CAUTION:
the defusing material at each end of the cable as this can result in
erroneous optics readings.
b. Reconnect the polychromator fiber bundle and fiber bundle to the
sample chamber interface assembly.
7. Check that the fiber bundle is not bunched or crimped.
8. Reassemble in the reverse order. Refer to Installing the Rear Cover,
Cartridge Interface Assembly, and Power Module Assembly on
page 378.
9. After reassembly, ensure that the optics head pivots, springs, and
moves easily from side-to-side.
10. Perform all system tests to ensure proper system operation.
366
Handle the fiber bundles with care being careful not to damage
Rev. C