c. With the spring pawl positioned in the slot, use a flat-blade
screwdriver to press the spring pawl down, toward the front of the
system, and hold it down while inserting the display support
through the slot until it locks into place.
1 Flat-blade screwdriver holds the spring pawl in place
2 Spring pawl down toward the front of the system
3 Display support
If the spring pawl jams, position a screwdriver behind the spring
NOTE:
pawl and push it forward while sliding the display support forward one tooth
at a time until correctly seated.
d. Tilt the display module back completely. Locate the hinge brackets
at the base of the display and pull the hinge brackets forward.
Tighten the two screws located at the base of the display.
e. Reconnect the display cable to the Main board.
14. Reassemble in the reverse order. Refer to Installing the Rear Cover,
Cartridge Interface Assembly, and Power Module Assembly on
page 378.
15. Close the front door and reinstall the ampule breaker.
Rev. C
Replacing Components 329