What Can Go Wrong?

If . . .
the message
FOR DEMONSTRATION ONLY
is not displayed
a yellow dialog box appears and
indicates that the cartridge needs
to be replaced
Attempting to go from a real cartridge in the USER mode to a
demonstration cartridge without removing power from the system
generates the diagnostic message, M Cartridge Not Valid.
If a real cartridge is installed in the FOR DEMONSTRATION ONLY mode,
a diagnostic error message is displayed, but no damage occurs to the
cartridge. There is no IDEE write capability in the FOR DEMONSTRATION
ONLY mode.

Setting the Rapidpoint 400 Series System to the USER Mode

The system must have power applied with no cartridges in place.
RULE:
There is no way to directly access the USER mode from the FOR
DEMONSTRATION ONLY mode.
1. Apply power to the system without a measurement or a wash/waste
cartridge in place.
When the system does not detect the presence of a cartridge IDEE
ROM, the USER mode software is installed.
2. The USER mode software loads. This takes a few minutes.
Rev. C
Then . . .
1. Touch
on the Cartridge Status screen.
Replace
2. Open the door and verify that the cartridges are
in fact demonstration cartridges.
3. Remove and replace the demonstration
cartridges with particular attention to proper
seating, and close the door.
4. Touch
Shutdown
the power switch.
5. Wait a minimum of 30 seconds.
6. Turn the power switch on and verify that
FOR DEMONSTRATION ONLY is displayed on
the Selection screen.
1. Touch
.
Continue
2. Touch
on the
Replace
The door opens.
3. Wait a few seconds and then close the door.
Rapidpoint 400 Series Cartridge Usage 423
and when prompted, turn off
Cartridge Status
screen.