Bayer HealthCare RapidPoint 400 Series Service Manual page 71

Table of Contents

Advertisement

Input Signal
-5_VDC
ID_PWR
ID_RET
PO2_SIG
GLU_WB
GLU_WA
LAC_WB
LAC_WA
CRE_WB
CRE_WA
PH_VO
K_VO
CA_VO
MG_CL_VO
NA_VO
BUN_VO
HCO3_VO
REF_VO
HCT1_HS
HCT1_LS
Rev. C
Description
Negative power supply for the potentiometric amplifiers.
Supplies -5.00 V ±5% at 5 mA.
Power supply for the EEPROM in the sensor module.
Supplies +5.00 V ±5% at 5 mA.
Power supply return signal for the EEPROM.
Current output signal of the pO
Current output signal of the glucose (active) electrode.
Current output signal for an inactivated twin of the glucose
electrode.
For future use.
For future use.
For future use.
For future use.
Voltage output signal of the pH sensor.
Voltage output signal of the K
Voltage output signal for the Ca
Voltage output of the Cl- sensor.
Voltage output signal of the Na
For future use.
Voltage output signal of the HCO
Voltage output signal of the reference sensor. The range is
-60 to 140 mV. Impedance is about 10K ohms.
Signal output of the hematocrit 1 sensor. This sensor
connects to the positive input of the synchronous sampling
differential amplifier on the Contact board. The amplifier
circuitry measures the impedance (conductance) between
the HCT_HF and the HCT_LF contacts of the hematocrit 1
sensor.
Signal reference of the hematocrit 1 sensor. This sensor
connects to the negative input of the synchronous sampling
differential amplifier on the Contact board.
electrode.
2
+
sensor.
++
sensor.
+
sensor.
sensor.
3
Electronic System Overview 71