Relocating The System - Bayer HealthCare RapidPoint 400 Series Service Manual

Table of Contents

Advertisement

d. Insert the cartridge into the system, and then push firmly on the
raised dots until you hear the cartridge lock in place.
e. Slide the connector to the left to close it.
The Wait screen appears while the system prepares the cartridge.
The Analysis screen appears when the cartridge is ready for use.
7. Define the schedule for AutomaticQC analysis as described in
Selecting AutomaticQC Analysis Options in Section 6 in the
Rapidpoint 400 Series Operator's Manual.

Relocating the System

Use this procedure to move the system for use in another area. If you plan
to move the system frequently, Bayer Diagnostics recommends that you
connect the system to an uninterruptible power supply and then move the
system while it is still connected to the uninterruptible power supply. This
ensures that you do not affect the performance of installed cartridges by
leaving the system without power for more than 60 minutes.
When relocating the system, a change in ambient temperature of
NOTE:
greater than 5°C may cause the screen to become out of calibration.
Calibrate the screen if required. Refer to Calibrating the Touch Screen in
Calibrations.
power from the system as described in this procedure.
1. If prompted, enter your password, the service password, or the
password of the day.
2. Touch the
3. Touch
Shutdown
CAUTION:
without power. To maintain cartridge stability, do not remove power from
the system for more than 60 minutes if a cartridge is installed.
4. When the Shutdown screen appears, turn off the power switch and
disconnect the power cord from the electrical outlet.
5. Disinfect the exterior surfaces of the system. Refer to Cleaning and
Disinfecting the Exterior surfaces in Section 3 in the Rapidpoint 400
Series Operator's Manual.
396
To prevent electrical shock or damage to the system, remove
button.
Status
.
Cartridges installed in the system remain stable for 60 minutes
Rev. C

Advertisement

Table of Contents
loading

Table of Contents