Retrospective Calibration (Retrocal) - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

The system measures three calibrators (Wash, 200 Cal, and Zero Cal)
for all sensors.The system measures the drift and calculates the
mathematical slope and the offset for all analytes. A 2-point calibration
takes 3 minutes and 55 seconds (3:55) to complete.
A full 2-point calibration (slope and offsets for all analytes):
A 2-point calibration adjusts both the offset and the slope drift for a
parameter by measuring two reagents of known concentration.
The system performs a full 2-point calibration every 8 hours.
The system measures four calibrators (Wash, 200 Cal, Zero Cal, and
Hct Cal) on all sensors. The system measures the drift and calculates
the mathematical slope and offset. A full 2-point calibration takes
5 minutes and 50 seconds (5:50) to complete. Every third full 2-point
calibration includes a blood/fiber removal sequence preceding the
calibration.

Retrospective Calibration (Retrocal)

A patient sample run during the initialization (init) period includes a
retrospective calibration (retrocal) before results are reported. A
retrospective calibration is a 1-point calibration immediately following the
sample. During the init period, if a patient sample is run, the following
occurs:
The sample is loaded.
An electronic measurement is performed on the sample.
The system washes out the sample with 200 Cal and loads 200 Cal
over the sensors.
The system performs a 1-point calibration by measuring drift for all
analytes.
The system calculates the patient data and displays sample results.
The system performs a wash.
This sequence from start of the sample being loaded until the wash is
complete takes 3 minutes. Sample results are posted in 2 minutes.
428
Rev. C

Advertisement

Table of Contents
loading

Table of Contents