Advertisement

3. Touch
Replace
The system releases the access panel that covers the connections for
the AutomaticQC cartridge.
Also, the system plays a video that shows how to replace the
AutomaticQC cartridge, refer to the following figure, Releasing the
Access Panel.
Releasing the Access Panel
1
The system releases the access panel. Remove the panel.
4. View the video before you begin. Observe the steps for replacing the
cartridge.
5. Remove the access panel.
6. Install the AutomaticQC cartridge:
Refer to Replacing the AutomaticQC Cartridge in Section 3 in the
Rapidpoint 400 Series Operator's Manual, if required.
a. Get the new AutomaticQC cartridge.
b. Remove the yellow card from under the lever.
Be sure the lever locks in place.
NOTE:
c. Push down on the lever firmly, near the raised dots, to close and
lock the lever in the cartridge.
Rev. C
.
Installing and Relocating the Rapidpoint 400 Series System 395