DO NOT apply power with a new real measurement cartridge in place.
DO NOT apply power with any simulator cartridge in place.
DO NOT attempt to reuse a partially used measurement cartridge that
was removed from the system.
DO NOT leave a cartridge on an instrument after it was removed to
perform diagnostics testing with the simulator cartridge.
Setting the Rapidpoint 400 Series System to the
FOR DEMONSTRATION ONLY Mode
RULE:
The system must have power applied with the demonstration cartridge
in place. There is no alternate way to access the FOR
DEMONSTRATION ONLY mode from the USER mode.
1. Apply power to the system without a measurement, a wash waste, or
an AutomaticQC cartridge in place, or when the system is already
running with real cartridges or no cartridge.
Do not use a real wash/waste cartridge.
CAUTION:
2. Touch
Replace
demonstration measurement cartridge, a demonstration wash waste
cartridge, and a demonstration AutomaticQC cartridge.
When in the USER mode, ignore the M Cartridge Not Valid
NOTE:
message generated by the demonstration cartridges.
3. Touch the
switch.
4. Wait a minimum of 30 seconds.
5. Turn the power switch on.
6. The system detects the IDEE ROM in the demonstration
measurement cartridge and selects the
FOR DEMONSTRATION ONLY
Rev. C
on the Cartridge Status screen, and install a
button and when prompted, turn off the power
Shutdown
software.
mode
Rapidpoint 400 Series Cartridge Usage 421