These parameters are available only on Rapidpoint 405 systems.
In addition to sensor chips, the sensor module contains:
•
fluid detector chips
•
a preheater chip
•
an IDEE ROM chip
The Rapidpoint 400 series systems use two Hct sensors as fluid detectors.
Fluid detector 1 (FD1) is located in front of the first sensor. Fluid detector 2
(FD2) is located after the last sensor.
The fluid detector chips detect the presence and the continuity of fluids in
the system, and measure sample fluid using alternating current (AC)
conductivity. The system uses information from the detectors to monitor
and control the movement and the positioning of fluids during the system
sequences.
Rev. C
Parameter
total hemoglobin; tHb
oxyhemoglobin; FO
2
carboxyhemoglobin;
†
FCOHb
methemoglobin; FMetHb
hemoglobin oxygen
†
saturation; sO
2
oxygen binding capacity;
†
BO
2
oxygen tension at 50%
†
saturation; p50
Description
†
the total of all measured hemoglobin
fractions
†
Hb
hemoglobin that is reversibly bound to
oxygen
hemoglobin covalently bound to
carbon monoxide
†
hemoglobin whose iron is oxidized to
its ferric state and is unable to bind
oxygen
the ratio of the amount of hemoglobin
bound to oxygen to the total amount of
hemoglobin able to bind oxygen
the maximum amount of oxygen that
can be carried by the hemoglobin in a
given quantity of blood
the partial pressure of oxygenated
which the hemoglobin oxygen
saturation is 50%
Mechanical Descriptions 39