate the
system.
When installing two or more new Rapidpoint 400 series systems to
NOTE:
the same network, connect the network cables during the Wait period when
the system is testing the connections between the Rapidpoint 400 series
systems and the Rapidlink systems.

Installing the System

Do not install the measurement and wash/waste cartridges until
CAUTION:
the appropriate system message appears indicating that the measurement
cartridge needs replacing. Installing the measurement cartridge before this
message appears, invalidates the measurement cartridge.
1. Unpack the shipping box and remove any packing materials from the
system.
2. Place the system on a bench top or other level work surface.
The system must equilibrate to room temperature to successfully
NOTE:
complete the installation. If the ambient temperature is near the limits of the
recommended operating temperature (15 to 32°C), the touch screen can
become out of calibration. Calibrate the screen if required. Refer to
Calibrating the Touch Screen in Calibrations.
3. If you purchased the optional bar code scanner, connect the bar code
scanner to the system:
a. Connect the bar code cable to the bar code scanner connector.
The bar code scanner connector is located on the rear panel of the
system and is labeled with a bar code symbol.
b. Tighten the hold-down screws on the connector.
c. Attach the holder for the bar code scanner to the right side of the
system.
Rev. C

Installing and Relocating the Rapidpoint 400 Series System 391