Bayer HealthCare RapidPoint 400 Series Service Manual page 325

Table of Contents

Advertisement

e. Slide diskette drive out through the slot.
When inserting the new diskette drive, the 100-pin display cable
CAUTION:
must be located beneath the diskette drive bracket or the display will not
pivot correctly when adjusted.
Do not cross-thread the screws that secure the diskette drive
CAUTION:
bracket to the interface wall. The screws seat in brass inserts and cross-
threading effectively welds the screws to the inserts.
5. Install the new diskette drive:
a. Slide the diskette drive into position and replace the two screws.
b. Reconnect the diskette power harness to the Main board.
Ensure cables do not catch on the RFI shielding fingers.
CAUTION:
c. Replace the diskette drive cable, reconnect the diskette drive
cable in the cable clamp and to the UIP board.
d. Ensure that the diskette drive is flush with the side of the system
and adjust the screws as necessary.
6. Reassemble in the reverse order. Refer to Installing the Rear Cover,
Cartridge Interface Assembly, and Power Module Assembly on
page 378.
Rev. C
Replacing Components 325