Temp Warning - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

Message

Temp Warning

Temp Warning Descriptions
In Retro Calibration Mode
If the measurement module temperature is outside the range 37 ± 0.4°C
Temp warning exception is set.
As soon as the temperature gets within the range 37 ± 0.40°C, the temp
warning message disappears.
Rev. C
Probable Cause and Corrective Action
The system detects that the temperature of the sensor module is
outside the range of 37
a new measurement cartridge is warming up, when the system is
warming up after being shut down, and when the door remains
open for too long.
If necessary, you can analyze samples while this message is
displayed.
1. If possible, wait until the system removes the message from the
Status screen, indicating the temperature is now within range,
before analyzing samples.
2. If the message appears again, record the message and call for
technical assistance.
Service Solutions
Perform service solutions only after all other corrective actions
have taken place
Ensure that the instrument and measurement cartridge have
reached normal operating temperature, then:
1. Check that the system location is not causing the ambient
temperature to be too high or too low.
2. View the events log for errors that may be causing the problem.
3. If the message occurs frequently:
a. Replace the connector block.
b. Replace the cartridge interface assembly.
°
C ± 0.20°. This message can occur when
Troubleshooting 225