a. Loosen the four captive screws securing the AutomaticQC latch
assembly. Do not remove the screws.
1 AQC latch assembly
2 Captive screws (4)
3 Cable to the AQC frame assembly
b. Remove the copper AutomaticQC clip from between the
AutomaticQC frame assembly and the base wall.
c. Disconnect the cable (J43) from the AutomaticQC frame
assembly.
While pulling the AutomaticQC latch assembly from the system,
guide the cable connector through the slot in the wall interface.
Rev. C
Replacing Components 321
2
3